Klaus Koren1, Maria Moßhammer2, Vincent V Scholz3, Sergey M Borisov4, Gerhard Holst5, Michael Kühl2,6.
Abstract
Luminescence lifetime based imaging is still the most reliable method for generating chemical images using chemical sensor technology. However, only few commercial systems are available that enable imaging lifetimes within the relevant nanosecond to microsecond range. In this technical note we compare the performance of an older time-domain (TD) based camera system with a frequency-domain (FD) based camera system regarding their measuring characteristics and applicability for O2 and pH imaging in environmental samples and with different indicator dye systems emitting in the visible and near-infrared part of the spectrum. We conclude that the newly introduced FD imaging system delivers comparable if not better results than its predecessor, now enabling robust and simple chemical imaging based on FD luminescence lifetime measurements.Entities:
Year: 2019 PMID: 30758940 DOI: 10.1021/acs.analchem.8b05869
Source DB: PubMed Journal: Anal Chem ISSN: 0003-2700 Impact factor: 6.986