| Literature DB >> 30444230 |
Zhengda Lu1, Elizabeth Rey1, Sasank Vemulapati1, Balaji Srinivasan2, Saurabh Mehta3, David Erickson4.
Abstract
Interest in developing paper-based devices for point-of-care diagnostics in resource-limited settings has risen remarkably in recent decades. In this paper, we demonstrate what we refer to as "high yield passive rrythrocyte removal" (HYPER) technology, which utilizes capillary forces in a unique cross-flow filtration for the separation of whole blood with performance comparable to centrifuges. As we will demonstrate, state-of-the-art passive blood separation methods implemented in paper-based systems exhibit rapid blood cell clogging on the filtration media or serum outlet and yield only about 10-30% of the total serum present in the sample. Our innovation results from the inclusion of a differentiation pad, which exploits hydrodynamic effects to reduce the formation of a fouling layer on the blood filtration membrane resulting in more than 60% serum yield with undiluted whole blood as direct input. To demonstrate the effectiveness of the HYPER technology we implement it in a lateral flow system and demonstrate the accurate quantification of vitamin A and iron levels in whole blood samples in 15 minutes.Entities:
Mesh:
Substances:
Year: 2018 PMID: 30444230 PMCID: PMC6287273 DOI: 10.1039/c8lc00717a
Source DB: PubMed Journal: Lab Chip ISSN: 1473-0189 Impact factor: 6.799