PURPOSE: Evaluation of the diagnostic accuracy of optic nerve sheath diameter (ONSD) and Doppler indices of central retinal arteries and veins for the detection of increased intracranial pressure (ICP) in intracerebral hemorrhage (ICH) and of the usefulness of a second assessment of these variables in the monitoring of ICH.
METHODS: A total of 46 acute ICH patients with (group 1, n = 25) and without (group 2, n = 21) clinical and radiological computed tomography signs of raised ICP and 40 healthy controls were recruited. The median binocular ONSD and Doppler indices of retinal vessels including resistive index (RI) and retinal venous pulsation (RVP) were compared among groups, both at admission and later during ICH monitoring.
RESULTS: Median binocular ONSD showed higher accuracy for the detection of increased ICP (sensitivity and specificity 100%), while Doppler indices were less accurate (sensitivity 48% and specificity 95% for RI; 80% and 62% for RVP). In ICH patients, ONSD was significantly elevated in group 1 both at admission (6.40 mm [interquartile range [IQR] = .70] vs. 4.70 [.40]) and at control time (6.00 [.55] vs. 4.55 [.40]; P < .01), as well as RI (.79 [.11] vs. .77 [.03] and .80 [.06] vs. .75 [.35]; P = .01). RVP was significantly increased in group 1 only at admission (3.20 cm/s [1.05] vs. 2.00 [1.55], P = .02).
CONCLUSIONS: Median binocular ONSD evaluation showed higher accuracy for the estimation of elevated ICP compared with Doppler indices of retinal vessels. The ONSD enlargement detected in the early phase of ICH persists at control time.
© 2019 by the American Society of Neuroimaging.

Entities:  

Keywords:  intracerebral hemorrhage; intracranial pressure; optic nerve sheath diameter; retinal vessels

Mesh:

Year:  2019        PMID: 30714259     DOI: 10.1111/jon.12604

Source DB:  PubMed          Journal:  J Neuroimaging        ISSN: 1051-2284            Impact factor:   2.486


  4 in total

Review 1.  Optic nerve sheath diameter: present and future perspectives for neurologists and critical care physicians.

Authors:  Piergiorgio Lochner; Marek Czosnyka; Andrea Naldi; Epameinondas Lyros; Paolo Pelosi; Shrey Mathur; Klaus Fassbender; Chiara Robba
Journal:  Neurol Sci       Date:  2019-07-31       Impact factor: 3.307

Review 2.  Ultrasound Detection of Intracranial Hypertension in Brain Injuries.

Authors:  Livio Vitiello; Giulio Salerno; Maddalena De Bernardo; Olga D'Aniello; Luigi Capasso; Giuseppe Marotta; Nicola Rosa
Journal:  Front Med (Lausanne)       Date:  2022-06-30

3.  Optic nerve sheath diameter asymmetry in healthy subjects and patients with intracranial hypertension.

Authors:  Andrea Naldi; Paolo Provero; Alessandro Vercelli; Mauro Bergui; Anna Teresa Mazzeo; Roberto Cantello; Giacomo Tondo; Piergiorgio Lochner
Journal:  Neurol Sci       Date:  2019-10-05       Impact factor: 3.307

4.  Comparative study on optic disc features of premature infants and full-term newborns.

Authors:  Xiaofen Feng; Yan Nan; Jiandong Pan; Ruitao Zou; Lijun Shen; Feng Chen
Journal:  BMC Ophthalmol       Date:  2021-03-06       Impact factor: 2.209

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.