Literature DB >> 30641305

Validation of a novel single lead ambulatory ECG monitor - Cardiostat™ - Compared to a standard ECG Holter monitoring.

Isabelle Nault1, Philippe André2, Benoit Plourde2, François Leclerc2, Jean-François Sarrazin2, François Philippon2, Gilles O'Hara2, Franck Molin2, Christian Steinberg2, Karine Roy2, Louis Blier2, Jean Champagne2.   

Abstract

BACKGROUND: Cardiostat™ is a single lead ambulatory ECG monitor. Recording is made through 2 electrodes positioned in a lead 1-like configuration. We first validated its accuracy for atrial fibrillation detection compared to a 12-lead ECG. In the second phase of the study, arrhythmia detection accuracy was compared between Cardiostat™ ambulatory ECG and a standard 24 h Holter ECG monitoring. METHOD/
RESULTS: Phase one of the study included patients undergoing cardioversion for atrial fibrillation (AF) or atrial flutter. Cardiostat™ tracings were compared with standard 12-lead ECG. In the second phase, patients undergoing 24 h ambulatory Holter ECG monitoring for control or suspicion of atrial fibrillation (AF) were included. Simultaneous Holter monitoring and Cardiostat™ ECG recordings were performed. Tracings were analysed and compared. Two hundred twelve monitoring were compared. AF was diagnosed in 73 patients. Agreement between Cardiostat™ ECG and standard Holter monitoring was 99% for AF detection with kappa = 0.99. Kappa correlation for atrial flutter detection was only moderate at 0.51. AF burden was similar in both recordings. Noise hindered analysis in a greater proportion with Cardiostat™ compared to Holter ambulatory ECG (8.5 vs 3.8%).
CONCLUSION: Cardiostat™ ambulatory ECG device showed excellent correlation with the standard Holter ECG monitoring for AF detection. Holter monitoring was however superior to discriminate premature atrial and ventricular beats and to qualify the morphology of PVCs since it has more vectors for analysis. Added value of Cardiostat™ includes longer monitoring duration, less cumbersome installation and water resistance.
Copyright © 2019 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Ambulatory ECG monitoring; Atrial fibrillation; Extended monitoring

Year:  2018        PMID: 30641305     DOI: 10.1016/j.jelectrocard.2018.12.011

Source DB:  PubMed          Journal:  J Electrocardiol        ISSN: 0022-0736            Impact factor:   1.438


  6 in total

1.  Methodology for the nocturnal cardiac arrhythmia ancillary study of the ADVENT-HF trial in patients with heart failure with reduced ejection fraction and sleep-disordered breathing.

Authors:  Christian M Horvath; Christoph Fisser; T Douglas Bradley; John S Floras; Samuel Sossalla; Gianfranco Parati; Florian Zeman; Paolo Castiglioni; Andrea Faini; Fiona Rankin; Michael Arzt
Journal:  Int J Cardiol Heart Vasc       Date:  2022-05-21

2.  Comparison Between the 24-hour Holter Test and 72-hour Single-Lead Electrocardiogram Monitoring With an Adhesive Patch-Type Device for Atrial Fibrillation Detection: Prospective Cohort Study.

Authors:  Soonil Kwon; So-Ryoung Lee; Eue-Keun Choi; Hyo-Jeong Ahn; Hee-Seok Song; Young-Shin Lee; Seil Oh; Gregory Y H Lip
Journal:  J Med Internet Res       Date:  2022-05-09       Impact factor: 7.076

3.  Validation of Adhesive Single-Lead ECG Device Compared with Holter Monitoring among Non-Atrial Fibrillation Patients.

Authors:  Soonil Kwon; So-Ryoung Lee; Eue-Keun Choi; Hyo-Jeong Ahn; Hee-Seok Song; Young-Shin Lee; Seil Oh
Journal:  Sensors (Basel)       Date:  2021-04-30       Impact factor: 3.576

Review 4.  Remote Cardiac Rhythm Monitoring in the Era of Smart Wearables: Present Assets and Future Perspectives.

Authors:  Anastasia Xintarakou; Vasileios Sousonis; Dimitrios Asvestas; Panos E Vardas; Stylianos Tzeis
Journal:  Front Cardiovasc Med       Date:  2022-03-01

5.  Diagnostic Efficacy of a Single-Lead Ambulatory 14-Day ECG Monitor in Symptomatic Children.

Authors:  Jean-Marc Côté; Philippe Chetaille; Sylvia Abadir; Louise Gosselin; David Simonyan; Frédéric Dallaire
Journal:  CJC Open       Date:  2021-06-24

6.  Effect of Mobile Internet on Attitude and Self-Efficacy of Patients with Coronary Heart Disease Diagnosed by 12-Lead Holter ECG.

Authors:  Haitao Sun; Jing Li; Yue Wang; Xiaoke Ma
Journal:  J Healthc Eng       Date:  2022-01-07       Impact factor: 2.682

  6 in total