Literature DB >> 30611149

A comprehensive biomechanical analysis of sacral alar iliac fixation: an in vitro human cadaveric model.

Bryan W Cunningham1, Paul D Sponseller2, Ashley A Murgatroyd1, Jun Kikkawa3, P Justin Tortolani1.   

Abstract

In BriefIn this in vitro investigation, we compared the multidirectional flexibility properties of sacral alar iliac fixation with conventional methods of sacral and sacroiliac fixation using nondestructive and destructive investigative methods. The study demonstrated that S1-2 sacral fixation alone significantly increases sacroiliac motion under all loading modalities, while sacral alar iliac fixation reduced motion in axial rotation at the sacroiliac joint and offers potential advantages of a lower instrumentation profile and ease of assembly compared to conventional sacroiliac screw instrumentation.

Entities:  

Keywords:  ROM = range of motion; biomechanical properties; iliac screws; sacral alar iliac fixation; sacroiliac junction

Mesh:

Year:  2019        PMID: 30611149     DOI: 10.3171/2018.8.SPINE18328

Source DB:  PubMed          Journal:  J Neurosurg Spine        ISSN: 1547-5646


  4 in total

1.  Biomechanics of sacropelvic fixation: a comprehensive finite element comparison of three techniques.

Authors:  Fabio Galbusera; Gloria Casaroli; Ruchi Chande; Derek Lindsey; Tomaso Villa; Scott Yerby; Ali Mesiwala; Matteo Panico; Enrico Gallazzi; Marco Brayda-Bruno
Journal:  Eur Spine J       Date:  2019-11-26       Impact factor: 3.134

2.  Innovative sacropelvic fixation using iliac screws and triangular titanium implants.

Authors:  Matteo Panico; Ruchi D Chande; Derek P Lindsey; Ali Mesiwala; Tomaso Maria Tobia Villa; Scott A Yerby; Enrico Gallazzi; Marco Brayda-Bruno; Fabio Galbusera
Journal:  Eur Spine J       Date:  2021-09-25       Impact factor: 3.134

3.  Novel Technique for Sacral-Alar-Iliac Screw Placement Using Three-Dimensional Patient-Specific Template Guide.

Authors:  Keitaro Matsukawa; Yuichiro Abe; Ralph Jasper Mobbs
Journal:  Spine Surg Relat Res       Date:  2021-03-10

4.  Comparative Analysis of Optoelectronic Accuracy in the Laboratory Setting Versus Clinical Operative Environment: A Systematic Review.

Authors:  Bryan W Cunningham; Daina M Brooks
Journal:  Global Spine J       Date:  2022-04
  4 in total