Francesca Bonelli1, Tommaso Vezzosi1, Mireille Meylan2, Irene Nocera1, Vincenzo Ferrulli1, Carlotta Buralli1, Valentina Meucci1, Rosalba Tognetti1.
Abstract
BACKGROUND: There is good diagnostic accuracy of smartphone-based ECG in the evaluation of heart rate (HR), heart rhythm, and ECG values for humans, horses, dogs, and cats.Entities:
Keywords: ECG; cow; iPhone; mobile; validation
Mesh:
Year: 2018 PMID: 30592089 PMCID: PMC6430901 DOI: 10.1111/jvim.15396
Source DB: PubMed Journal: J Vet Intern Med ISSN: 0891-6640 Impact factor: 3.333
Figure 1Dorsoventral orientation with a 30° cranial inclination of the smartphone, with the case maintained in touch with the skin and the smartphone screen pointed toward the investigator for a real‐time visualization of the ECG trace. The smartphone was held with the camera located ventrally while the “start button” was located dorsally
Figure 2Standard ECG (A) and smartphone ECG (B) tracings in a cow. Paper speed = 50 mm/s (A and B); gain = 10 mm/mV (A) and 20 mm/mV (B)
Median and range (minimum and maximum) of the values obtained with the standard ECG and smartphone ECG methods
| Parameter | Standard ECG | Smartphone ECG |
|---|---|---|
| Median (range) | Median (range) | |
| HR (bpm) | 80 (60‐110) | 80 (60–110) |
| HRApp (bpm) | … | 78 (62‐265) |
| P wave (ms) | 70 (60‐120) | 75 (50‐90) |
| PR interval (ms) | 220 (160‐260) | 205 (150‐250) |
| QRS complex (ms) | 70 (50‐130) | 60 (40‐120) |
| QT interval (ms) | 380 (280‐440) | 390 (270‐460) |
Abbreviations: bpm, beats per minute; HR, heart rate, HRApp, heart rate automatically calculated by the smartphone ECG application.
Figure 3Limits of agreement (Bland‐Altman) plots showing differences between heart rate (HR) values manually measured on standard ECG (HRst) and (A) smartphone ECG tracings (HRsm) and (B) values produced by the smartphone application (HRapp); (C) differences between duration of P wave measured on standard ECG (ECGst) and smartphone ECG tracings (ECGsm); (D) differences between PR interval measured on ECGst and ECGsm; (E) differences between QRS complex measured on ECGst and ECGsm; and (F) differences between QT interval measured on ECGst and ECGsm