| Literature DB >> 30652020 |
Christina A Sutherland1,1, Can Ozbal2,2, David P Nicolau1,3,1,3.
Abstract
AIM: We describe the validation of an HPLC-MS/MS method to analyze ceftolozane and tazobactam simultaneously in saline matrixes. MATERIALS &Entities:
Keywords: BAL; HPLC–MS; ceftolozane; tazobactam
Year: 2018 PMID: 30652020 PMCID: PMC6331755 DOI: 10.4155/fsoa-2018-0079
Source DB: PubMed Journal: Future Sci OA ISSN: 2056-5623
Precision and accuracy of ceftolozane in saline.
| Inter-run (n = 7) | |||
| Mean | 0.03 | 0.09 | 0.40 |
| SD | 0.002 | 0.005 | 0.009 |
| Precision | 6.69% | 6.26% | 2.19% |
| Accuracy | 97.75% | 95.77% | 99.83% |
| Intra-run (n = 10) | |||
| Mean | 0.03 | 0.08 | 0.40 |
| SD | 0.001 | 0.003 | 0.023 |
| Precision | 4.57% | 4.18% | 5.79% |
| Accuracy | 98.45% | 91.33% | 100.19% |
SD: Standard deviation.
Precision and accuracy of tazobactam in saline.
| Inter-run (n = 7) | |||
| Mean | 0.03 | 0.10 | 0.41 |
| SD | 0.001 | 0.003 | 0.009 |
| Precision | 2.66% | 2.64% | 2.26% |
| Accuracy | 103.37% | 105.68% | 102.55% |
| Intra-run (n = 10) | |||
| Mean | 0.03 | 0.09 | 0.41 |
| SD | 0.002 | 0.003 | 0.020 |
| Precision | 5.75% | 3.69% | 4.92% |
| Accuracy | 98.17% | 103.63% | 102.13% |
SD: Standard deviation.
Lower limit of quantification of 0.02 μg/ml.
Ceftolozane (top trace), tazobactam internal standard (middle trace) and tazobactam (bottom trace).