| Literature DB >> 30646622 |
Xiao-Yan Zhang1, Zhe-Yu Li2, Yu Zhang3, Xiao-Qian Zang4, Kosei Ueno5, Hiroaki Misawa6,7, Kai Sun8.
Abstract
Capacitively coupled contactless conductivity detection (C⁴D) is an improved approach to avoid the problems of labor-intensive, time-consuming and insufficient accuracy of plate count as well as the high-cost apparatus of flow cytometry (FCM) in bacterial counting. This article describes a novel electrode-integrated printed-circuit-board (PCB)-based C⁴D device, which supports the simple and safe exchange of capillaries and improves the sensitivity and repeatability of the contactless detection. Furthermore, no syringe pump is needed in the detection, it reduces the system size, and, more importantly, avoids the effect on the bacteria due to high pressure. The recovered bacteria after C⁴D detection at excitation of 25 Vpp and 60⁻120 kHz were analyzed by flow cytometry, and a survival rate higher than 96% was given. It was verified that C⁴D detection did not influence the bacterial viability. Moreover, bacteria concentrations from 10⁶ cells/mL to 10⁸ cells/mL were measured in a linear range, and relative standard deviation (RSD) is below 0.2%. In addition, the effects on bacteria and C⁴D from background solutions were discussed. In contrast to common methods used in most laboratories, this method may provide a simple solution to in situ detection of bacterial cultures.Entities:
Keywords: E. coli; bacterial concentration; capacitively coupled contactless conductivity detection (C4D); capillary; printed-circuit-board (PCB)
Year: 2019 PMID: 30646622 PMCID: PMC6356519 DOI: 10.3390/mi10010055
Source DB: PubMed Journal: Micromachines (Basel) ISSN: 2072-666X Impact factor: 2.891
Figure 1(A) Schematic illustration of capacitively coupled contactless conductivity detection (C4D) device on detecting bacterial concentration. (B) The printed-circuit-board (PCB) based electrodes. (C) Fluorescence-labeled Escherichia coli in capillary. (D) Scanning electron microscope (SEM) image of E. coli, the length of which on average is about 2 μm.
Figure 2Influence on bacteria viability due to C4D detection by analyzing the ratio of live to dead cells. The recovered bacteria were labeled with Propidium Iodide (PI) and analyzed by flow cytometry (FCM). Error bars are standard deviations of four measurements.
Figure 3The C4D response to the bacteria suspended in original phosphate-buffered saline (PBS) buffer with concentrations from 104 to 108 cells/mL. Error bars are standard deviations of four measurements.
Figure 4(A) Bacteria survival rate in 10 mM PBS solution, 0.2 mM PBS dilution, 0.1 mM PBS dilution and water after 2 h and 24 h were detected by FCM. Bacteria were dyed by PI for 10 min. Error bars are standard deviations of 3 measurements. (B) The relationship between PBS dilution and C4D intensity. Error bars are standard deviations of 4 measurements.
Figure 5The C4D response to the bacteria suspended in 0.2 mM PBS dilution and 0.1 mM PBS dilution with the different concentrations of 104–108 cells/mL. Error bars are standard deviations of four measurements.
The comparison of developed devices and this work.
| Method | Operation | Recording | Label | Accuracy | Reference |
|---|---|---|---|---|---|
| Optical density (OD) | simple | automatic monitoring | no | providing the growth trend | [ |
| Plate count | simple | manual reading | no | depending on labor | [ |
| Flow cytometry | skilled operator needed | automatic monitoring | fluorescence | good | [ |
| This work | simple | automatic monitoring | no | Better than OD in range of 106–108 cells/ml |