Clutch Cutter knife efficacy in endoscopic submucosal dissection for early gastric neoplasms.

Yasuyo Hayashi1, Mitsuru Esaki1, Sho Suzuki2, Eikichi Ihara1, Azusa Yokoyama3, Seiichiro Sakisaka3, Taizo Hosokawa3, Yoshimasa Tanaka3, Takahiro Mizutani3, Shinichi Tsuruta4, Aya Iwao2, Shun Yamakawa2, Akira Irie2, Yosuke Minoda1, Yoshitaka Hata1, Haruei Ogino1, Hirotada Akiho3, Yoshihiro Ogawa1.   

Abstract

AIM: To compare the outcomes of endoscopic submucosal dissection (ESD) for gastric neoplasms using Clutch Cutter (ESD-C) or other knives (ESD-O).
METHODS: This was a single-center retrospective study. Gastric neoplasms treated by ESD between April 2016 and October 2017 at Kitakyushu Municipal Medical Center were reviewed. Multivariate analyses and propensity score matching were used to reduce biases. Covariates included factors that might affect outcomes of ESD, including age, sex, underlying disease, anti-thrombotic drugs use, tumor location, tumor position, tumor size, tumor depth, tumor morphology, tumor histology, ulcer (scar), and operator skill. The treatment outcomes were compared among two groups. The primary outcome was ESD procedure time. Secondary outcomes were en bloc, complete, and curative resection rates, and adverse events rates including perforation and delayed bleeding.
RESULTS: A total of 155 patients were included in this study; 44 pairs were created by propensity score matching. Background characteristics were quite similar among two groups after matching. Procedure time was significantly shorter for ESD-C (median; 49 min) than for ESD-O (median; 88.5 min) (P < 0.01). However, there was no significant difference in treatment outcomes between ESD-C and ESD-O including en bloc resection rate (100% in both groups), complete resection rate (100% in both groups), curative resection rate (86.4% vs 88.6%, P = 0.730), delayed bleeding (2.3% vs 6.8%, P = 0.62) and perforation (0% in both groups).
CONCLUSION: ESD-C achieved shorter procedure time without an increase in complication risk. Therefore, ESD-C could become an effective ESD option for gastric neoplasms.

Entities:  

Keywords:  Clutch Cutter; Endoscopic submucosal dissection; Gastric neoplasm; Knife; Propensity score

Year:  2018        PMID: 30595802      PMCID: PMC6304306          DOI: 10.4251/wjgo.v10.i12.487

Source DB:  PubMed          Journal:  World J Gastrointest Oncol


Core tip: Propensity score matching was performed to compare the outcomes of endoscopic submucosal dissection (ESD) for gastric neoplasms using Clutch Cutter or other knives in this single-center retrospective study. Forty-four pairs were matched in this study. ESD using Clutch Cutter achieved shorter procedure time without an increase in complication risk (median procedure time; 49 min vs 88.5 min, P < 0.01). Therefore, ESD using Clutch Cutter could become an effective ESD option for gastric neoplasms.

INTRODUCTION

Endoscopic submucosal dissection (ESD) is the standard treatment for gastrointestinal tract tumors including gastric neoplasms, achieving a higher rate of en bloc resection and low rates of local recurrence even for large and ulcerated lesions, as compared with endoscopic mucosal resection (EMR)[1]. However, more advanced technical skills and greater experience are needed in ESD because of the longer procedure time and high risk of complications including bleeding and perforation[1]. Although various types of endo-knife including needle-type knife and insulated-tip knife were invented and used in ESD, this remains a challenging procedure and there is no consensus on the best knife to be used[2-5]. Clutch Cutter (DP2618DT, Fujifilm Medical, Tokyo, Japan; Figure 1) was invented as a scissor-type device for ESD, which allows for grasping of the targeted tissue and its subsequent cut with an electrosurgical unit[6]. This procedure is similar to the technique of a standard bite biopsy, which is a common procedure during routine endoscopy. Furthermore, Clutch Cutter allows re-grasping of the tissue anytime during ESD, which may prevent miscutting and perforation. Therefore, Clutch Cutter may contribute to easier and safer ESD than other endo-knives. ESD with Clutch Cutter (ESD-C) may then become an option of endo-knife for ESD. Favorable outcomes of ESD-C have been reported, including in a large single-center study with single arm trial[7-9]. However, few reports exist showing the comparison between scissors-type and non-scissors-type knives in the technical outcomes of ESD, which have been limited to non-experts in inclusion criteria[10,11]. The advantage of scissor-type knife in ESD is controversial at present because comparative studies are still lacking.
Figure 1

Clutch Cutter is a scissor-type device for endoscopic submucosal dissection.

Clutch Cutter is a scissor-type device for endoscopic submucosal dissection. We retrospectively compared the technical outcomes of ESD-C for gastric neoplasms with those of ESD with other knives (ESD-O) by using propensity score matching analysis, which compensated for differences in extraneous factors including baseline characteristics[2]. We hypothesize that the outcomes using ESD-C will be superior to those of ESD-O.

MATERIALS AND METHODS

Study design and ethical approval

This was designed as a retrospective, observational cohort study, which was conducted based on the ESD databases at a single-center, Kitakyushu Municipal Medical Center (Fukuoka, Japan). These cases represented a consecutive and unselected cohort. The protocol of this study was developed in accordance with the Declaration of Helsinki and approved by the Institutional Review Board of Kitakyushu Municipal Medical Center on November 2017 (No. 201711050).

Patients

We enrolled 191 consecutive patients with gastric neoplasms treated by ESD between April 2016 and October 2017 at the Kitakyushu Municipal Medical Center. Three patients were excluded from analysis for previously having undergone gastric surgery. Furthermore, 33 patients were excluded because two or more lesions were simultaneously resected. Finally, 155 patients were analyzed in this study. We classified the patients into two groups: one group included patients treated by ESD-C and the other group included patients treated by ESD-O. Either IT Knife2 (KD-611L, Olympus, Tokyo, Japan) or Splash M-Knife (DN-D2718A; HOYA Corp., Pentax, Tokyo, Japan) was mainly used in the patients enrolled between April 2016 and March 2017, while Clutch Cutter was mainly used in the patients enrolled between April 2017 and October 2017. The flow chart of the patients enrolled in the present study is shown in Figure 2.
Figure 2

Flowchart of patients and lesions enrolled in this study. ESD: Endoscopic submucosal dissection; ESD-C: Endoscopic submucosal dissection using Clutch Cutter; ESD-O: Endoscopic submucosal dissection using other knives.

Flowchart of patients and lesions enrolled in this study. ESD: Endoscopic submucosal dissection; ESD-C: Endoscopic submucosal dissection using Clutch Cutter; ESD-O: Endoscopic submucosal dissection using other knives.

ESD procedure

All patients were admitted at Kitakyushu Municipal Medical Center. All ESD procedures were carried out using a GIF-Q260J (Olympus, Tokyo, Japan) with a CO2 insufflation system. VIO 300D (ERBE Elektromedizin, GmbH, Tübingen, Germany) was used as the electrical power source. The ESD procedure was described in detail in previous reports[7,12,13]. In brief, marking dots were made 2 mm outside the lesion. A mixture of 4% hyaluronic acid and normal saline with a small amount of indigo carmine and epinephrine (0.001 mg/mL) was injected into the submucosa. After lifting the lesion, mucosal incision was conducted circumferentially using cutting and coagulation (Figure 3A). Once the circumferential mucosal incision was completed, submucosal dissection of the lesion was performed using cutting and coagulation (Figure 3B). Injection was added during dissection when needed. Prophylactic coagulation for visible vessels or hemostasis for active bleeding was conducted using endo-knives or hemostatic forceps (Figure 3C). When using Clutch Cutter, cutting was conducted by Endo Cut Mode (effect 1, duration 4, interval 1), while coagulation was conducted by Soft Coagulation Mode (80-100 W, effect; 5-6 in VIO300D) or Forced Coagulation Mode (30 W, effect 2). In this study, operators with an experience of performing at least 50 ESD procedures were defined as experts, while those who had performed less than 50 ESD procedures were defined as trainees. As a result, 4 operators were defined as experts and 5 as trainees in this study. All experts were familiar with using each device since they had used each device at least 10 times before this study.
Figure 3

Procedures of endoscopic submucosal dissection. A: Mucosal incision using Clutch Cutter; B: Submucosal dissection using Clutch Cutter; C: Prophylactic coagulation for visible vessels using Clutch Cutter.

Procedures of endoscopic submucosal dissection. A: Mucosal incision using Clutch Cutter; B: Submucosal dissection using Clutch Cutter; C: Prophylactic coagulation for visible vessels using Clutch Cutter.

Histology evaluation

ESD specimens were immediately stretched and fixed in 10% buffered formalin. The specimens were serially sectioned perpendicularly at 2-3 mm intervals. Then, histological type, depth of invasion, tumor size, lymphatic/vascular invasion, and resection margin were assessed. The pathological curability of the specimens was evaluated based on the Japanese Gastric Cancer Classification[14].

Outcome

The primary outcome of this study was the procedure time during ESD, which was defined as the time from the start of marking to the completion of dissection. En bloc resection rate, complete resection rate, curative resection rate, and the rate of complications (delayed bleeding and perforation) were evaluated as secondary outcomes. En bloc resection was defined as resection in one piece. Complete resection was defined as en bloc resection with the lateral and vertical resection margins free of neoplasm. Curative resection was evaluated according to the guideline[15]. Delayed bleeding was defined as clinical evidence of bleeding after ESD, requiring endoscopic hemostasis or blood transfusion. Perforation was diagnosed if mesenteric fat or the intra-abdominal space was observed during ESD procedure or free air was detected on chest and abdominal radiographs or computed tomography scans after ESD. All patients were given a proton pump inhibitor or potassium competitive acid blocker for a minimum of 4 wk.

Statistics

Background characteristics were not equal among two groups. Previous studies have reported some factors associated with the difficulty or complication of the ESD procedure, which may affect outcomes of this study[16-22]. Therefore, we adopted propensity score matching analysis to reduce bias. Logistic regression of the following factors with ESD device (Clutch Cutter vs other endo-knives) and calculation of propensity score were conducted: age (≥ 75 vs < 75 years old), sex (male vs female), underlying disease (presence vs none), anti-thrombotic drugs use (continuation vs not receiving or discontinuation), tumor location (upper third of the stomach vs middle or lower third), tumor position (lessor curvature of the stomach vs others), tumor size (> 20 mm vs ≤ 20 mm), tumor depth (mucosa vs submucosa), tumor morphology (flat or depressed vs others), tumor histology (differentiated type vs undifferentiated type), ulcer (scar) (presence vs absence), and operator skill (expert vs trainee). Underlying disease included cardiomyopathy, liver cirrhosis, and chronic kidney disease. Nearest neighbor matching in a 1:1 ratio from the ESD-C and ESD-O groups was made in calipers (0.12) with a width equal to 0.25 of the standard deviation of the logit of the propensity score. Baseline characteristics and outcomes were analyzed using Fisher’s exact test for categorial data, the Mann–Whitney U test for continuous data with non-normal distributions, and a t test for continuous data with normal distribution. P < 0.05 was considered statistically significant for all tests. All statistical analyses were performed using JMP Pro13.0 software (SAS Institute, Cary, NC, United States).

RESULTS

Propensity score matching

The area under the receiver operating characteristic curve, i.e., C statistic, was estimated to be 0.681, which indicated good predictive power. Propensity score matching created 44 pairs in this study. We compared two groups by using the absolute standardized differences (ASD) before and after matching to assess the propensity score balance. After matching, all ASDs ranged within 1.96√2/n, indicating that the characteristics were well-balanced[23].

Characteristics before and after propensity score matching

The background characteristics of 155 patients enrolled in this study are shown in Table 1. Patients in the ESD-C group had a significantly higher rate of undifferentiated adenocarcinoma than those in the ESD-O group (10.0% vs 0.95%; P = 0.014). The median tumor size of patients in ESD-C group was significantly smaller than that of patients in the ESD-O group (13.5 mm vs 18.0 mm; P = 0.027).
Table 1

Characteristics of enrolled patients before propensity score matching

ESD-C n = 50ESD-O n = 105P valueASD
Age, yr
Mean ± SD73.1 ± 8.5572.75 ± 8.040.80620.0422
Median (range)74.0 (46–91)73.0 (52–91)0.5803
Sex (n)
Male39710.25610.2350
Female1134
Underlying disease, positive, n (%)18 (36.0)29 (27.6)0.35010.1810
Anti-thrombotic drugs (n)
None or discontinuation481020.65810.0623
Continuation23
Tumor location (n)
Upper third9170.82010.0481
Middle or lower third4188
Tumor position (n)
Lessor21570.73110.2670
Others2948
Morphology (n)
Flat or depressed29630.86210.0407
Others2142
Histology (n)
Undifferentiated510.014140.4060
Others45104
Tumor size (mm)
Mean ± SD16.89 ± 12.6520.90 ± 13.300.07620.3090
Median (range)13.5 (3–67)18.0 (3–82)0.02734
Tumor depth (n)
Mucosa44911.00010.0401
Submucosa614
Ulceration positive, n (%)7 (14.0)22 (21.0)0.38110.1840
Operator skill
Experts34730.85410.0329
Trainees1632

P value was calculated using Fisher’s exact test;

P value was calculated using a t test;

P value was calculated using the Mann–Whitney U test;

Significant value. ESD-C: Endoscopic submucosal dissection with Clutch Cutter; ESD-O: Endoscopic submucosal dissection with another end-knife; SD: Standard deviation; ASD: Absolute standardized difference.

Characteristics of enrolled patients before propensity score matching P value was calculated using Fisher’s exact test; P value was calculated using a t test; P value was calculated using the Mann–Whitney U test; Significant value. ESD-C: Endoscopic submucosal dissection with Clutch Cutter; ESD-O: Endoscopic submucosal dissection with another end-knife; SD: Standard deviation; ASD: Absolute standardized difference. Matching factors between two groups after propensity score matching are shown in Table 2. No significant differences were found in any matching factors.
Table 2

Matching factors between two groups after propensity score matching

ESD-C n = 44ESD-O n = 44P valueASD
Variable matching between groups
Age, yr; mean ± SD73.16 ± 8.5971.11 ± 8.810.27320.2360
Sex: Male/female8/366/380.77210.1250
Underlying disease: No/yes28/1629/15110.0476
Anti-thrombotic drugs: No/yes2/423/41110.0983
Tumor location: Upper third/others7/372/420.25010.3810
Tumor position: Lessor/others19/2521/230.83110.0914
Morphology: Flat or depressed/others25/1928/160.66310.1400
Histology: Undifferentiated/others0/440/44-0
Tumor size, mm: mean ± SD16.89 ± 12.6520.90 ± 13.300.07620.3090
Tumor depth: Mucosa/submucosaJun-38Jun-38110
Ulceration, positive6 (13.6%)4 (9.1%)0.73910.1440
Operator skill: Expert/trainee14/3014/30110

P value was calculated using Fisher’s exact test;

P value was calculated using a t test for continuous data. ESD-C: Endoscopic submucosal dissection with Clutch Cutter; ESD-O: Endoscopic submucosal dissection with another end-knife; SD: Standard deviation; ASD: Absolute standardized difference.

Matching factors between two groups after propensity score matching P value was calculated using Fisher’s exact test; P value was calculated using a t test for continuous data. ESD-C: Endoscopic submucosal dissection with Clutch Cutter; ESD-O: Endoscopic submucosal dissection with another end-knife; SD: Standard deviation; ASD: Absolute standardized difference.

Technical outcomes after propensity score matching

Treatment outcomes after matching are shown in Table 3. A significantly shorter procedure time was observed for ESD-C than for ESD-O in the adjusted comparison (49.0 min vs 88.5 min; P < 0.001). En bloc resection rates and complete resection rates were 100% in both groups. All ESDs were completed without perforation. Curative resection rates were similar between the two groups. The delayed bleeding rates of ESD-C tended to be lower than those of ESD-O, but these rates did not reach statistical significance (2.3% vs 6.8%, P = 0.62).
Table 3

Treatment outcomes between two groups after propensity score matching n (%)

ESD-C n = 44ESD-O n = 44P value
Procedure time, min< 0.00123
Mean ± SD63.1 ± 41.9098.41 ± 51.77
Median (range)49 (9–190)88.5 (26–290)
En bloc resection44 (100)44 (100)-
Complete resection44 (100)44 (100)-
Curative resection38 (86.4)39 (88.6)0.7301
Perforation0 (0)0 (0)-
Delayed bleeding1 (2.3)3 (6.8)0.6201

P value was calculated using Fisher’s exact test;

P value was calculated using the Mann–Whitney U test;

Significant value. ESD-C: Endoscopic submucosal dissection with Clutch Cutter; ESD-O: Endoscopic submucosal dissection with another end-knife; SD: Standard deviation.

Treatment outcomes between two groups after propensity score matching n (%) P value was calculated using Fisher’s exact test; P value was calculated using the Mann–Whitney U test; Significant value. ESD-C: Endoscopic submucosal dissection with Clutch Cutter; ESD-O: Endoscopic submucosal dissection with another end-knife; SD: Standard deviation.

DISCUSSION

The present study is the first to show that the technical outcomes of ESD-C are superior to those of ESD-O for the endoscopic treatment of gastric neoplasms regardless of technical expertise, as shown by the propensity score matching analysis. Currently, a wide variety of ESD devices is available. These devices are roughly classified into two types: scissor-type knives or non-scissor-type knives. The scissor-type knives commonly used in ESD include Clutch Cutter, SB knife, and SB knife Jr, while non-scissor-type knives mainly include IT Knife2, Dual knife, Flush knife, and Splash M-Knife. However, it is yet to be determined which type of knife is superior, scissor-type or non-scissor-type. It has been reported that the scissor-type knives reduced the technical difficulty of gastrointestinal ESD for unexperienced as well as expert endoscopists[24,25]. Rescue usage of the SB Jr knife has been reported to increase the self-completion rate of ESD of colorectal neoplasms using the Flush knife (63% in the SB Jr knife group vs 39% in the Flush knife only group; P = 0.03), without increasing the procedure time (59 min vs 51 min; P = 0.14)[11]. In other studies, however, ESD-C was reported to be a time-consuming procedure compared with ESD with non-scissor-type knives, especially when performed by unexperienced endoscopists[26,27]. Therefore, we carried out this study using a propensity score matching analysis to determine which was superior, ESD-C or ESD-O. We found that ESD-C achieved significantly shorter procedure time than ESD-O, indicating that ESD-C is a time-saving rather than a time-consuming procedure. Clutch Cutter might reduce the technical difficulties in gastric ESD similarly to those in colorectal ESD, which might have contributed to the reduction in procedure time. The scissors-type knives were invented several years after the invention of non-scissors-type knives[4-6]. In general, ESD experts tended to use non-scissors-type knives rather than scissors-type knives. In previous studies on the usefulness of ESD-C, the ESD-C procedures were conducted mainly by trainees rather than experts. In this study, however, ESD procedures were performed by 4 experts and 5 trainees. After propensity score matching, up to 68.2% of ESD procedures were conducted by experts, which might explain the discrepancy in the outcomes between the present and previous studies. ESD with scissor-type knives is being widely used not only by trainees but also by experts. Both delayed bleeding and perforation are major complications of ESD. In the present study, delayed bleeding occurred in only 1 case with ESD-C while in 3 cases with ESD-O after propensity score matching, although the results were not statistically significant. Moreover, the delayed bleeding rate was 2.0% in ESD-C before propensity score matching (data not shown), which was lower than previously reported delayed bleeding rates[15]. Although it was unknown why delayed bleeding occurs, during ESD-C, hemostasis was conducted by grasping and coagulation, similar to the procedure by hemostatic forceps, which may contribute to the reduction in the delayed bleeding rate. On the other hand, no perforation occurred in ESD-C before propensity score matching in this study. By contrast, perforation occurred in one case of ESD-O before matching (data not shown), although this case was excluded after propensity score matching. Although the number of patients undergoing ESD-C was smaller than that of patients undergoing ESD-O, ESD-C might be a safer procedure than ESD-O. In ESD-C, tissue grasping and lifting were conducted before coagulation or cutting, which could reduce heat conduction to the muscular layer, contributing to the decreased risk of perforation. In terms of safety, it was reported that ESD-C could be preferred over ESD-O for elderly patients with some comorbidities[28]. In the present study, over 90% (46/50 before matching, 41/44 after matching) of patients were aged 65 years or older; furthermore, over 40% (22/50 before matching, 20/44 after matching) of patients were aged 75 years or older. No patient experienced worsening of general condition or developed any severe complications. However, further studies are required to clarify whether ESD-C is safer than ESD-O. In addition, ESD-C has been recently used not only for ESD but also for other endoscopic procedures such as endoscopic treatment of Zenker’s diverticulum and endoscopic necrosectomy for pancreatic necrosis[29-31]. In the future, Clutch Cutter could be widely applied in additional endoscopic procedures. This study had several limitations. First, this was a single-center retrospective study. Therefore, the sample size was relatively small. There might be a selection bias because lesions in the ESD-C group were significantly smaller than those in the ESD-O group and had significantly higher rate of undifferentiation in histology evaluation. Second, only 9 endoscopists conducted ESD. Therefore, a multicenter trial should be carried out to validate this outcome. Third, in some ESD procedures conducted by trainees, experts occasionally assisted in the procedure, which might affect the outcomes of this study. Fourth, we grouped other devices together, including needle-type and insulated-tip knives, for comparison with Clutch Cutter. Future studies are needed to compare each knife individually with Clutch Cutter. Fifth, there was a possibility of an institutional learning curve. We cannot compensate for this bias because the Clutch Cutter was used mainly in the latter phase of this study; other devices were used mainly in former phase of this study, which may also affect outcomes. In conclusion, ESD-C achieved shorter procedure time than ESD-O without an increase in complication rates. Therefore, ESD-C could become one of the best endoscopic procedure options in ESD for gastric neoplasms.

ARTICLE HIGHLIGHTS

Research background

Endoscopic submucosal dissection (ESD) is the standard treatment for early gastric neoplasms with negligible lymph node metastasis. However, it is a complex and difficult procedure. Many types of endo-knives have been invented and developed to improve the ESD procedure.

Research motivation

The Clutch Cutter is a novel scissor-type endo-knife, which may contribute to facilitating the ESD procedure. However, few studies have compared the technical outcomes of each knife.

Research objectives

The aim of this study was to compare the technical outcomes between ESD with the Clutch Cutter and ESD with other devices.

Research methods

Patients with early gastric neoplasms treated by ESD at Kitakyushu Municipal Medical Center between April 2016 and October 2017 were reviewed. ESD was performed using the Clutch Cutter (ESD-C group) or other devices (ESD-O group). Propensity score matching analysis was conducted to compensate for confounding differences between the two groups that may affect the outcomes. After matching, the technical outcomes of ESD were compared among the two groups.

Research results

A total of 155 patients were included and 44 pairs were matched. ESD with the Clutch Cutter achieved a significantly shorter procedure time (median, 49 min vs 88.5 min, P < 0.001). The other technical outcomes and complication rates were similar among the two groups.

Research conclusions

The Clutch Cutter contributed to shortening the ESD’s procedure time. ESD with the Clutch Cutter could be an effective option in ESD with endo-knives for early gastric neoplasms.

Research perspectives

This was a single-center, retrospective study with a relatively small number of ESD cases. Therefore, further large-scale, randomized, prospective studies are needed.

ACKNOWLEDGMENT

We thank Shuichi Itonaga and Kota Bussaka (Kitakyushu Municipal Medical Center) for their assistance in data collection.
  3 in total

1.  Reduction in the procedure time of hybrid endoscopic submucosal dissection for early gastric neoplasms: a multi-center retrospective propensity score-matched analysis.

Authors:  Mitsuru Esaki; Sho Suzuki; Toshiki Horii; Ryoji Ichijima; Shun Yamakawa; Hitoshi Shibuya; Chika Kusano; Hisatomo Ikehara; Takuji Gotoda
Journal:  Therap Adv Gastroenterol       Date:  2020-08-03       Impact factor: 4.409

2.  Assistant skill in gastric endoscopic submucosal dissection using a clutch cutter.

Authors:  Mitsuru Esaki; Toshiki Horii; Ryoji Ichijima; Masafumi Wada; Seiichiro Sakisaka; Shuichi Abe; Naru Tomoeda; Yusuke Kitagawa; Kei Nishioka; Yosuke Minoda; Shinichi Tsuruta; Sho Suzuki; Hirotada Akiho; Eikichi Ihara; Yoshihiro Ogawa; Takuji Gotoda
Journal:  World J Gastrointest Surg       Date:  2021-02-27

3.  Comparisons of outcomes between ProKnife injection endoscopic submucosal dissection and conventional endoscopic submucosal dissection for large gastric lesions in ex vivo porcine model study: A randomized controlled trial.

Authors:  Mitsuru Esaki; Eikichi Ihara; Misato Esaki; Kei Nishioka; Yusuke Kimura; Yoshitaka Hata; Hirotaka Tsuru; Masafumi Wada; Yosuke Minoda; Xiaopeng Bai; Yoshihisa Shoguchi; Takayuki Nasu; Shuzaburo Nagatomo; Kazumasa Muta; Haruei Ogino; Yoshihiro Ogawa
Journal:  DEN open       Date:  2022-01-26
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.