Literature DB >> 30674636

WEB embolization versus stent-assisted coiling: comparison of complication rates and angiographic outcomes.

Christoph Kabbasch1, Lukas Goertz2, Eberhard Siebert3, Moriz Herzberg4, Jan Borggrefe1, Boris Krischek2, Pantelis Stavrinou2, Franziska Dorn4, Thomas Liebig4.   

Abstract

BACKGROUND: Intrasaccular flow disruption represents a new paradigm in endovascular treatment of wide-necked bifurcation aneurysms.
OBJECTIVE: To perform a matched case-control study comparing complications and angiographic outcome using the Woven Endobridge (WEB) device and stent-assisted coiling (SAC).
METHODS: Sixty-six patients treated with the WEB at three German tertiary care centers were included and matched with 66 patients treated with SAC based on aneurysm location and unruptured/ruptured aneurysm status. Parameters were retrospectively analysed and compared between the treatment groups using inverse probability of treatment weighting (IPTW) with propensity scores.
RESULTS: Procedural complication rates were 12.1% in the WEB group and 21.2% in the SAC group, which was statistically significant after IPTW adjustment (OR=2.2, 95% CI 1.08 to 4.4, p=0.03). Favourable outcome (modified Rankin scale score ≤2) was achieved by 57/66 (86.4%) in the WEB group and 57/66 (86.4%) in the SAC group (p=1.0). At mid-term follow-up, a similar number of aneurysms achieved adequate occlusion (complete occlusion or neck remnant) in the WEB group (93.9%) and in the SAC group (93.9%, p=1.0). Re-treatment was performed in 10.6% after WEB embolization and 12.1% after SAC (p=1.0).
CONCLUSIONS: The WEB provides similar mid-term aneurysm occlusion rates to those of SAC, with no additional morbidity and potentially lower complication rates. Long-term outcome analysis will provide a definite conclusion on the use of WEB for intracranial aneurysms. © Author(s) (or their employer(s)) 2019. No commercial re-use. See rights and permissions. Published by BMJ.

Entities:  

Keywords:  aneurysm; coil; device; statistics

Mesh:

Year:  2019        PMID: 30674636     DOI: 10.1136/neurintsurg-2018-014555

Source DB:  PubMed          Journal:  J Neurointerv Surg        ISSN: 1759-8478            Impact factor:   5.836


  12 in total

1.  The RISE trial: A Randomized Trial on Intra-Saccular Endobridge devices.

Authors:  Jean Raymond; Anne-Christine Januel; Daniela Iancu; Daniel Roy; Alain Weill; Andrew Carlson; Tim E Darsaut
Journal:  Interv Neuroradiol       Date:  2019-11-05       Impact factor: 1.610

2.  Low-Profile Intra-Aneurysmal Flow Disruptor WEB 17 versus WEB Predecessor Systems for Treatment of Small Intracranial Aneurysms: Comparative Analysis of Procedural Safety and Feasibility.

Authors:  L Goertz; T Liebig; E Siebert; M Herzberg; L Pennig; M Schlamann; J Borggrefe; B Krischek; F Dorn; C Kabbasch
Journal:  AJNR Am J Neuroradiol       Date:  2019-09-05       Impact factor: 3.825

3.  Implantation of Large Diameter (5.5-6 mm) Derivo Embolization Devices for the Treatment of Cerebral Aneurysms.

Authors:  Waleed Butt; Cha-Ney Kim; Rajesh Ramaswamy; Aubrey Smith; Paul Maliakal
Journal:  Clin Neuroradiol       Date:  2021-09-08       Impact factor: 3.649

4.  Treatment of Proximal Posterior Inferior Cerebellar Artery Aneurysms by Intrasaccular Flow Disruption: A Multicenter Experience.

Authors:  L Goertz; T Liebig; E Siebert; Y Özpeynirci; L Pennig; E Celik; M Schlamann; F Dorn; C Kabbasch
Journal:  AJNR Am J Neuroradiol       Date:  2022-07-21       Impact factor: 4.966

5.  Comparison of angiographic outcomes and complication rates of WEB embolization and coiling for treatment of unruptured basilar tip aneurysms.

Authors:  Erkan Celik; Yigit Ozpeynirci; Thomas Liebig; Marc Schlamann; Franziska Dorn; Nils Lehnen; Eberhard Siebert; Lukas Goertz; Christoph Kabbasch
Journal:  Sci Rep       Date:  2022-06-28       Impact factor: 4.996

Review 6.  Comprehensive review of the recent advances in devices for endovascular treatment of complex brain aneurysms.

Authors:  Kavi Fatania; Dr Tufail Patankar
Journal:  Br J Radiol       Date:  2021-10-05       Impact factor: 3.039

7.  Volume-based sizing of the Woven EndoBridge (WEB) device: A preliminary assessment of a novel method for device size selection.

Authors:  Kevin A Shah; Timothy G White; Ina Teron; Thomas Link; Amir R Dehdashti; Jeffrey M Katz; Henry H Woo
Journal:  Interv Neuroradiol       Date:  2021-01-21       Impact factor: 1.764

8.  Intracranial aneurysm treatment with WEB and adjunctive stent: preliminary evaluation in a single-center series.

Authors:  Maher Sahnoun; Sébastien Soize; Pierre-François Manceau; Christophe Gelmini; Laurent Pierot
Journal:  J Neurointerv Surg       Date:  2021-03-30       Impact factor: 5.836

9.  Safety and Efficacy of the Woven EndoBridge Device for Treatment of Ruptured Intracranial Aneurysms: A Systematic Review and Meta-analysis.

Authors:  M A Essibayi; G Lanzino; W Brinjikji
Journal:  AJNR Am J Neuroradiol       Date:  2021-06-11       Impact factor: 4.966

10.  Ticagrelor versus Clopidogrel in the Dual Antiplatelet Regimen for Intracranial Stenting or Flow-Diverter Treatment for Unruptured Cerebral Aneurysms: A Single-Center Cohort Study.

Authors:  K Y Park; T Ozaki; A Kostynskyy; H Kortman; A Hilario; P Nicholson; R Agid; T Krings; V M Pereira
Journal:  AJNR Am J Neuroradiol       Date:  2021-07-08       Impact factor: 4.966

View more