Image quality and dose evaluation of MVCT TomoTherapy acquisitions: A phantom study.

P De Marco1, I Abdi Osman2, F Castellini3, R Ricotti3, M C Leonardi3, E Miglietta3, R Cambria4, D Origgi4, B A Jereczek-Fossa5, C Garibaldi6, F Cattani4.   

Abstract

BACKGROUND: The aim of this study is to evaluate the dose delivered and the image quality of pre-treatment MVCT images with Hi-Art TomoTherapy system, varying acquisition and reconstruction parameters.
MATERIALS AND METHODS: Catphan 500 MVCT images were acquired with all acquisition pitch and reconstruction intervals; image quality was evaluated in terms of noise, uniformity, contrast linearity, contrast-to-noise ratio (CNR) and spatial resolution with the Modulation Transfer Function (MTF). Dose was evaluated as Multi Slice Average Dose (MSADw) and measurements were performed with the Standard TomoTherapy® Quality Assurance Kit composed by the TomoTherapy Phantom, the Exradin A1SL ion chamber and TomoElectrometer. For each pitch-reconstruction interval, acquisitions were repeated 5 times.
RESULTS: Differences in noise and uniformity, though statistically significant in some cases, were very small: noise ranged from 2.3% for Coarse - 3 mm to 2.4% for Coarse - 6 mm, while uniformity passed from 99.5% for Coarse - 6 mm to 99.8% for Normal - 4 mm. No differences at all were found for CNR for high and low density inserts, while MTF was higher for pitch Coarse, even if no differences in spatial resolution were observed visually (spatial resolution was up to 4 lp/cm for all combinations of pitch and reconstruction interval). Dose was dependent on pitch, being 1.0 cGy for Coarse, 1.5 cGy for Normal and 2.85 cGy for Fine.
CONCLUSIONS: We observed negligible differences in image quality among different pitch and reconstruction interval, thus, considerations regarding pre-treatment imaging modalities should be based only on dose delivered and on the desired resolution along the cranio-caudal axis for image-guided radiotherapy and adaptive radiotherapy purposes.
Copyright © 2019 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Keywords:  Image guided radiotherapy; Image quality; Megavoltage computed tomography; Pre-treatment imaging; Tomotherapy

Mesh:

Year:  2019        PMID: 30738526     DOI: 10.1016/j.ejmp.2019.01.009

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  2 in total

1.  Clinical implementation of kVCT-guided tomotherapy with ClearRT.

Authors:  Bin Yang; Hui Geng; Tien Yee Amy Chang; Mei Yan Tse; Wai Wang Lam; Chen-Yu Huang; Tungho Wu; Ka Ki Lau; Kin Yin Cheung; George Chiu; Siu Ki Yu
Journal:  Phys Eng Sci Med       Date:  2022-08-04

2.  TomoMQA: Automated analysis program for MVCT quality assurance of helical tomotherapy.

Authors:  Min Cheol Han; Chae-Seon Hong; Kyung Hwan Chang; Jihun Kim; Su Chul Han; Dong Wook Kim; Kwangwoo Park; Min-Joo Kim; Han Back Shin; Se Hyung Lee; Jin Sung Kim
Journal:  J Appl Clin Med Phys       Date:  2020-04-08       Impact factor: 2.102

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.