Natalia Porporato1, Mani Baskaran1, Tin A Tun1, Rehena Sultana2, Marcus C L Tan3, Joanne H M Quah4, John Allen2, David S Friedman5, Ching-Yu Cheng6, Tin Aung7. 1. Singapore Eye Research Institute, Singapore National Eye Centre, Singapore. 2. DUKE-NUS Medical School, Singapore. 3. Yong Loo Lin School of Medicine, National University of Singapore, Singapore. 4. SingHealth Polyclinics, Singapore; SingHealth Duke-NUS Family Medicine Academic Clinic Program, Singapore. 5. Wilmer Eye Institute, Dana Center for Preventive Ophthalmology and Johns Hopkins Bloomberg School of Public Health, Baltimore, Maryland, USA. 6. Singapore Eye Research Institute, Singapore National Eye Centre, Singapore; DUKE-NUS Medical School, Singapore. 7. Singapore Eye Research Institute, Singapore National Eye Centre, Singapore. Electronic address: aung.tin@singhealth.com.sg.
Abstract
PURPOSE: To evaluate the diagnostic performance of swept-source optical coherence tomography (SS-OCT, CASIA SS-1000; Tomey Corporation, Nagoya, Japan) for angle closure detection, in comparison with gonioscopy, in a community setting. DESIGN: Reliability analysis. METHODS: A total of 2027 phakic subjects aged ≥50 years, with no previous history of glaucoma, laser (including peripheral iridotomy), intraocular surgery, or ocular trauma, were consecutively recruited from a community polyclinic in Singapore. Gonioscopy was performed by a single trained ophthalmologist. SS-OCT angle scans, which obtain radial scans for the entire circumference of the angle, were analyzed by a single examiner, masked to the subject's clinical details. On SS-OCT images, angle closure was defined as contact between the iris and any part of the angle wall anterior to the scleral spur. Different cutoff values of the degree of circumferential angle closure (≥35%, ≥50%, and ≥75%) were taken for analysis to assess SS-OCT performance in detecting angle closure. RESULTS: A total of 1857 subjects (91.6%) were included in the final analysis after excluding poor-quality SS-OCT scans. Almost 90% of the subjects were Chinese, with a mean age of 61.8 ± 6.7 years, and more than half were women (63.5%). The overall AUC of SS-OCT manual grading against gonioscopy was 0.84 (95% confidence interval, 0.81-0.88). The prevalence of angle closure on SS-OCT was 26.1% for the ≥35% definition, with an area under the curve of 0.80 (0.77-0.84), sensitivity of 82.5% (75.3%-88.4%), and specificity of 78.5% (76.5%-80.4%). The first-order agreement coefficient statistics for the 2-quadrant gonioscopic definition of angle-closure with corresponding ≥35%, ≥50%, and ≥75% angle closure definitions for SS-OCT were good at 0.89 (0.83-0.93), 0.88 (0.842-0.93), and 0.88 (0.831-0.99), respectively. CONCLUSIONS: In this large community-based study, SS-OCT exhibited moderate performance for angle closure detection compared to gonioscopy as the reference standard.
PURPOSE: To evaluate the diagnostic performance of swept-source optical coherence tomography (SS-OCT, CASIA SS-1000; Tomey Corporation, Nagoya, Japan) for angle closure detection, in comparison with gonioscopy, in a community setting. DESIGN: Reliability analysis. METHODS: A total of 2027 phakic subjects aged ≥50 years, with no previous history of glaucoma, laser (including peripheral iridotomy), intraocular surgery, or ocular trauma, were consecutively recruited from a community polyclinic in Singapore. Gonioscopy was performed by a single trained ophthalmologist. SS-OCT angle scans, which obtain radial scans for the entire circumference of the angle, were analyzed by a single examiner, masked to the subject's clinical details. On SS-OCT images, angle closure was defined as contact between the iris and any part of the angle wall anterior to the scleral spur. Different cutoff values of the degree of circumferential angle closure (≥35%, ≥50%, and ≥75%) were taken for analysis to assess SS-OCT performance in detecting angle closure. RESULTS: A total of 1857 subjects (91.6%) were included in the final analysis after excluding poor-quality SS-OCT scans. Almost 90% of the subjects were Chinese, with a mean age of 61.8 ± 6.7 years, and more than half were women (63.5%). The overall AUC of SS-OCT manual grading against gonioscopy was 0.84 (95% confidence interval, 0.81-0.88). The prevalence of angle closure on SS-OCT was 26.1% for the ≥35% definition, with an area under the curve of 0.80 (0.77-0.84), sensitivity of 82.5% (75.3%-88.4%), and specificity of 78.5% (76.5%-80.4%). The first-order agreement coefficient statistics for the 2-quadrant gonioscopic definition of angle-closure with corresponding ≥35%, ≥50%, and ≥75% angle closure definitions for SS-OCT were good at 0.89 (0.83-0.93), 0.88 (0.842-0.93), and 0.88 (0.831-0.99), respectively. CONCLUSIONS: In this large community-based study, SS-OCT exhibited moderate performance for angle closure detection compared to gonioscopy as the reference standard.
Authors: Jasmeen Randhawa; Michael Chiang; Natalia Porporato; Anmol A Pardeshi; Justin Dredge; Galo Apolo Aroca; Tin A Tun; Joanne HuiMin Quah; Marcus Tan; Risa Higashita; Tin Aung; Rohit Varma; Benjamin Y Xu Journal: Br J Ophthalmol Date: 2021-10-20 Impact factor: 5.908
Authors: Ivano Riva; Eleonora Micheletti; Francesco Oddone; Carlo Bruttini; Silvia Montescani; Giovanni De Angelis; Luigi Rovati; Robert N Weinreb; Luciano Quaranta Journal: J Clin Med Date: 2020-11-25 Impact factor: 4.241