| Literature DB >> 30782884 |
Yoonje Lee1, Hyungoo Shin1, Hyuk Joong Choi2, Changsun Kim2.
Abstract
OBJECTIVE: The purpose of this study was to assess whether a photoplethysmography (PPG) sensor in a smart watch can accurately recognise the return of spontaneous circulation (ROSC) in cardiac arrest patients compared with carotid artery palpation.Entities:
Keywords: health informatics
Mesh:
Year: 2019 PMID: 30782884 PMCID: PMC6367979 DOI: 10.1136/bmjopen-2018-023627
Source DB: PubMed Journal: BMJ Open ISSN: 2044-6055 Impact factor: 2.692
Figure 1Flow diagram.
Pulse checking results for the return of spontaneous circulation for all conventional methods and the three smart watch method
| Method of pulse checking | Sensitivity (%) | Specificity (%) | False positive (%) | False negative (%) |
| Carotid palpation | 78.6 | 90.4 | 9.6 | 21.4 |
| Pulse check by smart watch | ||||
| In the carotid region | 100 | 78.7 | 21.3 | 0 |
| In the forehead region | 92.9 | 80.9 | 19.1 | 7.1 |
| In the wrist region | 78.6 | 100 | 0 | 21.4 |
Figure 2Receiver operating characteristic curves of each method including manual palpation and the smart watch on the carotid, forehead and wrist area under the curve, (95% CI); carotid palpation: 0.847 (0.779 to 0.900); smart watch on the carotid: 0.894 (0.833 to 0.938); smart watch on the forehead: 0.867 (0.802 to 0.917); smart watch on the wrist: 0.893 (0.832 to 0.937). AUC, area under the curve.
Figure 3Comparison of the receiver operating characteristic curves for manual palpation and the smart watch on the carotid, forehead and wrist smart watch on the carotid versus wrist: p=0.99; smart watch on the carotid versus forehead: p=0.21; smart watch on the carotid versus carotid palpation: p=0.19; smart watch on the wrist versus forehead: p=0.44; smart watch on the wrist versus carotid palpation: p<0.01; smart watch on the forehead versus carotid palpation: p=0.51.
Individual evaluation of false-positive cases using photoplethysmography with a smart watch and the palpation method
| ROSC case number | Manual check | Smart watch | Vital sign with HR regularity | ||
| Carotid | Forehead | Wrist | |||
| ROSC 1 | No | Yes | Yes | Yes | BP 60/40, HR 98, regular rhythm |
| ROSC 2 | No | Yes | Yes | No | BP 64/41, HR 86, irregular rhythm |
| ROSC 3 | Yes | Yes | Yes | No | BP 72/46, HR 71, irregular rhythm |
| ROSC 4 | No | Yes | Yes | Yes | BP 69/41, HR 80, regular rhythm |
| ROSC 5 | Yes | Yes | Yes | Yes | BP 70/40, HR 56, regular rhythm |
| ROSC 6 | Yes | Yes | Yes | Yes | BP 118/69, HR 103, irregular rhythm |
| ROSC 7 | Yes | Yes | Yes | Yes | BP 102/55, HR 121, regular rhythm |
| ROSC 8 | Yes | Yes | Yes | Yes | BP 100/70, HR 62, regular rhythm |
| ROSC 9 | Yes | Yes | Yes | No | BP 107/80, HR 72, irregular rhythm |
| ROSC 10 | Yes | Yes | Yes | Yes | BP 172/83, HR 102, regular rhythm |
| ROSC 11 | Yes | Yes | Yes | Yes | BP 140/80, HR 55, regular rhythm |
| ROSC 12 | Yes | Yes | No | Yes | BP 142/72, HR 110, irregular rhythm |
| ROSC 13 | Yes | Yes | Yes | Yes | BP 104/81, HR 65, regular rhythm |
| ROSC 14 | Yes | Yes | Yes | Yes | BP 90/50, HR 58, regular rhythm |
BP, blood pressure; HR, heart rate; ROSC, return of spontaneous circulation.