Feasibility of 4DCBCT-based proton dose calculation: An ex vivo porcine lung phantom study.

Katharina Niepel1, Florian Kamp2, Christopher Kurz3, David Hansen4, Simon Rit5, Sebastian Neppl2, Jan Hofmaier2, David Bondesson6, Christian Thieke2, Julien Dinkel6, Claus Belka7, Katia Parodi1, Guillaume Landry8.   

Abstract

Inter-fractional variations of breathing pattern and patient anatomy introduce dose uncertainties in proton therapy. One approach to monitor these variations is to utilize the cone-beam computed tomography (CT, CBCT) scans routinely taken for patient positioning, reconstruct them as 4DCBCTs, and generate 'virtual CTs' (vCTs), combining the accurate CT numbers of the diagnostic 4DCT and the geometry of the daily 4DCBCT by using deformable image registration (DIR). In this study different algorithms for 4DCBCT reconstruction and DIR were evaluated. For this purpose, CBCT scans of a moving ex vivo porcine lung phantom with 663 and 2350 projections respectively were acquired, accompanied by an additional 4DCT as reference. The CBCT projections were sorted in 10 phase bins with the Amsterdam-shroud method and reconstructed phase-by-phase using first a FDK reconstruction from the Reconstruction Toolkit (RTK) and again an iterative reconstruction algorithm implemented in the Gadgetron Toolkit. The resulting 4DCBCTs were corrected by DIR of the corresponding 4DCT phases, using both a morphons algorithm from REGGUI and a b-spline deformation from Plastimatch. The resulting 4DvCTs were compared to the 4DCT by visual inspection and by calculating water equivalent thickness (WET) maps from the phantom's surface to the distal edge of a target from various angles. The optimized procedure was successfully repeated with mismatched input phases and on a clinical patient dataset. Proton treatment plans were simulated on the 4DvCTs and the dose distributions compared to the reference based on the 4DCT via gamma pass rate analysis. A combination of iterative reconstruction and morphons DIR yielded the most accurate 4DvCTs, with median WET differences under 2mm and 3%/3mm gamma pass rates per phase between 89% and 99%. These results suggest that image correction of iteratively reconstructed 4DCBCTs with a morphons DIR of the planning CT may yield sufficiently accurate 4DvCTs for daily time resolved proton dose calculations.
Copyright © 2018. Published by Elsevier GmbH.

Entities:  

Keywords:  Adaptive proton therapy; Cone-beam CT; Dose monitoring; Proton therapy

Mesh:

Year:  2018        PMID: 30448049     DOI: 10.1016/j.zemedi.2018.10.005

Source DB:  PubMed          Journal:  Z Med Phys        ISSN: 0939-3889            Impact factor:   4.820


  3 in total

Review 1.  Proton beam therapy for tumors of the upper abdomen.

Authors:  Ann Raldow; James Lamb; Theodore Hong
Journal:  Br J Radiol       Date:  2019-08-23       Impact factor: 3.039

Review 2.  Roadmap: proton therapy physics and biology.

Authors:  Harald Paganetti; Chris Beltran; Stefan Both; Lei Dong; Jacob Flanz; Keith Furutani; Clemens Grassberger; David R Grosshans; Antje-Christin Knopf; Johannes A Langendijk; Hakan Nystrom; Katia Parodi; Bas W Raaymakers; Christian Richter; Gabriel O Sawakuchi; Marco Schippers; Simona F Shaitelman; B K Kevin Teo; Jan Unkelbach; Patrick Wohlfahrt; Tony Lomax
Journal:  Phys Med Biol       Date:  2021-02-26       Impact factor: 4.174

3.  Technical Note: 4D cone-beam CT reconstruction from sparse-view CBCT data for daily motion assessment in pencil beam scanned proton therapy (PBS-PT).

Authors:  Lydia A den Otter; Kuanling Chen; Guillaume Janssens; Arturs Meijers; Stefan Both; Johannes A Langendijk; Lane R Rosen; Hsinshun T Wu; Antje-Christin Knopf
Journal:  Med Phys       Date:  2020-10-24       Impact factor: 4.071