Design, Material, and Seating Load Effects on In Vitro Fretting Corrosion Performance of Modular Head-Neck Tapers.

Eric S Ouellette1, Sachin A Mali2, Jua Kim3, Justin Grostefon4, Jeremy L Gilbert5.   

Abstract

BACKGROUND: The short-term corrosion and micromechanical behavior of 32 unique head-neck taper design/material/assembly conditions was tested using an incremental cyclic fretting corrosion (ICFC) test method previously developed.
METHODS: Seven materials, design, and simulated surgical parameters were evaluated, each being assigned 2 conditions for testing, using a 27-2 (7 factor, quarter factorial) design of experiments test matrix. The factors explored were (1) seating load, (2) head-neck offset, (3) material combination, (4) taper diameter, (5) taper roughness, (6) angular mismatch/engagement, and (7) taper length. Each sample underwent assembly, ICFC testing, pull off.
RESULTS: Low seating load and high head offset correlated with increased fretting corrosion (P < .05). High head offset also contributed to a lower onset load for fretting current and higher micromotion (P < .05). Head subsidence measured over the ICFC test for samples seated at 100 N was significantly higher than samples seated at 4000 N. Micromotion for 12-mm head offsets was statistically higher than samples with a 1.5-mm head offset. A number of interactive effects were observed. For example, samples seated at 4000 N were less sensitive to head offset than samples seated at 100 N in terms of the resulting fretting current.
CONCLUSION: Taper locking position, material combination, taper engagement length, taper roughness, and taper dimensions all had weak or no correlation with fretting current and taper micromotion. This test method and experimental design is a versatile means of assessing potential new taper designs in the future.
Copyright © 2019 Elsevier Inc. All rights reserved.

Keywords:  MACC; THA; corrosion; electrochemistry; fretting; micromotion

Year:  2019        PMID: 30827717     DOI: 10.1016/j.arth.2019.01.043

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  2 in total

1.  Effect of head size and rotation on taper corrosion in a hip simulator.

Authors:  Christian M Wight; Cari M Whyne; Earl R Bogoch; Radovan Zdero; Ryan M Chapman; Douglas W van Citters; William R Walsh; Emil Schemitsch
Journal:  Bone Jt Open       Date:  2021-11

2.  Importance of surgical assembly technique on the engagement of 12/14 modular tapers.

Authors:  A Wade; F Webster; A R Beadling; M G Bryant
Journal:  Proc Inst Mech Eng H       Date:  2021-10-25       Impact factor: 1.617

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.