Nishanee Rampersad1, Rekha Hansraj1.
Abstract
BACKGROUND: Accurate and repeatable measurements of the retinal nerve fibre layer (RNFL) thickness are important in the diagnosis and management of glaucoma and other disorders.Entities:
Keywords: Retinal nerve fibre layer thickness; iVue-100 OCT; optical coherence tomography; repeatability; reproducibility
Mesh:
Year: 2018 PMID: 30602957 PMCID: PMC6306963 DOI: 10.4314/ahs.v18i2.15
Source DB: PubMed Journal: Afr Health Sci ISSN: 1680-6905 Impact factor: 0.927
Figure 1RNFL thickness display of the optic nerve head scan
Intraclass correlation coefficient with 95% confidence intervals and CoV for each observer, for RNFL thickness measured for the different quadrants of the right eye
| OBSERVER 1 | OBSERVER 2 | |||
| ICC (95% CI) | CoV | ICC (95% CI) | CoV | |
| 0.976 (0.962–0.986) | 0.098 | 0.953 (0.925–0.972) | 0.091 | |
| 0.968 (0.948–0.981) | 0.137 | 0.919 (0.870–0.952) | 0.125 | |
| 0.972 (0.956–0.983) | 0.111 | 0.872 (0.795–0.923) | 0.108 | |
| 0.941 (0.906–0.965) | 0.134 | 0.829 (0.727–0.898) | 0.140 | |
| 0.958 (0.932–0.975) | 0.132 | 0.918 (0.868–0.951) | 0.132 |
The mean differences, and their standard deviations, of RNFL thickness (µm) between two observers, Bland and Altman upper and lower LoA, t-values and p-values from the Paired t-tests and the ICC
| Mean diff | SD(µm) | Upper | Lower | ICC | |||
| 1.047 | 2.959 | 6.847 | −4.753 | 2.501 | 0.016 | 0.976 | |
| 2.667 | 6.998 | 16.383 | −11.049 | 2.640 | 0.011 | 0.954 | |
| 2.507 | 8.121 | 18.424 | −13.410 | 2.183 | 0.034 | 0.912 | |
| −0.253 | 6.631 | 12.744 | −13.250 | −0.270 | 0.788 | 0.906 | |
| −0.403 | 4.993 | 9.383 | −10.189 | −0.559 | 0.579 | 0.932 |
Figure 2Bland-Altman plot comparing RNFL thickness measurements of the right eyes taken by observer one and observer two. The solid line represents the mean difference and the dashed lines represent the upper and lower LoA. Three points were outside the 95% LoA region.
The mean differences and their standard deviations, of RNFL thickness (µm) for observer one taken over two sessions, Bland and Altman upper and lower LoA, and t-values and p-values from Paired t-tests and ICC
| Mean diff | SD(µm) | Upper | Lower | ICC | |||
| 1.465 | 2.625 | 6.610 | −3.680 | 3.206 | 0.003 | 0.979 | |
| 4.061 | 8.841 | 21.389 | −13.267 | 2.639 | 0.013 | 0.914 | |
| 1.455 | 5.321 | 11.884 | −8.974 | 1.570 | 0.126 | 0.969 | |
| 0.222 | 4.421 | 8.887 | −8.443 | .289 | 0.775 | 0.942 | |
| 0.162 | 3.654 | 7.324 | −7.000 | .254 | 0.801 | 0.965 |
Figure 3Bland-Altman plot comparing average RNFL measurements of the right eyes taken by observer one in two separate sessions. The solid line represents the mean difference and the dashed lines represent the upper and lower LoA.