Literature DB >> 30538774

FPGA-Based Interface of Digital DAQ System for Double-Scattering Compton Camera.

Soo Mee Kim1, Young Soo Kim2.   

Abstract

PURPOSE: The double-scattering Compton camera (DSCC) is a radiation imaging system that can provide both unknown source energy spectra and 3D spatial source distributions. The energies and detection locations measured in coincidence with three CdZnTe (CZT) detectors contribute to reconstructing emission energies and a spatial image based on conical surface integrals. In this study, we developed a digital data acquisition (DAQ) board to support our research into coincidence detection in the DSCC.
METHODS: The main components of the digital DAQ board were 12 ADCs and one field programmable gate array (FPGA). The ADCs digitized the analog 96-channel CZT signals at a sampling rate of 50 MHz and transferred the serialized ADC samples and the bit and frame clocks to the FPGA. In order to correctly capture the ADC sample bits in the FPGA, we conducted individual sync calibrations for all the ADC channels to align the bit and frame clocks to the right positions of the ADC sample bits. The FPGA logic design was composed of IDELAY and IDDR components, six shift registers, and bit slip buffer resources.
RESULTS: Using a Deskew test pattern, the delay value of the IDELAY component was determined to align the bit clock to the center of each sample bit. We determined the bit slip in the 12-bit ADC sample using an MSB test pattern by checking where the MSB value of one is located in the captured parallel data.
CONCLUSIONS: After sync calibration, we tested the interface between the ADCs and the FPGA with a synthetic analog Gaussian signal. The 96 ADC channels yielded a mean R 2 goodness-of-fit value of 0.95 between the Gaussian curve and the captured 12-bit parallel data.

Entities:  

Keywords:  Double-scattering Compton camera (DSCC); FPGA-based digital DAQ system; Prompt gamma imaging; Sync calibration of interface between ADC and FPGA

Year:  2018        PMID: 30538774      PMCID: PMC6261862          DOI: 10.1007/s13139-018-0551-8

Source DB:  PubMed          Journal:  Nucl Med Mol Imaging        ISSN: 1869-3474


  10 in total

Review 1.  Molecular imaging of cancer with positron emission tomography.

Authors:  Sanjiv Sam Gambhir
Journal:  Nat Rev Cancer       Date:  2002-09       Impact factor: 60.716

2.  Reconstruction methods and completeness conditions for two Compton data models.

Authors:  Bruce Smith
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  2005-03       Impact factor: 2.129

3.  Optimizing a three-stage Compton camera for measuring prompt gamma rays emitted during proton radiotherapy.

Authors:  S W Peterson; D Robertson; J Polf
Journal:  Phys Med Biol       Date:  2010-11-03       Impact factor: 3.609

Review 4.  The Current Status of SPECT or SPECT/CT in South Korea.

Authors:  Ikdong Yoo; Eun Kyoung Choi; Yong-An Chung
Journal:  Nucl Med Mol Imaging       Date:  2016-04-22

Review 5.  Contrast-enhanced dual mode imaging: photoacoustic imaging plus more.

Authors:  Sungjo Park; Unsang Jung; Seunghyun Lee; Donghyun Lee; Chulhong Kim
Journal:  Biomed Eng Lett       Date:  2017-01-23

6.  Resolution recovery reconstruction for a Compton camera.

Authors:  Soo Mee Kim; Hee Seo; Jin Hyung Park; Chan Hyeong Kim; Chun Sik Lee; Soo-Jin Lee; Dong Soo Lee; Jae Sung Lee
Journal:  Phys Med Biol       Date:  2013-04-08       Impact factor: 3.609

7.  An electronically collimated gamma camera for single photon emission computed tomography. Part I: Theoretical considerations and design criteria.

Authors:  M Singh
Journal:  Med Phys       Date:  1983 Jul-Aug       Impact factor: 4.071

8.  Fully three-dimensional OSEM-based image reconstruction for Compton imaging using optimized ordering schemes.

Authors:  Soo Mee Kim; Jae Sung Lee; Chun Sik Lee; Chan Hyeong Kim; Myung Chul Lee; Dong Soo Lee; Soo-Jin Lee
Journal:  Phys Med Biol       Date:  2010-08-11       Impact factor: 3.609

Review 9.  FDG Whole-Body PET/MRI in Oncology: a Systematic Review.

Authors:  Hyun Woo Kwon; Ann-Katharina Becker; Jin Mo Goo; Gi Jeong Cheon
Journal:  Nucl Med Mol Imaging       Date:  2016-04-07

10.  Electrodeless conductivity tensor imaging (CTI) using MRI: basic theory and animal experiments.

Authors:  Saurav Z K Sajib; Oh In Kwon; Hyung Joong Kim; Eung Je Woo
Journal:  Biomed Eng Lett       Date:  2018-04-25
  10 in total