Communicating vessels volumeter to measure upper extremity lymphedema after breast cancer: reliability and criterion validity compared to the gold standard.

Rogério Mendonça de Carvalho1, Fausto Miranda2.   

Abstract

BACKGROUND: The traditional overflow method for measuring limb volume remains the gold standard, but many disadvantages still inhibit its routine use in clinical practice.
OBJECTIVE: To assess the intra-rater and inter-rater reliability and criterion validity of the 'communicating vessels volumeter' (CVV) for volume measurement of lymphedematous upper extremities (LUE) by using the overflow volumeter (OV) as the reference standard.
METHODS: Twelve LUE of 12 women undergoing mastectomy for breast cancer were measured three times each by three raters using both methods, totaling 216 volume measurements. Criterion validity was estimated by 33 volume measurements of one cylinder of known volume by three raters using both methods, totaling 198 measurements.
RESULTS: Measurement time was short with both CVV and OV. The intraclass correlation coefficient3,1 was high for both CVV and OV in intra-rater (0.99 vs 0.99) and inter-rater (0.99 vs 0.99) analyses. The bias between methods was low (7.50mL; 0.40%) and the limits of agreement were narrow (-5.80 to 6.50%). The volumes were statistically equal with a strong correlation (R2=0.98) between methods. CVV was more accurate than OV (0.00 vs 0.02%) in cylinder measurements.
CONCLUSION: The high intra-rater and inter-rater reliability rates of CVV were comparable to those of OV, and the volumes resulting from LUE measurements were statistically equal in the two methods. Criterion validity rates indicated that CVV measurements were closer to the actual value of the cylinder than those obtained with the OV.
Copyright © 2018 Associação Brasileira de Pesquisa e Pós-Graduação em Fisioterapia. Publicado por Elsevier Editora Ltda. All rights reserved.

Entities:  

Keywords:  Anthropometry; Biomedical technology; Measurement equipment; Medical devices

Mesh:

Year:  2018        PMID: 30448062      PMCID: PMC6849093          DOI: 10.1016/j.bjpt.2018.10.015

Source DB:  PubMed          Journal:  Braz J Phys Ther        ISSN: 1413-3555            Impact factor:   3.377


  31 in total

Review 1.  Non-invasive assessment of the lymphedematous limb.

Authors:  A W Stanton; C Badger; J Sitzia
Journal:  Lymphology       Date:  2000-09       Impact factor: 1.286

2.  The risk of developing arm lymphedema among breast cancer survivors: a meta-analysis of treatment factors.

Authors:  Rebecca J Tsai; Leslie K Dennis; Charles F Lynch; Linda G Snetselaar; Gideon K D Zamba; Carol Scott-Conner
Journal:  Ann Surg Oncol       Date:  2009-04-14       Impact factor: 5.344

3.  Volume measurement using the Polhemus FastSCAN 3D laser scanning: a novel application for burns clinical research.

Authors:  Dale Edgar; Rob Day; N Kathy Briffa; Joan Cole; Fiona Wood
Journal:  J Burn Care Res       Date:  2008 Nov-Dec       Impact factor: 1.845

4.  A technique for fast and accurate measurement of hand volumes using Archimedes' principle.

Authors:  S Hughes; J Lau
Journal:  Australas Phys Eng Sci Med       Date:  2008-03       Impact factor: 1.430

5.  Validation of an optoelectronic limb volumeter (Perometer).

Authors:  A W Stanton; J W Northfield; B Holroyd; P S Mortimer; J R Levick
Journal:  Lymphology       Date:  1997-06       Impact factor: 1.286

6.  Development and evaluation of a precision forearm and hand volumeter and measuring cylinder.

Authors:  R Boland; R Adams
Journal:  J Hand Ther       Date:  1996 Oct-Dec       Impact factor: 1.950

7.  Edema volume, not timing, is the key to success in lymphedema treatment.

Authors:  S M Ramos; L S O'Donnell; G Knight
Journal:  Am J Surg       Date:  1999-10       Impact factor: 2.565

8.  Measurement of upper extremity volume in women after axillary dissection for breast cancer.

Authors:  A M Megens; S R Harris; C Kim-Sing; D C McKenzie
Journal:  Arch Phys Med Rehabil       Date:  2001-12       Impact factor: 3.966

9.  Reliability and limits of agreement of circumferential, water displacement, and optoelectronic volumetry in the measurement of upper limb lymphedema.

Authors:  T Deltombe; J Jamart; S Recloux; C Legrand; N Vandenbroeck; S Theys; P Hanson
Journal:  Lymphology       Date:  2007-03       Impact factor: 1.286

10.  A new, effective and low-cost three-dimensional approach for the estimation of upper-limb volume.

Authors:  Roberto Buffa; Elena Mereu; Paolo Lussu; Valeria Succa; Tonino Pisanu; Franco Buffa; Elisabetta Marini
Journal:  Sensors (Basel)       Date:  2015-05-26       Impact factor: 3.576

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.