Use of bismuth shield for protection of superficial radiosensitive organs in patients undergoing computed tomography: a literature review and meta-analysis.

Parinaz Mehnati1,2, Reza Malekzadeh3,4, Mohammad Yousefi Sooteh2.   

Abstract

The study aimed to assess the effect of bismuth (Bi) shielding on dose reduction and image quality in computed tomography (CT) through a literature review. A search was conducted in the following databases: Web of Science, PubMed, Google Scholar, and Scopus. Studies that reported estimated dose reduction with bismuth shielding during imaging of the eye, thyroid, and breast were included, and a meta-regression analysis was used to examine the influence of the CT scanner type on the dose reduction. The studies included a total of 237 patients and 34 pediatric and adult anthropomorphic phantoms for whom the radiation dose was reported. Bismuth shielding was recommended in 88.89% of the studies based on the maintenance of appropriate image quality under shielding. Noise associated with Bi shielding was 7.5%, 263%, and 23.5% for the eye, thyroid, and breast, respectively. The fixed-effects pooled estimate of dose reduction was 34% (95% CI: 13-55; p < 0.001) for the eye, 37% (95% CI 14-61; p < 0.001) for the thyroid, and 36% (95% CI 36-55; p < 0.001) for the breast. The image quality, usage of foams, CT scanner type, beam energies, and backscatter radiation were important factors that directly affected the efficacy of Bi shielding to reduce the radiation dose at the superficial radiosensitive organs.

Entities:  

Keywords:  Bismuth shielding; Computed tomography; Dose reduction; Image quality

Mesh:

Substances:

Year:  2019        PMID: 30790174     DOI: 10.1007/s12194-019-00500-2

Source DB:  PubMed          Journal:  Radiol Phys Technol        ISSN: 1865-0333


  3 in total

1.  Lens dose reduction with a bismuth shield in neuro cone-beam computed tomography: an investigation on optimum shield device placement conditions.

Authors:  Satoru Kawauchi; Koichi Chida; Yusuke Hamada; Wataro Tsuruta
Journal:  Radiol Phys Technol       Date:  2021-11-18

2.  Shielding effect of radiation dose reduction fiber during the use of C-arm fluoroscopy: a phantom study.

Authors:  Hyemi Cha; Kisung Lee; Moon Seok Park; Kyoung Min Lee; Kyeyoung Cho; Ki Hyuk Sung
Journal:  J Radiat Res       Date:  2020-09-08       Impact factor: 2.724

3.  Bismuth Pelvic X-Ray Shielding Reduces Radiation Dose Exposure in Pediatric Radiography.

Authors:  Bow Wang; Chien-Yi Ting; Cheng-Shih Lai; Yi-Shan Tsai
Journal:  Biomed Res Int       Date:  2021-10-11       Impact factor: 3.411