Literature DB >> 36033130

A review of portable quantitative and semi-quantitative devices for measurement of vitamin A in biological samples.

Samantha L Huey1, Jesse T Krisher1, David Morgan2, Penjani Mkambula2, Bryan M Gannon1, Mduduzi N N Mbuya3, Saurabh Mehta1,4.   

Abstract

Background: We catalog and summarize evidence of the analytical performance of portable quantitative and semi-quantitative devices for the assessment of vitamin A status and vitamin A deficiency (VAD) in various biological samples-including whole blood, plasma, serum, and milk-in addition to VAD determination by functional indicators such as pupillary response.
Methods: We searched the literature for published research articles, patents, and information from manufacturers of mobile devices, particularly those appropriate for low-resource settings. The included devices were required to be portable (lightweight and ideally not needing a power outlet) and to measure vitamin A as well as define VAD. Eligible studies compared a portable device to a reference standard of high-performance liquid chromatography for blood and milk, or a Goldmann-Weekers dark adaptometer for eyes/vision. Where available, identified devices were compared with reference methods across several performance criteria. When possible, we compared the device's performance reported in published studies against the stated performance criteria from the manufacturers' websites.
Results: We catalogued 25 portable devices for measuring vitamin A and/or VAD via biological samples. We also identified 18 comparison studies (plus associated reports) assessing nine methods: the iCheck Fluoro, iCheck Carotene, CRAFTi, Tidbit with or without the HYPER filtration system, custom field-friendly immunoassays, and microfluidic assays for blood; the iCheck Fluoro and iCheck Carotene for milk; and the Scotopic Sensitivity Tester-1 for eye function. Conclusions: The iCheck Fluoro and iCheck Carotene are commercially available for use and are acceptable for measuring vitamin A in blood and milk samples, according to the available validation data. Many of the other identified devices, including other portable fluorometers, photometers, immunoassays, microfluidics-based devices, and dark adaptometers, were proofs of concept and not yet commercially available. Furthermore, none of these other devices included manufacturer-described device performance criteria to compare with descriptions from experimental studies. Several gaps remain, including studies comparing the other portable devices against a reference standard, particularly for functional indicators of vitamin A status/deficiency; available manufacturer-reported device performance criteria against which to compare future results of investigations; and more comprehensive reporting of validation metrics including sensitivity, specificity, precision, and Bland-Altman analysis.
© 2022 The Author(s).

Entities:  

Keywords:  Beta carotene; Field devices; Portable devices; Resource-limited settings; Retinol; Vitamin A

Year:  2022        PMID: 36033130      PMCID: PMC9407042          DOI: 10.1016/j.crbiot.2022.04.003

Source DB:  PubMed          Journal:  Curr Res Biotechnol        ISSN: 2590-2628


Introduction/background

Vitamin A deficiency (VAD) continues to be a major global health issue leading to poor health outcomes, including night blindness, greater severity of measles infection, and higher mortality risk from infectious diseases (Tanumihardjo et al., 2016). Most existing analytical techniques to assess vitamin A status by measuring serum retinol or retinol binding protein require access to a sophisticated laboratory and equipment such as high performance liquid chromatography (HPLC) (de Pee and Dary, 2002). These methods require extensive sample preparation, are time-consuming, and are potentially prohibitively expensive, depending on the number of samples to be analyzed. Furthermore, VAD is more prevalent in lower income countries, where such laboratory resources may be limited or might not yet exist; in recent vitamin A surveys, <20% of pregnant women at risk have been covered by population surveys globally, possibly partly because of a lack of diagnostics (WHO, 2009). Portable, field-friendly devices and tools for assessing vitamin A status in populations have the potential to overcome some of the limitations of traditional, laboratory-based testing. These methods may differ in their cost, accuracy, reliability, ease of use, and required consumables/reagents for performing the testing. A review cataloguing the range of portable tests for vitamin A status and VAD in biological samples, and summarizing these devices’ performance with respect to a reference standard method, is not available. Therefore, the goal of this review was to enable current manufacturers to modify and improve their products according to the gaps identified herein, and to set design goals for new products meeting the current demands of industry, regulators, and other stakeholders.

Materials and methods

In December 2020, we conducted a standardized search of the literature indexed in five databases (MEDLINE, EMBASE, World Health Organization Global Index Medicus, Scopus, and Web of Science) with no restrictions on language, location, or date of publication. We designed a search strategy for MEDLINE (PubMed) (Supplementary Table 1) and translated the search strategy for the remaining databases with guidance from the evidence synthesis specialists at Mann Library, Cornell University. We also used an online search engine to search for other sources such as manufacturers’ websites and patents, and we consulted with subject matter experts within our organizations to gain more information. We catalogued any portable devices measuring vitamin A or vitamin A deficiency in biological samples, either as reported in studies or provided on manufacturers’ websites. We included both portable devices/methods measuring vitamin A status and devices/methods that indicated VAD. Initially, we considered devices measuring skin carotenoids, as shown in our search strategy; however, because of the lack of established guidance or consensus regarding the conversion of skin carotenoid measurements via Raman resonance spectroscopy (e.g., BioPhotonic Scanner (Pharmanex/Nuskin Enterprises, 2018)) to blood carotenoid measurements and overall vitamin A status (von Lintig, 2020), we determined that these devices were beyond the scope of the review. The inclusion criteria for our analysis of device performance included certain study designs such as proof-of-concept development studies, method comparison studies, and diagnostic test accuracy studies; studies involving human participants (e.g., observational studies or randomized controlled trials) were considered if the authors described using a portable method for analyzing vitamin A in biological samples. Animal studies were also included. Eligible studies were required to measure vitamin A in any biological sample, including blood, eyes, or breast milk, with a portable device and to compare the device performance with that of a reference method, such as HPLC, depending on the sample type. Studies detailing field friendly methods of sample collection (e.g., dried blood spots) necessitating the use of a non-portable device or a laboratory for analysis were considered beyond the scope of this review. We contacted the authors to request raw data or more information as needed. We also re-analyzed raw data, when available, as needed.

Results and discussion

Catalog of portable devices

From our search (Fig. 1), we catalogued 25 portable devices, kits, and/or field-friendly assays able to assess a variety of biological sample and vitamin A biomarker types in Table 1a (blood, milk) and Table 1b (eyes/vision assessment).
Fig. 1

PRISMA Diagram for study identification and screening. ().

Table 1a

Catalog of all portable devices quantifying vitamin A and vitamin A deficiency: blood, milk.

Device (manufacturer)Principle/method# Tests per kitVitamin A biomarkerPricingTechnical requirementsSample volume, preparation and setupOverall time requiredPortabilityIncluded in kitSpecial storage conditionsConsumables:Reagents requiredPower sourceShelf lifeOperational rangeQuantificationOutputsTarget settingManufacturer support availableGlobal availability
iCheck Fluoro (BioAnalyt GmBH, Teltow, Germany) (BioAnalyt, 2021)Fluorescence100 tests per kitRetinol, retinyl palmitate, retinyl acetate and other estersPricing not published+1 day trainingWhole blood, serum, breast milk: 0.5 mL; no preparation required<10 min+++Compact and lightweight (11 × 4 × 20 cm); 0.45 kgDevice + test kit (iEX MILA reaction vials), syringeNone+++Optional:50 mL conical tubes, weighing dishes, reference samplesRechargeable battery12 months at 20–30 °C, no direct sunlight, upright+++50–3000 µg RE/LQuantitativeSample #, batch #, result, date, time (in transferred data); results (µg RE/L) are stored in the device and transferred to a PC via USB+++Lab and fieldYes>80 countries+++
iCheck Carotene (BioAnalyt GmBH, Teltow, Germany) (BioAnalyt, 2021)Photometry100 tests per kitBeta-carotenePricing not published+1 day trainingColostrum, cattle whole blood, cattle serum: 0.4 mL; no preparation required<10 min+++Compact and lightweight (11 × 4 × 20 cm); 0.45 kgDevice + test kitNone+++Optional:50 mL conical tubes, weighing dishes, reference samplesRechargeable battery12 months at 20–30 °C, no direct sunlight, upright+++0.15–15 mg/LQuantitativeSample #, batch #, result, date, time (in transferred data); results (µg RE/L) are stored in the device and transferred to a PC via USB+++Lab and fieldYes>80 countries+++
CRAFTi (Eurofins CRAFT Technologies Inc., Wilson, NC, USA) (Chaimongkol et al., 2011, Eurofins Craft Technologies, 2020)FluorescenceNRRetinolPricing not published+Minimal trainingSerum: 25 µL; requires serum separation30 min++Compact and lightweight (13 × 16.5 × 35 cm); 2.1 kgNRNR++Fluorometer cuvettes or Durham tubesBattery (12 V and inverter) or line current (115–230 V)Fluorescent dye fades; must check periodically++0.5–1.5 µmol/LQuantitativeFluorescence readings; no detail on data appearance++Lab and fieldYesNR++
Tidbit (Lu and Erickson, 2017, Lu et al., 2017), ± HYPER filtration system (Lu et al., 2018) (Cornell University, Ithaca, NY, USA)Fluorescence, multicolor lateral flowNRRBPEstimated: $95 manufacturing cost; $1.50 per test; Using HYPER platform, <$1 per test++Meant for consumer, clinical, and research useSerum: 15–20 µL, separated from RBCs with a centrifuge Whole blood: 60 µL, using HYPER (Lu et al., 2018)Serum: 15 minWhole blood: 5–20 min for HYPER separation)+++NR; meant for field useTidbit reader, disposable test strip(s)None+++Lightening-Link Conjugation Kits (Innova Bioscience Ltd., HF180 cards (EMD Millipore); Running buffer (60 µL)Battery; connect to mobile device, Wi-Fi optionalNR+++2.2–20 µg/mL (0.10–0.95 µmol/L)QuantitativeResult from each sample to smartphone (Nutriphone app) or laptop; stores results internally via 16-GB SD card+++Lab and fieldCorresponding authors: David Erickson or Saurabh Mehta; not commercially availableNR++
Electronics-enabled (EE)-µPAD (Diagnostics for All) (Lee et al., 2016)Paper-based microfluidics for immune detectionNRRBPEstimated: $20 for prototyping; $0.41 per test, but price expected to decrease below $10 per unit and at $1 per biomarker per unit++Meant for clinicians and researchersWhole blood: 35 µL; no preparation required13 min+++Size of a credit cardNRNR++µPADBatteryStored at room temperature in desiccator box until use+++∼10 µg/mL to < 70 µg/mL, according to on graph (Fig. 5)QuantitativeMeasurements are wirelessly transferred to a mobile phone application that geo-tags the data and transmits it to a remote server for real time tracking of micronutrient deficiencies; NFC-enabled smartphone required+++Resource limited settingsYes, but not commercially availableNR++
RBP-EIA (Scimedx Corp., Dover, NJ, USA) (Hix et al., 2004, Hix et al., 2006, 2020)Antigen competition assay8–96 tests per kitRBPEstimated: <$3.00 per test; pricing not published++Meant for health care workersSerum: 10 µL; portable battery-operated centrifuge to separate whole blood, vortex serum samples and store on ice until assay completed40 min++Size of 96-well plate; requires sink for washing stepNRNR++Well plate, monoclonal anti-RBP antibody, wash buffer, substrateNone requiredNR++10–40 µg/mL (0.48–1.92 µmol/L)QuantitativeRead optical densities using EIA plate reader (Revelation, Dynex)++Lab and fieldYesNR+++
Antigen-antibody reaction based on liquid-semisolid phase (custom) (Ciaiolo et al., 2015)Custom antigen–antibody reaction based on liquid-semisolid phase + visualization systemNRRBPPricing not published+Meant for research or diagnosticsSerum: 5 µL; requires serum separation30 min++Petri dishes, pipette, portable viewer, and glass gel holderNRNR++Dilution: PBS due to high protein concentration; gel; antibodiesNone requiredNR++Depends on time for reaction; range 64 mg/L to 1 mg/LQualitative, semi-quantitativeImmunoprecipitates, scored as: “-, +, ++, +++, ++++”++Lab and fieldCorresponding author: Carlo Ciaiolo; but not commercially availableNR++
RID plate reader (The Binding Site, San Diego, CA, USA) (Hix et al., 2004, The Binding Site, 2020)Radial immunodiffusion of antigen–antibody precipitin rings1–3 plates per kitRBPPricing not published+Meant for researchersSerum: 5 µL; requires serum separationIncubation for 3 days+Compact and lightweight (22 × 14 × 16 cm, 1.14 kg)User guide and installation CD, USB-A to USB-B cable, power supply, plate reader calibration plateIndoor use only, altitude < 2000 m, 5–40 °C, relative humidity ≤ 80% at < 31 °C (or ≤ 50% if > 31 °C)++Microsoft Windows computerPower adapter, USB portNR (warranty: 1 year)++Depends on analyte; range for RBP not reportedQuantitativePrecipitin ring diameters, mm++LabYesNR++
Reference method: HPLCChromatography$20000–$50000 per machine$50–$100 per test+Meant for researchers,≥500 µL, requires HPLC solvent and other preparation≥65 min+Not portablen/aControlled conditions+Can be used for different analyses or when the procurement of vials is difficultRequires external power sourceRequires routine maintenance+Depends on analyteQuantitativeExact concentration output on attached computer, chromatogram with quantified absorbance for vitamin A concentration+++LabYes+++

Notes: EIA, enzyme immunoassay; HPLC, high-performance liquid chromatography; NR, not reported; RBP, retinol-binding protein; RID, radial immunodiffusion.

+++ = best.

++ = acceptable.

+ = not acceptable.

Table 1b

Catalog of all portable devices quantifying vitamin A and vitamin A deficiency: eyes.

Device (manufacturer)Principle/methodVitamin A biomarkerPricing (estimated, list price from manufacturer website)Technical requirementsSample site; time for full chargePortabilityIncluded in kitPower sourceUsage duration per chargeSlit lamps:MagnificationDioptic rangeInterpupillary rangeSlit image width(s)FiltersTarget settingManufacturer support availableGlobal availability
BA 904, BA 904C (Haag-Streit, Harlow, Essex, UK) Haag-Streit, 1900Slit lampOcular morbiditiesPricing not published+Meant for researchers and ophthalmologistsAnterior segment; 4–5 h++“Lightweight”BA 904: head and chin rest stand, two energy packs, charger, power supply and large case;BA 904C: two energy packs, charger, power supply, parking unit and small case+++Batteries and chargers45 min+++10×, 16×−8 to + 853–95 mmNRBlue, yellow+++Field, clinicYesYes+++
Hand-held digital slit lamp (HSL-100, HSL-150) Portable slit lamp (Heine®) (Melo et al., 2004, Heine Optotechnik GmbH, 2018)Slit lampOcular morbiditiesPricing not published+Meant for researchers and ophthalmologistsAnterior segment; NR++70 gBETA4 SLIM NT rechargeable handle and NT4 table charger (included reducer insert), spare bulb, hard case+++Rechargeable or battery handleNR+++10×, 16×NRNR10 × 0.2 to 14 × 4 mmCobalt blue+++Field, clinicYesYes+++
Portable slit lamp(SL-17) (Kowa Ophthalmic Diagnostic Products, Torrance, CA, USA) (KOWA New Lighter)Slit lampOcular morbiditiesPricing not published+Meant for researchers and ophthalmologistsAnterior segment; NR++<800 g; 220 × 95 × 220 mm4 AAA batteries, dust cover, stand, instruction manual; optional: forehead rest, camera connection adapter+++4 AAA rechargeable or dry cell batteries130–140 min+++10×, 16×NR50–72 mm1 × 1, 0.15, 0.5, 0.8, 1.6, 12Cobalt blue+++Field, clinicYesYes+++
Binocular hand held biomicroscope slit lamp(PSL One, PSL Classic) (Keeler, Malvern, PAUSA)Slit lampOcular morbiditiesPricing not published+Meant for researchers and ophthalmologistsAnterior segment; 2.5 h+++900 g; 238 × 116 × 210 mmBase charger unit, power supply, user instructions, lens cloth+++AC-powered50 min++10×, 16×−7 to + 750–72 mm0.15 mm, 0.5 mm, 0.8 mm and 1.6 mm slits, 12 mm circle and a 1 mm squareRed free, blue, neutral density 0.8 and clear+++Field, clinicYesYes+++
Handheld Slit Lamp S200 (Digital Eye Center, Miami, FL, USA) (Digital Eye Center, 2021, Digital Eye Center, 2021, Digital Eye Center, 2021)Slit lampOcular morbidities$2090++Meant for researchers and ophthalmologistsAnterior segment; NR+++40 g; NRUniversal smartphone adapter, metallic case, accessories.+++Rechargeable battery7 h++10×, 16×Diopter adjustment (not specified)50–74 mmSlit width adjustment (not specified)Red free, green, cobalt blue, heat absorption, clear, neutral density+++Field, clinicYesYes+++
Handheld Slit LampS2 (Digital Eye Center, Miami, FL, USA) (Digital Eye Center, 2021, Digital Eye Center, 2021, Digital Eye Center, 2021)Slit lampOcular morbidities$1500++Meant for researchers and ophthalmologistsAnterior segment; NR+++750 g; 19 × 105 × 230 mmSmartphone adapter, metallic case, accessories+++Rechargeable battery2 h++10×, 16×−5 to + 545–70 mm0–10 mmHeat-absorption, gray, red-free, cobalt blue+++Field, clinicYesYes+++
Digital portable slit lamp Microclear Hyperion (Digital Eye Center, Miami, FL, USA) (Digital Eye Center, 2021, Digital Eye Center, 2021, Digital Eye Center, 2021)Slit lampOcular morbidities$3800++Meant for researchers and ophthalmologistsAnterior segment; NR+++600 g; NR4″ touch screen, 16 GB internal memory, two lithium batteries (4 h each), software and manual+++Rechargeable battery4 h+++10×NRNR0–10 mmHeat-absorption, gray, red-free, cobalt blue+++Field, clinicYesYes+++
Hand held slit lamp (SL280) (Opticlar, Poole, Dorset) (Optical Visionmed)Slit lampOcular morbidities$3900++Meant for researchers and ophthalmologistsAnterior segment; 2 h+++880 g; 163 × 124 × 205 mmBase plate, aluminum case+++Rechargeable battery6 h+++10×, 16×−7 to + 750–78 mm0.15/0.5/0.8/1.6 mm. Circle 12 mm dia. 1 mm squareGreen (red free), cobalt blue, neutral density 0.8, clear+++Field, clinicYesYes+++
Portable slit lamp (PSL) (Reichert Technologies Inc., Depew, NY, USA) (Reichert Technologies, 2021)Slit lampOcular morbiditiesPricing not published+Meant for researchers and ophthalmologistsNR+680 g; fits in palm of handTwo batteries, battery charger+++Rechargeable batteries2 h+++10×, 16×−7 to + 750–70 mm0–11 mmCobalt blue, red free, color temperature conversion+++Field, clinicYesYes+++
Handy Slit Lamp XL-1 (Shin-Nippon by Rexxam Co., Ltd.) (Rexxam, 2021)Slit lampOcular morbiditiesPricing not published+Meant for researchers and ophthalmologistsAnterior segment; NR+++700 g (195 × 105 × 230 mm)Carrying case, one battery, battery charger, forehead support, diopter adjustment bar, instruction manual+++Rechargeable battery2 h+++10×, 16×−7 to + 750–70 mm0–11 mmCobalt blue, green, conversion+++Field, clinicYesYes+++
Portable slit lamp S150 (Medi-Works, Shanghai, China) (Mediworks, 2021)Slit lamp, attachment for phoneOcular morbiditiesPricing not published+Meant for researchers and ophthalmologistsAnterior segment; 3.5 h++240 g; NRNR++Rechargeable batteries6 h+++NRNR0–12 mmCobalt blue+++Field, clinicYesYes+++
SK-LS-1B portable slit lamp (Coburn Technologies, Inc. South Windsor, CT, USA) (Coburn Technologies Inc.)Slit lampOcular morbiditiesPricing not published+Meant for researchers and ophthalmologistsAnterior segment; NR+++835 g; 320 × 310 × 205 mmNR, optional iPhone adapter+++Rechargeable batteries≥4 h+++10×, 16×−7 to + 749–75 mm0.1, 0.2, 0.8, 1, 5, 12 mmNeutral density, red-free, cobalt blue+++Field, clinicYesYes+++
Device (manufacturer)Principle/methodVitamin A biomarkerPricing (estimated, list price from manufacturer website)Technical requirementsSample site; time for full chargePortabilityIncluded in kitPower sourceUsage duration per chargeDark adaptometers and other devices: Other attributesTarget settingManufacturer support availableGlobal availability
RetEval (LKC Technologies, Gaithersburg, MD, USA) (Technologies, 2019)Non-mydriatic flash and flicker ERG/VEP deviceOcular morbiditiesPricing not published+Meant for researchers and ophthalmologistsPupils, retina; 4 h full recharge time+++240 g; 7 × 10 × 23 cmStorage case, charging stand+++Battery-powered8 h+++Pupil measurements: 1.3–9 mm, <0.1 mm resolution, 28.3 HzUSB connectivity+++Field, clinicYesYes+++
Mobile eye testing unit (Agrawal and Sahu, 2020)Vision drum, trial box, retinoscope, slit-lamp bio-microscope, applanation tonometer, and non-mydriatic fundus cameraOcular morbidities: conjunctival xerosis with Bitot’s spot (X1B) or keratomalacia (X3B) (World Health, 2014)Pricing not published+Meant for researchers, optometrists, ophthalmologistsVarious parts of eye+++Indicated to be mobileNR++Varies by device in kit++See (Agrawal and Sahu, 2020)+++FieldN/A; not commercially availableNo/custom++
Scotopic Sensitivity Tester-1 (LKC Technologies, Gaithersburg, MD, USA) (Congdon et al., 1995, Congdon et al., 2000, Sanchez et al., 1997)Dark adaptometerVisual score/thresholdPupillary score/thresholdPupillary dynamicsRod functionPricing not published+Meant for researchers and ophthalmologistsRetina; binocular partial bleaching with camera flash (3433 cd-s/m2)+++Hand-heldN/A+++NRNR+Yellow-green LED light with wavelength at 572 nm, 12 intensity settings, calibrated with EG + G DR 2550 digital radiometer-photometer+++Field, clinicAppears discontinuedNo+
Portable field dark adaptometer (custom)(Labrique et al., 2015, Palmer et al., 2015, Palmer et al., 2016)Dark adaptometerVisual score/thresholdPupillary score/thresholdPupillary dynamicsPricing not published+Meant for researchers and ophthalmologistsRetina; binocular partial bleaching with camera flash (>3400 cd-s/m2)+++Portable: “Its size and weight allowed it to be carried long distances to areas unreachable by car.”Digital camera, a retinal bleaching flash, and a Ganzfeld light source inside a pair of light-obscuring goggles+++Laptop-powered10 tests per day++See (Labrique et al., 2015)Assess impaired pupillary responses to a graded series of Ganzfeld light stimuli applied within a pair of “darkroom” goggles with an embedded microcircuit design and regulated by a laptop-powered controller box++FieldN/A; not commercially availableNo/custom?+
Emtech A meter V.01 (custom)(Mehta, 2018, Mehta, 2019b, Mehta, 2019a, Mehta and Mehta, 2018, Banerjee, 2019)Dark adaptometerDark adaptation; identify pictorial representation of objects at low light intensityPricing not published+Meant for researchers and ophthalmologistsRetina; NR+++HandheldElectronic paper module, LCD to display test object, microSD card, keypad+++NRNR+Results output to microSD card+++FieldN/A; not commercially availableNo/custom?+
Custom-built portable field dark adaptometer(Steven and Wald, 1941, Wald, 1941)Dark adaptometerVisual thresholdDark adaptationPricing not published+Meant for researchers and ophthalmologistsRetina; NR+++8.4 kg; 21.6 × 21.6 × 29.2 cm; “approximate size and shape of a pocket lamp”Eyepiece, test unit, cord, cabinet+++Three 2-volt discharge storage cellsNR+++Results in log units++Field, labN/A; not commercially availableNo/custom?+
Reference method: Goldmann-Weekers dark adaptometer (Haag-Streit)Visual thresholdDark adaptationNot available for purchase+Meant for researchers and ophthalmologists; some models require conversion table for a calibration error (Maggiano et al., 1978)Retina, with pupils dilated; NRRequires 60–120 min in a dark room+Large size and complex; not portableN/A+Requires external power sourceN/A+Results: luminance in units of log microapostilbs, which requires conversion to the more contemporary unit of luminance, cd/m2+ClinicNoOut of production and not available to order+

Notes: NR, not reported.

+++ = best.

++ = acceptable.

+ = not acceptable.

PRISMA Diagram for study identification and screening. (). Catalog of all portable devices quantifying vitamin A and vitamin A deficiency: blood, milk. Notes: EIA, enzyme immunoassay; HPLC, high-performance liquid chromatography; NR, not reported; RBP, retinol-binding protein; RID, radial immunodiffusion. +++ = best. ++ = acceptable. + = not acceptable. Catalog of all portable devices quantifying vitamin A and vitamin A deficiency: eyes. Notes: NR, not reported. +++ = best. ++ = acceptable. + = not acceptable.

Vitamin A deficiency biomarkers

In Table 2, we list definitions of VAD used across studies for a variety of biological sample types, from humans or cattle, including cows, calves, and bulls (Table 2). We also note which studies used particular definitions (e.g., VAD measured as RBP ≤ 0.70 µmol/L was measured by Hix et al. 2004). A previous review by Tanumihardjo (2016) has outlined the utility of biomarkers for vitamin A nutrition status (Tanumihardjo et al., 2016), which we adapted for Table 2. We outline the biomarkers of vitamin A as identified in our literature search below.
Table 2

Definitions of vitamin A deficiency, by sample type and device.

Blood (whole, serum, plasma)a
BiomarkerTypebDevice (studies using)Deficiency or insufficiency definitions usedc
RetinolStatusiCheck Fluoro (BioAnalyt)d (Boateng et al., 2018, Elom et al., 2015, Ghaffari et al., 2019, Raila et al., 2017, Schweigert et al., 2011a, Bechir et al., 2012, Crump et al., 2017, Schweigert et al., 2011b, Whang et al., 2012, Zambo et al., 2012)Spectrophotometer model 450 (Sequoia-Turner) (Marinovic et al., 1997)CRAFTI (Craft Technologies) (Chaimongkol et al., 2011)Severe/clinical deficiency: ≤0.35 µmol/L (10 µg/dL) (WHO, 2011)Low/subclinical deficiency: ≤0.70 µmol/L (20 µg/dL) (WHO, 2011)Insufficiency: ≤1.05 µmol/L (30 µg/dL) (de Pee and Dary, 2002)
RBPStatusCustom REI (Hix et al., 2004)EE-µPAD (Lee et al., 2016)Tidbit (Lu and Erickson, 2017, Lu et al., 2017) ± HYPER filtration (Lu et al., 2018)Custom Ag-Ab reaction (Ciaiolo et al., 2015)Deficiency: ≤0.70 µmol/L (Hix et al., 2004)Deficiency: <16.3 µg/mLe (Lee et al., 2016)Deficiency: <14.7 µg/mL (correlated with retinol ≤ 0.70 µmol/L) (Lu et al., 2017)Not defined (Ciaiolo et al., 2015)
Beta-caroteneNot defined (indicator of recent dietary intake)iCheck Carotene (BioAnalyt)f (Ghaffari et al., 2019, Hye et al., 2020, Klein et al., 2013, Livingston et al., 2020, Meinke et al., 2016, Raila et al., 2012, Madureira et al., 2020)Humans:No official cut-off defined (von Lintig, 2020)Cattle (Klein et al., 2013, De Ondarza and al., 2009, Schweigert and Immig, 2007):Deficient: 0.6–1.5 mg/L or < 1.5 mg/LMarginal: ≥1.5 mg/L to < 3.5 mg/LOptimal: ≥3.5 mg/L
Milk
BiomarkerTypebDeviceDeficiency or insufficiency definitions usedc
RetinolStatus, exposureiCheck Fluoro (BioAnalyt)d (Jans et al., 2018, Abebe et al., 2019, Engle-Stone et al., 2014, Schweigert et al., 2011a, Schweigert et al., 2011b, Bechir et al., 2012, Crump et al., 2017)Humans:Inadequate: <1.05 µmol/L (Blaner, 2020)or milk fat < 8 µg/g (Blaner, 2020)Cattle: not defined
Beta caroteneNot defined (indicator of recent dietary intake)iCheck Carotene (BioAnalyt)d (no studies)Humans: not definedCattle: not defined
Eyes
BiomarkerTypebDevice (studies using)Deficiency or insufficiency definitions usedc
Visual score/thresholdFunctionScotopic sensitivity hand-held illuminator (LKC Technologies, Inc.) (Congdon et al., 1995, Sanchez et al., 1997, Reilly et al., 2006)EmTech A meter V.01g (Mehta, 2018)Portable visual adaptometer (Wald, 1941, Steven and Wald, 1941)Scotopic Sensitivity Tester-1TM (SST-1) (Peters et al., 2000)Abnormal:≥stimulus #10 (Congdon et al., 1995)≥-3.76 log cd/m2 h (Congdon et al., 1995)Highly abnormal:≥stimulus #11≥-3.39 log cd/m2 (Congdon et al., 1995)A decrease of ≥ 0.3 log units after administration of vitamin A supplementation (Wald, 1941, Steven and Wald, 1941)
Dark adaptation: pupillary score/responsiveness [lowest light intensity that stimulated percentage relative change in pupil diameter (Labrique et al., 2015)]FunctionScotopic sensitivity hand-held illuminator (LKC Technologies, Inc.) (Congdon et al., 1995, Sanchez et al., 1997, Peters et al., 2000)Portable field dark adaptometer (PFDA) or digital pupillometer (Labrique et al., 2015, Palmer et al., 2015, Palmer et al., 2016)Normal:≥-1.24 log cd/m2 (Congdon and West, 2002)Abnormal:≥stimulus #9 (Congdon et al., 1995)≥-0.575 log cd/m2 (Congdon et al., 1995) i.e.,≥20% (Labrique et al., 2015)≥-1.11 log cd/m2 (Congdon et al., 2000)≥-0.9 log cd/m2 (Palmer et al., 2016)≥15% relative change in diameter (Labrique et al., 2015)≥10% contraction in pupil size (Palmer et al., 2016)Threshold:≥15 cd/m2 (Khan et al., 2019)
Pupillary dynamics [i.e., response time: absolute value of difference in frame numbers from pre- to post-stimulus divided by number of frames per second (Labrique et al., 2015)]FunctionPortable field dark adaptometer (PFDA) or digital pupillometer (Labrique et al., 2015, Palmer et al., 2015, Palmer et al., 2016)No official cut-off defined
Rod function [dark-adapted rod full-field electro-retinogram responses (Peters et al., 2000)]FunctionScotopic Sensitivity Tester-1TM (SST-1) (Peters et al., 2000)No official cut-off defined
Ocular morbiditiesFunctionMobile eye unit (comprised of vision drum, trial box, retinoscope, slit-lamp bio-microscope, applanation tonometer, non-mydriatic fundus camera) (Agrawal and Sahu, 2020)Heine HSL-100 biomicroscope equipped with portable slit lamp (Melo et al., 2004)Night blindness, conjunctival xerosis with Bitot’s spots (X1B), keratomalacia (X3B), ocular lesions (Agrawal and Sahu, 2020) (stages as designated by World Health Organization grading system (World Health, 2014)Ocular lesions (Melo et al., 2004)

Ag-Ab, antigen–antibody; EE-µPAD, electronics enabled microfluidic paper-based analytical device; HYPER, High-yield paper-based quantitative blood separation system; RBP, retinol-binding protein; REI, rapid enzyme immunoassay.

h cd/m2 is the SI unit of luminance (Congdon et al., 2000).

Adapted from reference (Tanumihardjo et al., 2016).

Whereas serum or plasma is required to measure circulating vitamin A, some devices can use whole blood as the sample input.

Defined by global standards (e.g., World Health Organization) or by study authors.

An earlier version of this device is referenced as iCheck Ret 435–1 (Bechir et al., 2012).

Correlated with retinol ≤ 0.70 µmol/L, when sandwich ELISA is used for RBP measurement (Erhardt et al., 2004).

An earlier version of this device is referenced as iCheck Ret 515–2 (Bechir et al., 2012).

Device also referenced as “dark adaptometer” (Banerjee, 2019) or “In-Direct method and system for Vitamin A deficiency detection” (Mehta, 2019b, Mehta, 2019a, Mehta and Mehta, 2018).

Definitions of vitamin A deficiency, by sample type and device. Ag-Ab, antigen–antibody; EE-µPAD, electronics enabled microfluidic paper-based analytical device; HYPER, High-yield paper-based quantitative blood separation system; RBP, retinol-binding protein; REI, rapid enzyme immunoassay. h cd/m2 is the SI unit of luminance (Congdon et al., 2000). Adapted from reference (Tanumihardjo et al., 2016). Whereas serum or plasma is required to measure circulating vitamin A, some devices can use whole blood as the sample input. Defined by global standards (e.g., World Health Organization) or by study authors. An earlier version of this device is referenced as iCheck Ret 435–1 (Bechir et al., 2012). Correlated with retinol ≤ 0.70 µmol/L, when sandwich ELISA is used for RBP measurement (Erhardt et al., 2004). An earlier version of this device is referenced as iCheck Ret 515–2 (Bechir et al., 2012). Device also referenced as “dark adaptometer” (Banerjee, 2019) or “In-Direct method and system for Vitamin A deficiency detection” (Mehta, 2019b, Mehta, 2019a, Mehta and Mehta, 2018). Vitamin A liver concentration (µmol vitamin A/g liver) is the gold standard for vitamin A status but requires invasive techniques such as biopsy to be measured (Tanumihardjo et al., 2016). Sampling blood enables the quantification of serum retinol, serum retinol-binding protein (RBP), or provitamin A in the form of beta-carotene; however, each measure has trade-offs. Serum retinol reflects liver stores only at extremes of deficiency (≤0.07 µmol/g liver) or elevation (>1.05 µmol/g liver) (WHO, 2011), because serum retinol is homeostatically regulated by the body. The World Health Organization defines VAD as serum retinol ≤ 0.70 µmol/L (WHO, 2011). RBP is commonly assumed to have a 1:1 ratio with serum retinol, and therefore the same cut-offs are sometimes used for both retinol and RBP. However, this ratio can be affected by the extent of VAD, zinc deficiency, acute phase response, protein-energy malnutrition, liver disease, acutely stressful situations, high fever, antibiotic use, or obesity (Tanumihardjo et al., 2016, de Pee and Dary, 2002). Therefore, previous studies have proposed other deficiency cut-offs, such as 0.69 µmol/L (Semba et al., 2002) or 0.83 µmol/L (Engle-Stone et al., 2011, Gorstein et al., 2008). Recently, the Global Alliance for Vitamin A has recommended analysis of a subsample by HPLC to confirm the cut-off point for VAD; furthermore, given the acute phase response, inflammation markers such as C-reactive protein and alpha-1-acid-glycoprotein must also be measured (Global Alliance for Vitamin A, 2019). Beta carotene is one of several dietary provitamin A carotenoids, a plant-derived form of vitamin A. The body converts dietary provitamin A carotenoids into retinol with the following conversion factors: 1 µg retinol activity equivalent (RAE) equals 1 retinol equivalent (RE), 1 µg retinol, 2 µg β-carotene in oil, 12 µg β-carotene in mixed foods, or 24 (12–26) µg other provitamin A carotenoids in mixed foods (Institute of Medicine, 2001, Combs and McClung, 2017, Blaner, 2020). The conversion efficiency ratio of beta carotene to RAE is still debated. For example, the European Food Safety Authority suggests that the conversion is 6:1 rather than 12:1 (EFSA Panel on Dietetic Products, Nutrition Allergies, 2015). Carotenoids can be measured in blood, milk, or skin, and several studies have found a positive association of skin carotenoid concentrations with serum or plasma carotenoid status (Zidichouski et al., 2009, Aguilar et al., 2014, Morgan et al., 2019, Hayashi et al., 2020). However, a consensus has not been reached regarding a conversion factor or how the measurements equate to vitamin A status (von Lintig, 2020). Because carotenoids tend to reflect recent dietary intake rather than long-term status, recommended serum carotenoid deficiency cut-offs have not been established in humans (von Lintig, 2020). Deficiency in β-carotene in cow’s blood has been defined as 0.6–1.5 mg/L (Klein et al., 2013, De Ondarza and Al, 2009, Schweigert and Immig, 2007). In breast milk, retinol may be measured to estimate both the maternal vitamin A status and intake, and the infant intake of vitamin A (Engle-Stone et al., 2014, Tanumihardjo et al., 2016). Additionally, breast milk retinol measurement is influenced by the stage of lactation, time of day, “fullness” of the breast, feeding status if milk from both breasts is analyzed, and whether the milk is hindmilk compared with foremilk (Tanumihardjo et al., 2016). VAD is defined as a milk retinol concentration ≤ 1.05 µmol/L, or ≤ 8 µg/g milk fat (Tanumihardjo et al., 2016). In cows, milk β-carotene levels are often measured and linked to bovine fertility and health. Because of vitamin A’s role in in producing rhodopsin, the visual pigment of rods in the eyes, VAD can cause ocular manifestations resulting in poor vision (World Health, 2014). These include night blindness, conjunctival xerosis, Bitot’s spots, corneal xerosis, and keratomalacia. Impaired adaption to the dark is among the first symptoms of VAD, and it can be used as a screening tool (World Health, 2014). Tests such as pupillary and visual thresholds can assess dark adaptation by determining the lowest-intensity level of light required to cause pupillary dilation or to visualize an image (World Health, 2014, Labrique et al., 2015).

Comparison studies

From 3230 studies (after de-duplication), we identified 18 studies (19 reports) comparing nine portable methods/devices (index 1) to a reference standard method (Fig. 1); we were unable to retrieve an additional two reports (Craft, 2005, Fujita, 2007). No studies compared two portable methods (i.e., index 1 vs. index 2). Thirteen studies (15 reports) measured human or cattle blood samples (BioAnalyt, 2020, Chaimongkol et al., 2011, Ciaiolo et al., 2015, Elom et al., 2015, Ghaffari et al., 2019, Hix et al., 2004, Hix et al., 2006, Lee et al., 2016, Lu et al., 2017, Lu et al., 2018, Raila et al., 2012, Raila et al., 2017, Schweigert et al., 2011a, Chaimongkol et al., 2008, Lu and Erickson, 2017); four studies measured human or cattle milk samples (Schweigert et al., 2011b, Schweigert et al., 2011a, Engle-Stone et al., 2014, Abebe et al., 2019); and one study measured eye function (Peters et al., 2000). Study details are listed in Table 3. We also re-analyzed the data presented in supplemental Tables S1 and S2 in one publication to calculate the descriptive statistics for plasma and whole blood retinol in samples analyzed by HPLC and iCheck Fluoro (Raila et al., 2017).
Table 3

Description of included studies comparing a portable method against a reference standard method.

AuthorYearDeviceManufacturerSample testedBiomarkerStudy populationTest location (field/laboratory, country)Reference methodRef.
Chaimongkol 2011CRAFTiEurofin Craft TechnologiesSerumRetinolStudy cohortsThailandHPLC(Chaimongkol et al., 2011)
Chaimongkol 2008aCRAFTiEurofin Craft TechnologiesSerumRetinolStudy cohortsThailandHPLC(Chaimongkol et al., 2008)
Ciaiolo 2015Custom Ab-Ag reactionCustomSerum,RBPPatientsItalyNephelometry(Ciaiolo et al., 2015)
Lee 2016EE-µPADCustomSerumRBPCommercial (ProMedDx)USAELISA(Lee et al., 2016)
BioAnalyt report (year: NR)iCheck CaroteneBioAnalytPlasmaDairy cows and calvesNRHPLC(BioAnalyt, NR)
Raila 2012iCheck CaroteneBioAnalytWhole blood or plasmaBeta-caroteneHolstein-Friesian cows, local farmGermany, Ireland, FranceHPLC(Raila et al., 2012)
Ghaffari 2019iCheck CaroteneBioAnalytPlasmaBeta-caroteneHolstein cows and calves from institutional farmsGermanyHPLC(Ghaffari et al., 2019)
iCheck FluoroBioAnalytWhole bloodRetinol
Raila 2017iCheck FluoroBioAnalytWhole blood, serumRetinolDairy cows and bulls, institutional farmsGermany, JapanHPLC(Raila et al., 2017)
Schweigert 2011aiCheck FluoroBioAnalytMilkRetinolStudy cohorts and local cowsGermanyHPLC(Schweigert et al., 2011b)
Schweigert 2011baiCheck FluoroBioAnalytPlasma or milkRetinolStudy cohortsLow resource settingHPLC(Schweigert et al., 2011a)
Abebe 2019ICheck FluoroBioAnalytMilkRetinolStudy cohortsEthiopiaHPLC(Abebe et al., 2019)
Elom 2015iCheck FluoroBioAnalytSerumRetinolStudy cohortsMoroccoHPLC(Elom et al., 2015)
Engle-Stone 2014iCheck FluoroBioAnalytMilkRetinolStudy cohortsCameroonHPLC(Engle-Stone et al., 2014)
Hix 2004RBP-EIAScimedx CorpSerumRBPStudy cohorts, commercialPapua New Guinea, NicaraguaHPLC(Hix et al., 2004)
RIDThe Binding SiteStudy cohortsNicaraguaHPLC
Hix 2006RBP-EIAScimedx CorpSerumRBPStudy cohortsCambodiaHPLC(Hix et al., 2006)
Peters 2000SST-1LKC TechnologiesEyesDark-adapted final thresholds; rod functionPatients of Retina Foundation of the SouthwestUSAGoldmann-Weekers Dark Adaptometer(Peters et al., 2000)
Lu 2017Tidbit, ± HYPER filtration systemCustomSerumRBPCommercial (Research Blood Components LLC)USAELISA(Lu and Erickson, 2017, Lu et al., 2017, Lu et al., 2018)

Notes: Ag-Ab, antigen–antibody; EE-µPAD, electronics enabled microfluidic paper-based analytical device; EIA, enzyme immunoassay; HYPER, high-yield paper-based quantitative blood separation system; RBP, retinol-binding protein; RID, radial immunodiffusion assay; SST-1, Scotopic Sensitivity Tester-1.

a Meeting abstract, therefore some details are not reported.

b RID may be considered a second index test, because it is not a reference standard; however, in the study, only the first index test, RBP-EIA was the assay undergoing development and validation.

Description of included studies comparing a portable method against a reference standard method. Notes: Ag-Ab, antigen–antibody; EE-µPAD, electronics enabled microfluidic paper-based analytical device; EIA, enzyme immunoassay; HYPER, high-yield paper-based quantitative blood separation system; RBP, retinol-binding protein; RID, radial immunodiffusion assay; SST-1, Scotopic Sensitivity Tester-1. a Meeting abstract, therefore some details are not reported. b RID may be considered a second index test, because it is not a reference standard; however, in the study, only the first index test, RBP-EIA was the assay undergoing development and validation. We also identified many studies that used a portable device for assaying samples but did not compare the results to those of a reference method and instead cited previous validation studies. Although the devices used are catalogued and described (Table 1a, Table 1b), these studies are not further detailed in this review. Study populations were mostly from the US and Germany, in addition to Thailand, Italy, France Ireland, Japan, Ethiopia, Morocco, Cameroon, Papua New Guinea, Nicaragua, Cambodia, and Oman. Portable fluorometers, photometers, enzyme-based assays or immunoassays, microfluidics-based approaches, and a dark adaptometer for eye function were assessed and compared with their respective reference standards. Table 4 compares the stated performance criteria described by the device manufacturers’ websites to reporting from individual studies using the devices, according to the WHO Affordable, Sensitive, Specific, User-friendly, Rapid and robust, Equipment-free and Deliverable to end-users (ASSURED) criteria for diagnostic tests in resource-limited settings (Kosack et al., 2017). Only the iCheck Fluoro and iCheck Carotene stated performance criteria on the BioAnalyt website and are included in this table. No studies described the cost of the devices. Most devices did not have published cost information, aside from some of the slit lamps showing list prices. Other devices described by the studies were developed as proofs of concept and are either not on the market or are on the market, but lacking performance criteria on the manufacturer’s website.
Table 4

Assessment of devices against manufacturer-reported performance, according to ASSURED* criteria. (Ghaffari et al., 2019, Raila et al., 2017, Schweigert et al., 2011b, Schweigert et al., 2011a, Abebe et al., 2019, Elom et al., 2015, Engle-Stone et al., 2015, Raila et al., 2012, Ghaffari et al., 2019.)

Assessment of devices against manufacturer-reported performance, according to ASSURED* criteria. (Ghaffari et al., 2019, Raila et al., 2017, Schweigert et al., 2011b, Schweigert et al., 2011a, Abebe et al., 2019, Elom et al., 2015, Engle-Stone et al., 2015, Raila et al., 2012, Ghaffari et al., 2019.) We note that the study by Ghaffari et al. (2019) reported both measurements of retinol in whole blood and beta carotene in plasma, but only directly references the iCheck Fluoro device and not the iCheck Carotene (Ghaffari et al., 2019). Whereas the iCheck Carotene requires 0.4 mL of sample, the iCheck Fluoro requires 0.5 mL; the authors stated using only 0.5 mL of sample. Additionally, the manufacturer (BioAnalyt) lists only “colostrum, cattle whole blood, and serum” as appropriate sample types for the iCheck Carotene; however all studies using the iCheck Carotene, including a BioAnalyt report, also analyzed plasma for beta carotene content (BioAnalyt, 2020, Ghaffari et al., 2019, Raila et al., 2012). A major gap across all devices is the lack of reporting on sensitivity and specificity compared with a reference standard method. Most studies compared a portable device to a reference method. Table 5a, Table 5b, Table 5c, Table 5d, Table 5e, Table 5f, Table 5g show the performance of these devices against their reference standards for measuring vitamin A and VAD. Additional analyses conducted with other index (e.g., index 2) tests are described in the text.
Table 5a

Portable fluorometers: device performance in human blood samples.

Portable device vs. referenceiCheck Fluoro vs. HPLCiCheck Fluoro vs. HPLCCRAFTi vs. HPLCCRAFTi vs. HPLC
Vitamin A biomarkerRetinolaVitamin AaRetinolaRetinola
Sample typePlasmaSerumSerumSerum
Study population89 children56 samples38 women75 women, 143 children
Concentration differenceMD: 0 min: 1.9 µg/L ± 23.2MD: 15 min: −8.0 µg/L + 22.7NRMD: −0.07MD: −0.07
Correlation coefficient0 min: 0.9815 min: 0.98NRNR0.77
R2NR>0.95NRNR
Regression equationNRNRSlope = 0.81NR
Operational rangeNRNRNR0.5–1.5 µmol/L
VAD or VAI (%), index vs. refNot defined: N = 2/89 vs. NRNRNR<0.7 µmol/L: 9.2% vs. 2.8%<1.05 µmol/L: 49.5% vs. 43.6%
Precision
SensitivityNRNRNRVAD ≤ 0.7 µmol/L: 66.7%VAD ≤ 1.05 µmol/L: 85.3%
SpecificityNRNRNRVAD ≤ 0.7 µmol/L: 92.4%VAD ≤ 1.05 µmol/L: 78.0%
Intra-assay %CVNR2.5–6.4 %bAgreement noted but not quantified3.97% vs. 3.45%
Inter-assay %CVNRNR
Inter-observer %CVNRNR
Bland Altman analysis commentsNo commentary. At 0 min, 3 values fell outside 2 SDs. At 15 min, 4 values fell outside 2 SDsNRNo systematic biasNo systematic bias; most values within ± 0.5 with normally distributed serum retinol values
Reference(Elom et al., 2015)(Schweigert et al., 2011a)c(Chaimongkol et al., 2008)c(Chaimongkol et al., 2011)

MD, mean difference; NR, not reported; RE, retinol equivalents defined as the sum of retinol and retinyl esters, equal to 3.3 International Units (IU) of vitamin A or as 1 µg (units reported by manufacturer—however, retinol activity equivalents (RAE) are the preferred unit for reporting (Institute of medicine, 2001); SD, standard deviation; VAD, vitamin A deficiency; VAI, vitamin A insufficiency.

Units: µg/L or µmol/L.

Specific %CVs not distinguished.

Study abstract, lacking some details.

Table 5b

Portable fluorometers: device performance in human and bovine milk samples.

Human milk
Bovine milk
Portable device vs. referenceiCheck Fluoro vs. HPLCiCheck Fluoro vs. HPLCiCheck Fluoro vs. HPLCiCheck Fluoro vs. HPLC
Vitamin A biomarkerRetinola, milk fatbRetinola, milk fatbRetinolaRetinola
Study population104 women75 women, 154 samples1 woman, 16 samples21 cows
Concentration differenceMD: 0.01 µmol/L, 0.03 µg/g fatMD: −0.83 ± 0.14 µmol/L,−5.6 ± 0.7 µg/g fatExpressed milk, MD: 103% ± 13Expressed milk, MD: 105% ± 9Powdered milk (n = 5), MD: 144% ± 15Liquid whole milk (n = 5), MD: 118% ± 13Liquid skim milk (n = 4), MD: 95% ± 10
Correlation coefficient0.57unadj, 0.59adjc0.85unadj, 0.79adjcNRNR
R20.32unadj, 0.35adjc0.72unadj, 0.62adjcNRNR
Regression equationNRNRNRNR
Operational range50–3000 µg RE/L50–3000 µg RE/L50–3000 µg RE/LdNR
VAD or VAI (%), index vs. ref<1.05 µmol/L: 87% vs. 76%<8 µg/g fat%: 89% vs. 81%<1.05 µmol/L: 3.9% vs. 2.60%<8 µg/g fat %: 0% vs. 2%NRNR
Precision
SensitivityNRToo few VAD cases to examineNRNR
SpecificityNRToo few VAD cases to examineNRNR
Intra-assay %CV1.1% vs. 1.5–1.6%0.6 %eNRNR
Inter-assay %CVNRNRNR
Inter-observer %CVNRNRNR
Bland Altman analysis commentsUsed to present mean difference between measurements; mean difference not significantly different from zeroPlotted but no conclusion drawn; appears to show 8 values outside of 2 SDs (µmol/L retinol) and 8 values outside of 2 SDs (µg/g fat)NRNR
Reference(Abebe et al., 2019)(Engle-Stone et al., 2014)(Schweigert et al., 2011b)(Schweigert et al., 2011b)

MD, mean difference; NR, not reported; RE, retinol equivalents defined as the sum of retinol and retinyl esters, equal to 3.3 International Units (IU) of vitamin A or as 1 µg (units reported by manufacturer—however, retinol activity equivalents (RAE) are the preferred unit for reporting (Institute of medicine, 2001); VAD, vitamin A deficiency; VAI, vitamin A insufficiency.

Units: µmol/L or µg RE/L.

Units: µg/g fat%.

Adjusted for breast milk fat content.

Not reported but based on previous studies using same device.

Specific %CVs not distinguished.

Table 5c

Portable immunoassays: device performance in human blood.

Portable device vs. referenceRBP-REI vs. HPLCaRBP-REI vs. HPLCaRID vs. HPLCaImmunoassay vs. nephelometry
Vitamin A biomarkerRBPbRBPbRBPbRBPbRBPb
Sample typeSerumSerumSerumSerumSerum
Study population24 children70 mothers and children359 children40 mothers and children2 healthy adults(Serum A, B)
Concentration differenceMD: 0.22 µmol/LNRNRNRIndex (dilution):
Serum A:(1:10): +(1:100): -(1:1000): -(1:10000): -Serum B:(1:10): ++(1:100): -(1:1000): -(1:10000): -

Ref:Serum A: 46 mg/LRef:Serum B: 42 mg/L
Correlation coefficient0.930.910.890.84NR
R20.860.820.790.71NR
Regression equationy = 0.95x + 0.36y = 0.62x + 0.32y = 0.65x + 0.27NRNR
Operational range10–40 µg RBP/mL10–40 µg RBP/mL10–40 µg RBP/mLcNRImmune precipitates:Neg: -2 mg/L: +10 mg/L: ++100 mg/L: +++1000–10000 mg/L: ++++
VAD or VAI (%), index vs. ref≤0.70 µmol/L: 32% vs. 36%NR<0.35 µmol/L: 0.6% vs. 2.2%≤0.70 µmol/L: 20.9% vs. 22.3%NRNR
Precision
SensitivityNRNR70%NR“Good” (“can detect presence of [RBP] at concentration of few µg/mL”)
SpecificityNRNR93.2%NRNR
Intra-assay %CV6.7 %dNRNRNRNR
Inter-assay %CV8.9 %dNRNRNRNR
Inter-observer %CV13.0 %dNRNRNRNR
Bland Altman analysis commentsNRNRNRNRNR
Reference(Hix et al., 2004)(Hix et al., 2006)(Hix et al., 2004)(Ciaiolo et al., 2015)

MD, mean difference; NR, not reported; RBP, retinol binding protein; RE, retinol equivalents defined as the sum of retinol and retinyl esters, equal to 3.3 International Units (IU) of vitamin A or as 1 µg (units reported by manufacturer—however, retinol activity equivalents (RAE) are the preferred unit for reporting (Institute of Medicine, 2001); REI, rapid enzyme immunoassay; VAD, vitamin A deficiency; VAI, vitamin A insufficiency.

Reference analyte is retinol.

Units: µmol/L or µg RE/L.

Not reported but based on previous studies using same device.

Reported from separate analysis among unknown total # of samples (“5 adult volunteers and a commercially available source”) analyzing device performance, without reference to HPLC.

Table 5d

Portable microfluidics-based methods: device performance in human blood.

Portable device vs. referenceEE-µPAD, vs. ELISATidbit with HYPER platform, vs. ELISAaTidbit without HYPER platform, vs. ELISA
Vitamin A biomarkerfRBPbRBPbRBPb
Sample typeWhole bloodWhole bloodSerum
Study population95 adults (commercial)12 adults43 adults (commercial)
Concentration differenceNRNRNR
Correlation coefficientNRNR0.75
R2 (index, unless specified)NRIndex: 0.81 vs. ref: >0.990.56
Regression equationNRSlope = 0.99Slope = 0.97
RMSE, index vs. refNR3.75 vs. 1.3 µg/mL4.34 µg/mL vs. NR
Operational range∼10–70 µg/mL (graph)∼5–20 µg/mL (graph)2.2–20 µg/mL (0.10–0.95 µmol/L)
VAD or VAI (%), index vs. ref<16.3 µg/mL: AUC = 0.7139 vs. 17.2%NR<14.7 µg/mL (≤0.70 µmol/L): NR vs. 9.3%
Precision
Sensitivity75% at MFR cutoff, 0.831NR100%
Specificity62.3% at MFR cutoff, 0.831NR100%
Intra-assay %CV10.8% vs. 3.9%20.3% deviation per test strip, recommend taking average of 3 test stripsNR
Inter-assay %CVNRNRNR
Inter-observer %CVNRNRNR
Bland Altman analysis commentsNRNRBias at −0.05 µg/mL (-2.3 nmol/L)
Reference(Lee et al., 2016)(Lu and Erickson, 2017, Lu et al., 2017, Lu et al., 2018)(Lu and Erickson, 2017, Lu et al., 2017)

MD, mean difference; MFR, multi-faceted ratio i.e., the ratio of the light transmission in the test area to that in the background control area, calculated for RBP for each sample repeat. NR, not reported; RE, retinol equivalents defined as the sum of retinol and retinyl esters, equal to 3.3 International Units (IU) of vitamin A or as 1 µg (units reported by manufacturer—however, retinol activity equivalents (RAE) are the preferred unit for reporting (Institute of medicine, 2001); RMSE, root mean squared error; VAD, vitamin A deficiency; VAI, vitamin A insufficiency.

Reference ELISA utilized samples that were filtered using HYPER system.

Units: µg/mL, mg/L, or µmol/L.

Table 5e

Other portable devices: device performance in for assessing eye function (vision).

Eyes
Index 1avs. referenceb,c
Validation of portable deviceSST-1 vs. Goldmann-Weekers dark adaptometera
Vitamin A biomarkerDark adaptation final thresholdb
Study population87 patientsc and 24 healthy children and adults
Concentration differenceNR
Correlation coefficient0.88 (adjusted for ceiling effect)
R2 (index, unless specified)0.77
Regression equation“intercept close to zero”
Operational range0–30 dB stimulus intensity range (0–3 log units)
VAD or VAI (%), index vs. refElevated final thresholds: 75% vs. 82%
Precision
SensitivityFinal threshold elevated: 74.7%
SpecificityNR
Intra-assay %CVNR
Inter-assay %CVNR
Inter-observer %CVNR
Bland Altman analysis commentsNR
Reference(Peters et al., 2000)

MD, mean difference; MFR, multi-faceted ratio i.e., the ratio of the light transmission in the test area to that in the background control area, calculated for RBP for each sample repeat. NR, not reported; RE, retinol equivalents defined as the sum of retinol and retinyl esters, equal to 3.3 International Units (IU) of vitamin A or as 1 µg (units reported by manufacturer—however, retinol activity equivalents (RAE) are the preferred unit for reporting (Institute of medicine, 2001); SST-1, Scotopic Sensitivity Tester-1; VAD, vitamin A deficiency; VAI, vitamin A insufficiency.

Reference analyte is dark adaptation final threshold.

Units: log units.

Patients had retinal degeneration with mild to severe loss of rod function from full-field ERG results.

Table 5f

Portable fluorometers: device performance in bovine blood samples.

Portable device vs. referenceiCheck Fluoro vs. HPLCaiCheck Fluoro vs. HPLCaiCheck Fluoro vs. HPLCaiCheck Fluoro vs. HPLCaiCheck Fluoro vs. HPLCb
Vitamin A biomarkerVitamin AcVitamin AcRetinoldVitamin AcVitamin AcRetinoldRetinoldRetinoldRetinold
Sample typeWhole bloodWhole bloodWhole bloodPlasmaPlasmaPlasmaPlasmaSerumSerum
Study population28 cows11 calves10 cows28 cows11 calves40 cows92 bulls29 cows32 black cattle
Concentration differenceRange: 184–336(see note)eMD: −0.013 ± -0.020MD: 19.3MD: 26.5MD: 0.01MD: 0.00
Correlation coefficient0.780.900.920.880.960.940.93
R20.610.810.840.770.920.88g0.87
0.880.94
Regression equationy = 0.77 + 11.26NRy = 1.18x − 72.64y = 0.80x + 1.32NRNR
y = 1.03–30.11
Operational rangeNRNRNRNRNR
VAD or VAI (%)NRNRNRNRNRNRNRNRNR
Precision
SensitivityNRNRStates test is sensitive and specific but does not quantify theseNRNRStates test is sensitive and specific but does not quantify theseStates test is sensitive and specific but does not quantify theseStates test is sensitive and specific but does not quantify theseStates test is sensitive and specific but does not quantify these
SpecificityNRNRNRNR
Intra-assay %CVNRNRNRNRNR2.3% vs. 5.3 %f2.1% vs. 3.3 %f
Inter-assay %CVNRNRNRNRNR
Inter-observer %CVNRNRNRNRNRNRNRNRNR
Bland-Altman analysis commentsGood level of agreement and no systematic error. 5% of total values fell outside 95% acceptability limitsGood level of agreement and no systematic error; 4% of total values fell outside the 95% acceptability limitsGood level of agreement and no systematic error. 1 value fell outside of 95% acceptability limitsGood level of agreement and no systematic error. 1 value fell outside of 95% acceptability limitsGood level of agreement and no systematic error. 4% of values fell outside the 95% acceptability limits
Reference(Ghaffari et al., 2019)(Raila et al., 2017)(Ghaffari et al., 2019)(Raila et al., 2017)

MD, mean difference; NR, not reported; RE, retinol equivalents defined as the sum of retinol and retinyl esters, equal to 3.3 International Units (IU) of vitamin A or as 1 µg (units reported by manufacturer—however, retinol activity equivalents (RAE) are the preferred unit for reporting (Institute of medicine, 2001); VAD, vitamin A deficiency; VAI, vitamin A insufficiency.

Reference sample is in plasma.

Reference sample is serum.

Units: µg RE/L.

Units: µmol/L.

Reported value from this study appears to be a repeated value for cow whole blood beta carotene content given as 2.09–8.15 mg/L, instead of the calf whole blood vitamin A reported in µg RE/L.

Values appear to be an average of intra- and inter-assay %CV.

Table 5g

Portable photometers: device performance in bovine blood samples.

Portable device vs. referenceiCheck Carotene vs. HPLCaiCheck Carotene vs. HPLCaiCheck Carotene vs. HPLCaiCheck Carotene vs. HPLCaiCheck Carotene vs. HPLCaiCheck Carotene vs. HPLCaiCheck Carotene vs. HPLCa
Vitamin A biomarkerβ-carotenebβ-carotenebβ-carotenebβ-carotenebβ-carotenebβ-carotenebβ-caroteneb
Sample typeWhole bloodWhole bloodWhole bloodPlasmaPlasmaPlasmaPlasma
Study population28 cows11 calves23 cows28 cows11 calvesNR, cows and calves166 cows
Concentration differenceNRNRMD: 0.21MD: −0.29MD: 0.02NRMD: 0.26
Correlation coefficient0.980.980.990.970.98NR0.99
R20.970.960.990.930.96c0.970.98
0.99c0.98d
Regression equationy = 1.01x + 0.17cNRy = 0.88x + 0.31ey = 0.97x + 0.40ey = 1.05x + 0.04ey = 0.90x + 0.04ey = 0.90x + 0.17y = 0.98x + 0.31
Operational rangeNR0.4–18 mg/LNR∼0–9 mg/L (graph)0.4–18 mg/L
VAD (%)NRNRNRNRNRNRNR
Precision
SensitivityNRNRNRNRNRNRNR
SpecificityNRNRNRNRNRNRNR
Intra-assay %CVNRNR3.5% vs. 2.3 %fNRNRNR3.5% vs. 2.3 %f
Inter-assay %CVNRNRNRNRNRNRNR
Inter-observer %CVNRNRNRNRNRNRNR
Bland Altman analysis commentsA good level of agreement and no systematic error for β-carotene and vitamin A; “only 5% of the differences in measured values fell outside the 95% acceptability limits for β-carotene in dairy cows”Systematic error did not occur between methods:4% of differences outside 95% limitsA good level of agreement and no systematic error for β-carotene and vitamin A; “only 5% of the differences in measured values fell outside the 95% acceptability limits for β-carotene in dairy cows”Graph presented, no comment (appears to have good agreement)Systematic error did not occur between methods:4% of differences outside 95% limits
Reference(Ghaffari et al., 2019)(Raila et al., 2012)(Ghaffari et al., 2019)(BioAnalyt, NR)(Raila et al., 2012)

Notes: MD, mean difference; NR, not reported; RE, retinol equivalents defined as the sum of retinol and retinyl esters, equal to 3.3 International Units (IU) of vitamin A or as 1 µg (units reported by manufacturer—however, retinol activity equivalents (RAE) are the preferred unit for reporting (Institute of Medicine, 2001).

Reference sample is in plasma.

Units: mg/L.

Average from reference analyses done in Germany and Switzerland.

Reference analysis done in Germany.

Reference analysis done in Switzerland.

Appears to represent the average CV for both whole blood and plasma samples.

Portable fluorometers: device performance in human blood samples. MD, mean difference; NR, not reported; RE, retinol equivalents defined as the sum of retinol and retinyl esters, equal to 3.3 International Units (IU) of vitamin A or as 1 µg (units reported by manufacturer—however, retinol activity equivalents (RAE) are the preferred unit for reporting (Institute of medicine, 2001); SD, standard deviation; VAD, vitamin A deficiency; VAI, vitamin A insufficiency. Units: µg/L or µmol/L. Specific %CVs not distinguished. Study abstract, lacking some details. Portable fluorometers: device performance in human and bovine milk samples. MD, mean difference; NR, not reported; RE, retinol equivalents defined as the sum of retinol and retinyl esters, equal to 3.3 International Units (IU) of vitamin A or as 1 µg (units reported by manufacturer—however, retinol activity equivalents (RAE) are the preferred unit for reporting (Institute of medicine, 2001); VAD, vitamin A deficiency; VAI, vitamin A insufficiency. Units: µmol/L or µg RE/L. Units: µg/g fat%. Adjusted for breast milk fat content. Not reported but based on previous studies using same device. Specific %CVs not distinguished. Portable immunoassays: device performance in human blood. MD, mean difference; NR, not reported; RBP, retinol binding protein; RE, retinol equivalents defined as the sum of retinol and retinyl esters, equal to 3.3 International Units (IU) of vitamin A or as 1 µg (units reported by manufacturer—however, retinol activity equivalents (RAE) are the preferred unit for reporting (Institute of Medicine, 2001); REI, rapid enzyme immunoassay; VAD, vitamin A deficiency; VAI, vitamin A insufficiency. Reference analyte is retinol. Units: µmol/L or µg RE/L. Not reported but based on previous studies using same device. Reported from separate analysis among unknown total # of samples (“5 adult volunteers and a commercially available source”) analyzing device performance, without reference to HPLC. Portable microfluidics-based methods: device performance in human blood. MD, mean difference; MFR, multi-faceted ratio i.e., the ratio of the light transmission in the test area to that in the background control area, calculated for RBP for each sample repeat. NR, not reported; RE, retinol equivalents defined as the sum of retinol and retinyl esters, equal to 3.3 International Units (IU) of vitamin A or as 1 µg (units reported by manufacturer—however, retinol activity equivalents (RAE) are the preferred unit for reporting (Institute of medicine, 2001); RMSE, root mean squared error; VAD, vitamin A deficiency; VAI, vitamin A insufficiency. Reference ELISA utilized samples that were filtered using HYPER system. Units: µg/mL, mg/L, or µmol/L. Other portable devices: device performance in for assessing eye function (vision). MD, mean difference; MFR, multi-faceted ratio i.e., the ratio of the light transmission in the test area to that in the background control area, calculated for RBP for each sample repeat. NR, not reported; RE, retinol equivalents defined as the sum of retinol and retinyl esters, equal to 3.3 International Units (IU) of vitamin A or as 1 µg (units reported by manufacturer—however, retinol activity equivalents (RAE) are the preferred unit for reporting (Institute of medicine, 2001); SST-1, Scotopic Sensitivity Tester-1; VAD, vitamin A deficiency; VAI, vitamin A insufficiency. Reference analyte is dark adaptation final threshold. Units: log units. Patients had retinal degeneration with mild to severe loss of rod function from full-field ERG results. Portable fluorometers: device performance in bovine blood samples. MD, mean difference; NR, not reported; RE, retinol equivalents defined as the sum of retinol and retinyl esters, equal to 3.3 International Units (IU) of vitamin A or as 1 µg (units reported by manufacturer—however, retinol activity equivalents (RAE) are the preferred unit for reporting (Institute of medicine, 2001); VAD, vitamin A deficiency; VAI, vitamin A insufficiency. Reference sample is in plasma. Reference sample is serum. Units: µg RE/L. Units: µmol/L. Reported value from this study appears to be a repeated value for cow whole blood beta carotene content given as 2.09–8.15 mg/L, instead of the calf whole blood vitamin A reported in µg RE/L. Values appear to be an average of intra- and inter-assay %CV. Portable photometers: device performance in bovine blood samples. Notes: MD, mean difference; NR, not reported; RE, retinol equivalents defined as the sum of retinol and retinyl esters, equal to 3.3 International Units (IU) of vitamin A or as 1 µg (units reported by manufacturer—however, retinol activity equivalents (RAE) are the preferred unit for reporting (Institute of Medicine, 2001). Reference sample is in plasma. Units: mg/L. Average from reference analyses done in Germany and Switzerland. Reference analysis done in Germany. Reference analysis done in Switzerland. Appears to represent the average CV for both whole blood and plasma samples. In human blood samples, both the iCheck Fluoro and the CRAFTi portable fluorometers were used to measure retinol (Table 5a). The iCheck Fluoro studies showed a high correlation (0.98) and an R-squared values over 0.95 with respect to HPLC. Both CRAFTi studies found a mean difference in serum retinol of −0.07 µmol/L, and the 2011 study found moderate sensitivity and specificity in identifying VAD at either ≤ 0.70 µmol/L or ≤ 1.05 µmol/L. Few additional comparative data were available between studies. Bias analysis indicated an acceptable level of agreement (within two SDs or 95% acceptability limits) between these devices’ performance and HPLC. Of note, we identified an additional report examining “vitamin A… in plasma” as measured by the CRAFTi compared with HPLC (Craft, 2005). The correlation between methods was 0.82. However, these data came from a summary of a poster submitted to a conference, and we were unable to find the full version of the poster; therefore, the report was excluded from our primary results and is not in Table 5a above. Four studies analyzed either human or cow’s milk samples with the iCheck Fluoro (Table 5b). The device performance varied: some studies reported lower, equivalent, or higher retinol values than those of HPLC. The R2 values for the correlation between the device and HPLC ranged between 0.35 and 0.79 after adjustment for milk fat content. In one study (Schweigert et al., 2011b), the authors tested increasingly diluted cow’s milk samples with 3.5% fat by using the iCheck, which showed linearity at an R2 of > 0.99 between 100 and 2500 µg RE/L. The same study also showed a positive correlation between percentage milk fat and µg RE/L milk with the iCheck Fluoro. Precision was tested over an operational range of 60 to 600 µg RE/L, and the inter-assay CV was < 3.5% (not shown in table). RBP was measured in blood with two field-friendly immunoassays reported across three studies comparing a portable device to a reference method (Table 5c). A rapid enzyme immunoassay (RBP-REI), available from Scimedx Corp, was able to detect serum RBP within a range of 10–40 µg/mL, which correlated with the HPLC results (R2 = 0.79 to 0.86) (Hix et al., 2004, Hix et al., 2006). The RBP-REI assay was also compared with another portable, laboratory-based device (index 2), a commercially available radial immunodiffusion plate reader (RID; The Binding Site, San Diego), by measuring RBP in 40 serum samples (Hix et al., 2004). Compared with the higher R2 values in validation against HPLC (R2 = 0.82 and 0.86; Table 5c), the RBP-REI had a lower, but still acceptable, correlation with the RID method (R2 = 0.73; linearity: y = 0.50x + 0.45) (not shown in table). Comparison of RID and HPLC indicated a slightly lower correlation (R2 = 0.71) (Table 5c). Other validity and precision data were not reported for the comparisons of REI vs. RID, or RID vs. HPLC. From the current manufacturer’s website (accessed date: March 15, 2021) (The Binding Site, 2020), RBP was not listed among the human proteins for assessment with the RID plate reader. A semi-quantitative antigen–antibody binding assay allowed for detection of low concentrations of RBP in serum samples (Ciaiolo et al., 2015). However, because only six samples were used for validation, drawing a conclusion regarding the efficacy of this method is difficult. We identified two microfluidics-based devices, the EE-µPAD (Lee et al., 2016) and the Tidbit with HYPER filtration (Lu and Erickson, 2017, Lu et al., 2017, Lu et al., 2018), both of which were able to separate whole blood into serum, detect VAD at high sensitivity and specificity with respect to the reference ELISA test, and send results to a mobile device (Table 5d). We note that the ELISA test may not be a suitable reference method for assessing VAD, owing to inherent problems with antibodies to RBP. Neither device is currently on the market. Although we identified several portable dark adaptometers (Table 1b), we found only one validation study between a portable dark adaptometer, the Scotopic Sensitivity-Tester 1 by LKC Technologies, and a reference standard, the Goldmann-Weekers dark adaptometer used in clinical settings (Peters et al., 2000) (Table 5e). The portable device was comparable to the reference standard in its sensitivity in identifying elevated final thresholds for dark adaptation, with a correlation (R2) of 0.77. However, this study was performed in the US in an eye clinic, and it remains to be tested and compared with the reference standard in field settings. The iCheck Fluoro, a portable fluorometer, was used to measure bovine blood samples for retinol (Table 5f). Compared with HPLC, mean differences in whole blood, plasma, or serum retinol ranged from −0.01 µmol/L to 26.5 µmol/L, and the iCheck generally displayed higher values than HPLC. The correlation between the iCheck Fluoro and HPLC was positive, ranging in R2 values from 0.61 to 0.96. Weaker correlations were observed in cows (range: 0.78–88) than calves (0.90–0.96). Raila and et al. (2017) also compared the correlation between bovine whole blood retinol (n = 10) and plasma retinol (n = 10), both measured by the index test iCheck Fluoro, and found a significant positive correlation (R2 = 0.87) (Raila et al., 2017). No studies reported sensitivity and specificity, or distinguished specific %CVs. Bias analysis indicated acceptable agreement between the device performance and HPLC. The iCheck Carotene, a portable photometer, was used to measure carotenoids in bovine whole blood and plasma. Mean differences in beta-carotene concentration ranged from −0.29 mg/L to 0.26 mg/L in plasma samples in cows and calves (Table 5g). The correlation between iCheck Carotene and HPLC was high, with R2 between 0.93 and 0.99. No studies reported sensitivity, specificity, or specific %CVs. Bias analysis revealed an acceptable level of agreement between the device performance and that of HPLC.

Future perspectives and recommendations

Gaps and recommendations

On the basis of our review of the literature, portable devices fell into five categories: Portable fluorometers Portable photometers Field-friendly immunoassays and/or microfluidics-based devices Slit lamps Dark adaptometers We found that, although many portable devices for quantifying vitamin A have been developed and described, only a few devices appear to be currently on the market or commercially available; of these, only two had easily accessible performance criteria information on the manufacturers’ websites related to vitamin A measurement. Studies tended not to report on portable device characteristics. Some major gaps involve the lack of data reported by studies. Few studies have reported the portable device’s sensitivity and specificity in detecting VAD compared with the reference standard method—a necessary metric for validation and adoption by randomized trials. Furthermore, only the iCheck devices were assessed in more than two studies; other devices should be analyzed further for validation.

Minimal set of criteria for point-of-need devices

See Fig. 2.
Fig. 2

Minimal set of criteria for point-of-need devices. Adapted with permission (Huey, 2022).

Minimal set of criteria for point-of-need devices. Adapted with permission (Huey, 2022). The device should: Be lightweight with a small form factor for easy transport to the necessary location as needed. Be standalone without needing additional equipment and self-powered, and should pre-store all the required reagents for the test, and use common reagents that are available on the market. Be easy to use with minimal processing steps in the protocol, and should require minimal training effort. Have analytical performance (e.g., %CV < 5% or within Bland Altman 95% limits of agreement) comparable to those of the current laboratory standards, with a capability to test various biological samples. Be affordable and capable of scaling up with locally available consumables where needed. Be able to connect to the internet or an external hard drive with a built-in data management system to allow the test results to be reliably stored and transferred. Be able to output test results quickly and present in a format that is easy to interpret.

Conclusions

In this review, we identified 25 portable methods or devices for a variety of biological sample types including those of human (blood, milk, and eye/vision) and animal (blood and milk) origin. These included nine methods measuring biochemical markers of vitamin A or VAD (serum retinol, RBP, milk retinol, retinyl palmitate, and retinyl esters) and 17 portable methods measuring functional biomarkers (measures of eye health, for example dark adaptation). The iCheck devices, including iCheck Carotene and iCheck Fluoro—for measuring total carotenoids or beta-carotene, or for measuring retinol, retinyl palmitate, retinyl acetate, or other esters, respectively, in blood or milk—were the only devices with manufacturer-reported performance metrics as well as the most information and data available to ascertain the method’s accuracy and precision with respect to those of a gold standard such as HPLC. These methods, in addition to the CRAFTi portable fluorometer, as compared with HPLC, were thus considered acceptable for measuring both blood and milk for biochemical biomarkers of vitamin A and detecting vitamin A deficiency. In measuring human or cow milk samples’ retinol concentration, the iCheck Fluoro had variable performance across studies, including both lower and higher values than the gold standard HPLC, thus leading to weaker correlation values than those calculated for blood samples. However, the mean differences were<1 µmol/L, and the values were considered to be within the expected variance. Correlation was improved by diluting the samples; dilution may be required for higher accuracy when the portable method is used. Several portable immunoassays (RBP-REI, RID, general immunoassay) and microfluidics-based methods (EE-µpad, TIDBIT with or without HYPER platform) for measuring RBP in human blood had acceptable correlations with HPLC reference methods and similar detection of VAD. However, these assays appeared not to be commercially available. One study has measured eye function with a portable dark adaptometer (Scotopic Sensitivity Tester-1), which had comparable results to the gold standard, a Goldmann-Weekers dark adaptometer. However, field studies using this device in comparison to a reference remain to be performed. Given the importance of eye health as a functional indicator of vitamin A deficiency, this gap in the literature is substantial. Finally, the iCheck Fluoro was used for measuring bovine blood samples for retinol. Generally, the retinol values were higher than those in samples tested by HPLC. Retinol measurements in calves appeared to have stronger correlations than retinol in cow’s blood. Several studies examined the accuracy of the iCheck Carotene, as compared with HPLC, in determining carotenoid content in cow’s blood. Strong correlations with acceptable levels of agreement were observed between device performance and HPLC performance. In summary, the iCheck devices are commercially available and are acceptable for measuring vitamin A in blood and milk, on the basis of the available data. Many of the other identified devices were proofs of concept and not yet commercially available. Several gaps remain, including studies comparing the other portable devices against a gold standard, particularly for functional indicators of vitamin A status/deficiency; available manufacturer-reported device performance criteria against which to compare the results of future investigations; and more comprehensive reporting of sensitivity, specificity, precision, and other validation metrics.

CRediT authorship contribution statement

Samantha L. Huey: Data curation, Formal analysis, Investigation, Methodology, Writing – original draft, Writing – review & editing. Jesse T. Krisher: Data curation, Formal analysis, Investigation, Methodology, Writing – original draft, Writing – review & editing. David Morgan: Project administration, Supervision, Writing – review & editing. Penjani Mkambula: Project administration, Writing – review & editing. Bryan M. Gannon: Conceptualization, Methodology, Writing – review & editing. Mduduzi N.N. Mbuya: Writing – review & editing. Saurabh Mehta: Conceptualization, Funding acquisition, Methodology, Supervision, Writing – review & editing.

Declaration of Competing Interest

The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
  45 in total

1.  Comparison of breast milk vitamin A concentration measured in fresh milk by a rapid field assay (the iCheck FLUORO) with standard measurement of stored milk by HPLC.

Authors:  R Engle-Stone; M J Haskell; M R La Frano; A O Ndjebayi; M Nankap; K H Brown
Journal:  Eur J Clin Nutr       Date:  2014-04-16       Impact factor: 4.016

2.  Responsiveness of dark-adaptation threshold to vitamin A and beta-carotene supplementation in pregnant and lactating women in Nepal.

Authors:  N G Congdon; M L Dreyfuss; P Christian; R C Navitsky; A M Sanchez; L S Wu; S K Khatry; M D Thapa; J Humphrey; D Hazelwood; K P West
Journal:  Am J Clin Nutr       Date:  2000-10       Impact factor: 7.045

3.  Estimating the Effective Coverage of Programs to Control Vitamin A Deficiency and Its Consequences Among Women and Young Children in Cameroon.

Authors:  Reina Engle-Stone; Martin Nankap; Alex O Ndjebayi; Stephen A Vosti; Kenneth H Brown
Journal:  Food Nutr Bull       Date:  2015-09       Impact factor: 2.069

4.  Development of a rapid enzyme immunoassay for the detection of retinol-binding protein.

Authors:  John Hix; Carolina Martinez; Ian Buchanan; Jeff Morgan; Milton Tam; Anuraj Shankar
Journal:  Am J Clin Nutr       Date:  2004-01       Impact factor: 7.045

5.  Physiologic indicators of vitamin A status.

Authors:  Nathan G Congdon; Keith P West
Journal:  J Nutr       Date:  2002-09       Impact factor: 4.798

6.  Feasibility of using retinol-binding protein from capillary blood specimens to estimate serum retinol concentrations and the prevalence of vitamin A deficiency in low-resource settings.

Authors:  Jonathan L Gorstein; Omar Dary; Bettina Shell-Duncan; Tim Quick; Emorn Wasanwisut
Journal:  Public Health Nutr       Date:  2008-05       Impact factor: 4.022

7.  Technical note: Rapid field test for the quantification of vitamin E, β-carotene, and vitamin A in whole blood and plasma of dairy cattle.

Authors:  Morteza H Ghaffari; Katrin Bernhöft; Stephane Etheve; Irmgard Immig; Michael Hölker; Helga Sauerwein; Florian J Schweigert
Journal:  J Dairy Sci       Date:  2019-10-16       Impact factor: 4.034

8.  A novel device for assessing dark adaptation in field settings.

Authors:  Alain B Labrique; Amanda C Palmer; Katherine Healy; Sucheta Mehra; Theodor C Sauer; Keith P West; Alfred Sommer
Journal:  BMC Ophthalmol       Date:  2015-07-09       Impact factor: 2.209

9.  Prevalence of ocular morbidities among school children in Raipur district, India.

Authors:  Deepanshu Agrawal; Anupam Sahu; Deepshikha Agrawal
Journal:  Indian J Ophthalmol       Date:  2020-02       Impact factor: 1.848

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.