Literature DB >> 30546640

Traumatic pseudoaneurysm resulting from ruptured outflow graft due to disconnection of HeartMate II LVAD Bend Relief.

Samuel H Cho1, Timothy Fritz2, Asghar Khaghani2.   

Abstract

We report a rare case of a 59-year-old male who presented with a pulsating pseudoaneurysm 15 months after placement of a HeartMate II left ventricular assist device (Thoratec Corporation, Pleasanton, CA, USA). Computed tomography and echocardiogram imaging and surgical exploration revealed that bend relief spontaneously detached from the outflow conduit, and the sharp edges of the snap ring component of the bend relief caused perforation to the outflow graft. <Learning objective: Left ventricular assist devices (LVADs) are of growing popularity within the field of cardiology as a means of bridge to transplant and as a destination therapy. This case highlights a rare, but potentially fatal complication of a popular LVAD, the HeartMate II. Since this case report, the US Food and Drug Administration has released a statement of this particular complication, and Thoratec has made changes to the design. However, many patients still have the older model in place, and are at risk of this rare but potentially fatal complication.>.

Entities:  

Keywords:  HeartMate; Left ventricular assist device; Pseudoaneurysm

Year:  2016        PMID: 30546640      PMCID: PMC6280743          DOI: 10.1016/j.jccase.2016.02.001

Source DB:  PubMed          Journal:  J Cardiol Cases        ISSN: 1878-5409


  3 in total

1.  Sixth INTERMACS annual report: a 10,000-patient database.

Authors:  James K Kirklin; David C Naftel; Francis D Pagani; Robert L Kormos; Lynne W Stevenson; Elizabeth D Blume; Marissa A Miller; J T Baldwin; J Timothy Baldwin; James B Young
Journal:  J Heart Lung Transplant       Date:  2014-04-21       Impact factor: 10.247

2.  Endovascular stent grafting of a left ventricular assist device outflow graft pseudoaneurysm.

Authors:  Asvin M Ganapathi; Nicholas D Andersen; Deyanira J Prastein; Zubair A Hashmi; Joseph G Rogers; Carmelo A Milano; Richard L McCann; G Chad Hughes
Journal:  Circ Heart Fail       Date:  2013-01       Impact factor: 8.790

3.  Prevalence and timing of bend relief disconnection in patients supported by the late version HeartMate II left ventricular assist device.

Authors:  Melana Yuzefpolskaya; Nir Uriel; Daniel S Chow; Susan W Restaino; Donna M Mancini; Margaret Flannery; Rosie Te-Frey; JoAnne Chichetti; Takeyoshi Ota; Hyonah Kim; Drew D Dano; Gregory D Pearson; Hiroo Takayama; Yoshifumi Naka; Ulrich P Jorde
Journal:  J Heart Lung Transplant       Date:  2013-01-16       Impact factor: 10.247