Myocardial blood flow quantitation in patients with congestive heart failure: head-to-head comparison between rapid-rotating gantry SPECT and CZT SPECT.

Rongzheng Ma1, Lei Wang1, Dayong Wu1, Meng Wang1, Xiaoxin Sun2, Bailing Hsu3,4, Wei Fang5.   

Abstract

PURPOSE: Recently, the feasibility of myocardial blood flow (MBF) quantitation using rapid-rotating gantry (RRG) and cadmium-zinc-telluride (CZT) SPECT cameras has been demonstrated. We compared MBF quantitation using these two camera systems.
METHODS: Twenty patients with congestive heart failure (CHF) and 20 patients without CHF (non-CHF) were included. On two consecutive days, dynamic SPECT imaging was performed after a bolus injection of 20 mCi of 99mTc-Sestamibi (MIBI) with RRG-SPECT and list-mode acquisition with CZT-SPECT. All dynamic SPECT images were reconstructed with full physical corrections, corrections for ventricular spillover and partial volume effect, using one-tissue kinetic modeling. Resting MBF converted from K1 was then corrected for MIBI extraction fraction and adjusted for rate-pressure product.
RESULTS: In both patient groups, there was no significant difference between resting MBF values measured with RRG-SPECT and CZT-SPECT systems (P = 0.06, P = 0.2 respectively). For CHF patients, linear regression (LR) was y(RRG) = 1.0412x (CZT) (r = 0.97) with a small systemic difference (Δ = 0.03 mL·min-1·g-1, 95% CI - 0.11 to 0.20) by Bland-Altman analysis. For non-CHF patients, LR was y(RRG) = 1.025x (CZT) (r = 0.89) with also small systemic difference (ΔT= 0.02 mL·min-1·g-1, 95% CI - 0.14 to 0.19) in BA analysis.
CONCLUSION: Physical corrections along with other image corrections can provide comparable MBF quantitations in both CHF and non-CHF patients, regardless of the type of SPECT systems used.

Entities:  

Keywords:  CZT SPECT; SPECT myocardial blood flow quantitation; congestive heart failure; correction of 99mTc-Sestamibi extraction fraction; corrections of left and right ventricle spillovers; physical corrections; rapid-rotating SPECT

Mesh:

Substances:

Year:  2019        PMID: 30783997     DOI: 10.1007/s12350-019-01621-2

Source DB:  PubMed          Journal:  J Nucl Cardiol        ISSN: 1071-3581            Impact factor:   5.952


  1 in total

1.  Compton scatter compensation using the triple-energy window method for single- and dual-isotope SPECT.

Authors:  T Ichihara; K Ogawa; N Motomura; A Kubo; S Hashimoto
Journal:  J Nucl Med       Date:  1993-12       Impact factor: 10.057

  1 in total
  2 in total

1.  Leveraging latest computer science tools to advance nuclear cardiology.

Authors:  Piotr Slomka
Journal:  J Nucl Cardiol       Date:  2019-09-05       Impact factor: 5.952

2.  Diagnostic efficiency of quantification of myocardial blood flow and coronary flow reserve with CZT dynamic SPECT imaging for patients with suspected coronary artery disease: a comparative study with traditional semi-quantitative evaluation.

Authors:  Jiao Wang; Shuai Li; Weiqiang Chen; Yue Chen; Zekun Pang; Jianming Li
Journal:  Cardiovasc Diagn Ther       Date:  2021-02