| Literature DB >> 30756263 |
Stefan van Duijvenboden1,2, Ben Hanson3, Nick Child4, Pier D Lambiase5,6, Christopher A Rinaldi4, Gill Jaswinder4, Peter Taggart5, Michele Orini3,6.
Abstract
Mechanical alternans (MA) is a powerful predictor of adverse prognosis in patients with heart failure and cardiomyopathy, but its use remains limited due to the need of invasive continuous arterial pressure recordings. This study aims to assess novel cardiovascular correlates of MA in the intact human heart to facilitate affordable and non-invasive detection of MA and advance our understanding of the underlying pathophysiology. Arterial pressure, respiration, and ECG were recorded in 12 subjects with healthy ventricles during voluntarily controlled breathing at different respiratory rate, before and after administration of beta-blockers. MA was induced by ventricular pacing. A total of 67 recordings lasting approximately 90 s each were analyzed. Mechanical alternans (MA) was measured in the systolic blood pressure. We studied cardiovascular correlates of MA, including maximum pressure rise during systole (dPdtmax), pulse arrival time (PAT), pulse wave interval (PI), RR interval (RRI), ECG QRS complexes and T-waves. MA was detected in 30% of the analyzed recordings. Beta-blockade significantly reduced MA prevalence (from 50 to 11%, p < 0.05). Binary classification showed that MA was detected by alternans in dPdtmax (100% sens, 96% spec), PAT (100% sens, 81% spec) and PI (80% sens, 81% spec). Alternans in PAT and in PI also showed high degree of temporal synchronization with MA (80 ± 33 and 73 ± 40%, respectively). These data suggest that cardiac contractility is a primary factor in the establishment of MA. Our findings show that MA was highly correlated with invasive measurements of PAT and PI. Since PAT and PI can be estimated using non-invasive technologies, these markers could potentially enable affordable MA detection for risk-prediction.Entities:
Keywords: Electrical alternans; Mechancial alternans; Pulse arrival time; Pulse transit time
Mesh:
Year: 2019 PMID: 30756263 PMCID: PMC6453876 DOI: 10.1007/s10439-019-02221-4
Source DB: PubMed Journal: Ann Biomed Eng ISSN: 0090-6964 Impact factor: 3.934
Figure 1Overview of the obtained measurements from the blood pressure and ECG waveforms. SBP systolic blood pressure, dPdt (dP/dt)max, Rw peak of the R-wave, Tp peak of the T-wave, PI pulse interval, PAT pulse arrival time, RRI RR interval. Note the broad QRS complexes as a result of ventricular pacing (corresponding pacing artefacts are visible at t = 0 s and t = 0.5 s).
Figure 2Temporal agreement between mechanical alternans (MA) and alternans in the pulse arrival time (PATA). Upper panel shows the raw BP trace with temporal episodes of MA around 10 and 20 s. The lower panel show the temporal agreement between MA (indicated by grey shading) and PATA: TP true positive: alternans in both systolic blood pressure (SBP) and PAT, TN true negative: absence of alternans in both markers, FP false positive: PATA during absence of MA, FN false negative: Absence of PATA during MA.
Figure 3Examples of blood pressure and ECG traces during presence of MA (a) and absence (b). SBP systolic blood pressure, PAT pulse arrival time.
Figure 4Prevalence of MA for respiratory frequency (a) and control vs. beta-blockade (b).
Figure 5ROC curves for tested markers to predict recordings with mechanical alternans. RRI RR interval, PAT pulse arrival time, PI pulse wave interval, QRSA QRS-complex alternans, TWA T-wave alternans.
Accuracy, sensitivity, and specificity of investigated cardiovascular markers to detect mechanical alternans evaluated in the testing data-set.
| Prev. (%) | MA | DPDTA | PATA | PIA | RRIA | TWA | QRSA | |
|---|---|---|---|---|---|---|---|---|
| MA | 30 | 1 (1/1) | 0.97 (1/0.96) | 0.87 (1/0.81) | 0.81 (0.8/0.81) | 0.78 (0.5/0.89) | 0.79 (0.85/0.77) | 0.7 (0.3/0.87) |
| DPDTA | 33 | 1 (1/1) | 0.87 (0.95/0.82) | 0.84 (0.82/0.84) | 0.75 (0.45/0.89) | 0.79 (0.82/0.78) | 0.67 (0.27/0.87) | |
| PATA | 43 | 1 (1/1) | 0.76 (0.66/0.84) | 0.64 (0.34/0.87) | 0.81 (0.76/0.84) | 0.63 (0.28/0.89) | ||
| PIA | 37 | 1 (1/1) | 0.73 (0.44/0.9) | 0.72 (0.68/0.74) | 0.66 (0.28/0.88) | |||
| RRIA | 22 | 1 (1/1) | 0.69 (0.73/0.67) | 0.78 (0.4/0.88) | ||||
| TWA | 42 | 1 (1/1) | 0.73 (0.39/0.97) | |||||
| QRSA | 18 | 1 (1/1) |
MA mechanical alternans, DPDTA dPdtmax alternans, PATA pulse arrival time alternans, PIA pulse wave interval alternans, TWA T-wave alternans, QRSA QRS-complex alternans
Figure 6Accuracy (Acc), sensitivity (Sens), and specificity (Spec) to predict mechanical alternans using alternans in pulse arrival time (PAT) and pulse width interval (PIA) for different analysis window lengths (L). The percentage of recordings with alternans is shown between brackets on top of the bars.
Temporal agreement between mechanical alternans and alternans of the investigated cardiovascular markers: mean tSens ± standard deviation and mean tSpec ± standard deviation.
| tSens | tSpec | |
|---|---|---|
| DPDTA | 0.93(± 0.15) | 0.7(± 0.38) |
| PATA | 0.8(± 0.3) | 0.74(± 0.38) |
| PIA | 0.73(± 0.4) | 0.54(± 0.43) |
| RRIA | 0.48(± 0.46) | 0.68(± 0.46) |
| QRSA | 0.51(± 0.43) | 0.62(± 0.43) |
| TWA | 0.64(± 0.42) | 0.55(± 0.43) |
For abbreviations see Table 1