| Literature DB >> 30799977 |
Dora Vuljanić1, Ana Dojder1, Valentina Špoljarić1, Andrea Saračević1, Lora Dukić1, Jasna Leniček-Krleža2,3, Jelena Vlašić-Tanasković4,3, Ivana Maradin5,3, Ana Grzunov2,3, Željka Vogrinc6, Ana-Maria Šimundić1.
Abstract
INTRODUCTION: Variability among manufacturers of urine dipsticks, respective to their accuracy and measurement range, may lead to diagnostic errors and thus create a serious risk for the patient. Our aims were to determine the level of agreement between 12 most commonly used urine dipsticks in Croatia, examine their accuracy for glucose and total protein and to test their repeatability.Entities:
Keywords: accuracy; comparability; repeatability; urine dipsticks; verification
Mesh:
Substances:
Year: 2019 PMID: 30799977 PMCID: PMC6366948 DOI: 10.11613/BM.2019.010708
Source DB: PubMed Journal: Biochem Med (Zagreb) ISSN: 1330-0962 Impact factor: 2.313
Most common urine dipstick brands and manufacturers in Croatia, used in this study
| 1 | Combur 10 Test M | Roche (Mannheim, Germany) |
| 2 | ChoiceLine 10 | Roche (Mannheim, Germany) |
| 3 | Combur 10 Test UX | Roche (Mannheim, Germany) |
| 4 | ComboStik 10M | DFI Co., Ltd. (Gimhae, South Korea) |
| 5 | ComboStik 11M | DFI Co., Ltd. (Gimhae, South Korea) |
| 6 | CombiScreen 11SYS | Analyticon (Lichtenfels, Germany) |
| 7 | CombiScreen 10SL | Analyticon (Lichtenfels, Germany) |
| 8 | Combina 13 | Human (Wiesbaden, Germany) |
| 9 | Combina 11S | Human (Wiesbaden, Germany) |
| 10 | Combina 10M | Human (Wiesbaden, Germany) |
| 11 | UriGnost 11 | BioGnost Ltd. (Zagreb, Croatia) |
| 12 | Multistix 10SG | Siemens (Erlangen, Germany) |
Agreement between 11 most common dipstick brands in Croatia with the reference Combur 10 Test M (Roche)
| ChoiceLine 10 | 0.90 | 0.89 | 0.76 | 0.82 | 0.73 | / | 0.89 | 0.97 | 0.71 | 0.81 |
| Combur 10 Test UX | 0.99 | 0.93 | 0.94 | 0.85 | 0.92 | / | 0.90 | 0.97 | 0.95 | 0.90 |
| ComboStik 10M | 0.89 | 0.87 | 0.75 | 0.71 | 0.71 | / | 0.51 | 0.97 | 0.40 | 0.31 |
| ComboStik 11M | 0.86 | 0.87 | 0.72 | 0.78 | 0.69 | / | 0.46 | 0.97 | 0.43 | 0.32 |
| CombiScreen 11SYS | 0.90 | 0.87 | 0.79 | 0.71 | 0.71 | 0.54 | 0.78 | 1.00 | 0.87 | 0.64 |
| CombiScreen 10SL | 0.89 | 0.87 | 0.76 | 0.70 | 0.80 | 0.51 | 0.74 | 1.00 | 0.87 | 0.62 |
| Combina 13 | 0.84 | 0.79 | 0.60 | 0.71 | 0.84 | 0.16 | 0.36 | 0.97 | 0.46 | 0.42 |
| Combina 11S | 0.88 | 0.81 | 0.76 | 0.68 | 0.71 | 0.44 | 0.81 | 0.97 | 0.79 | 0.60 |
| Combina 10M | 0.91 | 0.87 | 0.78 | 0.72 | 0.80 | / | 0.19 | 1.00 | 0.53 | 0.41 |
| UriGnost 11 | 0.83 | 0.87 | / | 0.78 | 0.85 | 0.33 | 0.85 | 0.97 | 0.88 | 0.49 |
| Multistix 10SG | 0.80 | 0.87 | 0.71 | 0.71 | 0.78 | / | 0.89 | 0.97 | 0.56 | 0.54 |
| Darker grey fields represent the highest κ-values (κ ≥ 0.80); lighter grey fields show lower κ-values (κ < 0.80); white fields represent excluded parameters (/). Glc – glucose. Prot – total protein. Erc – erythrocytes. Leu – lekocytes. Ket – ketones. Bil – bilirubin. Ubg – urobilinogen. Nit – nitrite. pH - acidity or basicity. SG – specific gravity. | ||||||||||
Repeatability of 12 most common dipstick brands in Croatia (assessed on 20 replicates for all parameters).
| Combur 10 Test M | 20/20 | 19/20 | 20/20 | 20/20 | 20/20 | 20/20 | 20/20 | 20/20 | 20/20 | 20/20 |
| ChoiceLine 10 | 19/20 | 17/20 | 20/20 | 20/20 | 20/20 | 20/20 | 20/20 | 20/20 | 20/20 | 20/20 |
| Combur 10 Test UX | 20/20 | 20/20 | 20/20 | 20/20 | 20/20 | 20/20 | 20/20 | 20/20 | 20/20 | 20/20 |
| ComboStik 10M | 20/20 | 20/20 | 19/20 | 20/20 | 20/20 | 20/20 | 20/20 | 20/20 | 20/20 | 20/20 |
| ComboStik 11M | 20/20 | 20/20 | 19/20 | 20/20 | 20/20 | 18/20 | 18/20 | 20/20 | 20/20 | 20/20 |
| CombiScreen 11SYS | 16/20 | 20/20 | 20/20 | 20/20 | 19/20 | 18/20 | 15/20 | 20/20 | 20/20 | 20/20 |
| CombiScreen 10SL | 20/20 | 16/20 | 20/20 | 20/20 | 20/20 | 20/20 | 20/20 | 20/20 | 20/20 | 20/20 |
| Combina 13 | 19/20 | 11/20 | 20/20 | 20/20 | 19/20 | 20/20 | 20/20 | 19/20 | 20/20 | 18/20 |
| Combina 11S | 18/20 | 20/20 | 20/20 | 20/20 | 19/20 | 20/20 | 20/20 | 20/20 | 20/20 | 20/20 |
| Combina 10M | 20/20 | 20/20 | 18/20 | 20/20 | 19/20 | 20/20 | 20/20 | 20/20 | 20/20 | 20/20 |
| UriGnost 11 | 19/20 | 20/20 | 17/20 | 20/20 | 20/20 | 20/20 | 20/20 | 18/20 | 20/20 | 20/20 |
| Multistix 10SG | 20/20 | 18/20 | 20/20 | 20/20 | 20/20 | 20/20 | 20/20 | 13/20 | 20/20 | 20/20 |
| Grey fields represent parameters that did not meet the acceptance criteria. SG – specific gravity. pH – acidity or basicity. Leu – lekocytes. Nit – nitrite. Prot – proteins. Glc – glucose. Ket – ketones. Bil – bilirubin. Ubg – urobilinogen. Erc – erythrocytes. | ||||||||||
The analytical sensitivities and specificities for glucose for 12 most common dipsticks in Croatia with hexokinase method as a reference
| Combur 10 Test M | Roche | 97.0% | 81.0% |
| ChoiceLine 10 | Roche | 96.3% | 75.0% |
| Combur 10 Test UX | Roche | 97.0% | 83.3% |
| ComboStik 10M | DFI Co., Ltd. | 80.0% | 80.0% |
| ComboStik 11M | DFI Co., Ltd. | 73.3% | 80.0% |
| CombiScreen 11SYS | Analyticon | 89.3% | 76.6% |
| CombiScreen 10SL | Analyticon | 85.7% | 76.6% |
| Combina 13 | Human | 69.7% | 92.9% |
| Combina 11S | Human | 95.8% | 70.6% |
| Combina 10M | Human | 93.1% | 80.4% |
| UriGnost 11 | BioGnost Ltd. | 72.7% | 97.6% |
| Multistix 10SG | Siemens | 67.7% | 93.2% |
| Grey fields represent acceptable sensitivity or specificity (light grey fields ≥ 80%, darker grey > 90%). | |||
The analytical sensitivities and specificities for urinary total protein for 12 most common dipsticks in Croatia with pyrogallol red molybdate assay and turbidimetric method with benzethonium chloride as a references
| Combur 10 Test M | Roche | 69.8% | 75.0% | 87.2% | 72.2% |
| ChoiceLine 10 | Roche | 69.2% | 69.4% | 81.8% | 64.3% |
| Combur 10 Test UX | Roche | 66.7% | 66.7% | 85.7% | 65.0% |
| ComboStik 10M | DFI Co., Ltd. | 60.0% | 71.4% | 77.8% | 69.2% |
| ComboStik 11M | DFI Co., Ltd. | 60.0% | 71.4% | 77.8% | 69.2% |
| CombiScreen 11SYS | Analyticon | 61.5% | 69.4% | 75.8% | 66.7% |
| CombiScreen 10SL | Analyticon | 60.0% | 71.4% | 73.5% | 68.3% |
| Combina 13 | Human | 41.0% | 72.2% | 55.9% | 70.7% |
| Combina 11S | Human | 75.0% | 45.7% | 85.3% | 41.5% |
| Combina 10M | Human | 70.0% | 62.9% | 91.7% | 66.7% |
| UriGnost 11 | BioGnost Ltd. | 70.7% | 70.6% | 86.1% | 66.7% |
| Multistix 10SG | Siemens | 67.5% | 74.3% | 80.0% | 67.5% |
| Light grey fields represent the highest (≥ 80%) and dark grey fields the lowest (< 60%) sensitivities and specificities. | |||||