Ultrasound tomography in bone mimicking phantoms: Simulations and experiments.

Timothe Falardeau1, Pierre Belanger1.   

Abstract

Bone quality assessment for osteoporosis diagnosis is usually performed using dual energy X-ray absorptiometry or X-ray quantitative computed tomography. Recent research demonstrated that both methods are inaccurate in diagnosing osteoporosis since they rely only on the bone mineral density. The literature on bone quantitative ultrasound suggests that ultrasonic waves are sensitive to multiple significant bone parameters such as mechanical properties, the bone volume fraction, and the micro-architecture. Typical ultrasound tomography techniques are limited to image objects with a low speed of sound contrast relative to a background medium. In this study, the possibility of adapting a more advanced ultrasound inversion technique referred to as the hybrid algorithm for robust breast ultrasound tomography for velocity mapping of bone mimicking phantoms was examined. The cortical bone thickness and the cortical bone speed of sound, which are directly related to the bone elastic properties, are parameters strongly correlated with the overall bone quality. A finite element model and an experimental test bench were developed to adapt the hybrid algorithm for robust breast ultrasound tomography to bone quality assessment. Although artefacts were present in the images generated, the results obtained enabled discrimination of a healthy bone phantom over an osteoporotic bone phantom based on the cortical bone thickness and the average cortical bone velocity. The speed of sound inside the cortical region of the bone phantoms was underestimated by 9.38% for the osteoporotic phantom, and by 10.68% for the healthy phantom relative to the values supplied by the bone phantom manufacturer, but there was a difference of 3.97% between the two samples. The difference between the measured cortical bone thickness of the reconstructed image and the X-ray computed tomography images was on average 0.4 mm.

Entities:  

Mesh:

Year:  2018        PMID: 30522285     DOI: 10.1121/1.5079533

Source DB:  PubMed          Journal:  J Acoust Soc Am        ISSN: 0001-4966            Impact factor:   1.840


  3 in total

Review 1.  Criteria for the design of tissue-mimicking phantoms for the standardization of biophotonic instrumentation.

Authors:  Lina Hacker; Heidrun Wabnitz; Antonio Pifferi; T Joshua Pfefer; Brian W Pogue; Sarah E Bohndiek
Journal:  Nat Biomed Eng       Date:  2022-05-27       Impact factor: 25.671

2.  Ultrasound Computed Tomography.

Authors:  Philippe Lasaygues; Luis Espinosa; Simon Bernard; Philippe Petit; Régine Guillermin
Journal:  Adv Exp Med Biol       Date:  2022       Impact factor: 2.622

3.  Analysis of Bone Strength and Bone Turnover Markers in Ankylosing Spondylitis with Radiological Hip Involvement.

Authors:  Wei Liu; Hui Song; Siliang Man; Hongchao Li; Liang Zhang
Journal:  Med Sci Monit       Date:  2021-06-27
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.