Literature DB >> 30427951

Accuracy of Implant Placement with a Navigation System, a Laboratory Guide, and Freehand Drilling.

Chang-Kai Chen, Da-Yo Yuh, Ren-Yeong Huang, Earl Fu, Cheng-Feng Tsai, Cheng-Yang Chiang.   

Abstract

PURPOSE: Computer-aided surgery under navigation system guidance is widely applied in dental implant procedures. However, the accuracy of drilling with such navigation systems has not been comparatively evaluated alongside those of laboratory guide-based and freehand drilling. Therefore, this study aimed to compare the accuracies of these three drilling systems.
MATERIALS AND METHODS: A navigation system, a laboratory guide, and freehand drilling were used to drill 150 holes on 30 cast models. Two master models-one each for the maxilla and mandible-were prepared with the idea of placing five implants per cast. After drilling five holes on each cast, postoperative cone beam computed tomography images were acquired to measure the magnitude of errors.
RESULTS: The navigation system and laboratory guide were more accurate than freehand placement with respect to total errors at the entry and apex, lateral error at the apex, and angular error. The navigation system was more accurate than the laboratory guide with respect to angular error. Laboratory guide-based drilling was more accurate than freehand drilling in terms of lateral error at entry.
CONCLUSION: In comparison with the laboratory guide and freehand placement, the navigation system exhibited lower angular and axial errors. Despite its higher accuracy, the navigation system requires the operator to pay greater attention.

Entities:  

Mesh:

Substances:

Year:  2018        PMID: 30427951     DOI: 10.11607/jomi.6585

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  17 in total

Review 1.  Accuracy assessment of dynamic computer-aided implant placement: a systematic review and meta-analysis.

Authors:  Adrià Jorba-García; Albert González-Barnadas; Octavi Camps-Font; Rui Figueiredo; Eduard Valmaseda-Castellón
Journal:  Clin Oral Investig       Date:  2021-02-26       Impact factor: 3.573

2.  Accuracy and primary stability of tapered or straight implants placed into fresh extraction socket using dynamic navigation: a randomized controlled clinical trial.

Authors:  Shi-Min Wei; Jun-Yu Shi; Shi-Chong Qiao; Xiao Zhang; Hong-Chang Lai; Xiao-Meng Zhang
Journal:  Clin Oral Investig       Date:  2021-11-19       Impact factor: 3.573

3.  Accuracy of Computer-Aided Dynamic Navigation Compared to Computer-Aided Static Navigation for Dental Implant Placement: An In Vitro Study.

Authors:  Alfonso Mediavilla Guzmán; Elena Riad Deglow; Álvaro Zubizarreta-Macho; Rubén Agustín-Panadero; Sofía Hernández Montero
Journal:  J Clin Med       Date:  2019-12-02       Impact factor: 4.241

4.  The Workflow of a New Dynamic Navigation System for the Insertion of Dental Implants in the Rehabilitation of Edentulous Jaws: Report of Two Cases.

Authors:  Armando Lopes; Miguel de Araújo Nobre; Diogo Santos
Journal:  J Clin Med       Date:  2020-02-04       Impact factor: 4.241

Review 5.  Efficacy of Computer-Aided Static Navigation Technique on the Accuracy of Endodontic Microsurgery. A Systematic Review and Meta-Analysis.

Authors:  Álvaro Zubizarreta-Macho; César Castillo-Amature; José María Montiel-Company; Jesús Mena-Álvarez
Journal:  J Clin Med       Date:  2021-01-15       Impact factor: 4.241

6.  Factors Influencing the Accuracy of Freehand Implant Placement: A Prospective Clinical Study.

Authors:  Sigmar Schnutenhaus; Marie Wagner; Cornelia Edelmann; Ralph G Luthardt; Heike Rudolph
Journal:  Dent J (Basel)       Date:  2021-05-10

7.  The influence of dental experience on a dental implant navigation system.

Authors:  Ting-Mao Sun; Huey-Er Lee; Ting-Hsun Lan
Journal:  BMC Oral Health       Date:  2019-10-17       Impact factor: 2.757

8.  Comparing Accuracy of Implant Installation with a Navigation System (NS), a Laboratory Guide (LG), NS with LG, and Freehand Drilling.

Authors:  Ting-Mao Sun; Huey-Er Lee; Ting-Hsun Lan
Journal:  Int J Environ Res Public Health       Date:  2020-03-22       Impact factor: 3.390

9.  Accuracy of Computer-Aided Dynamic Navigation Compared to Computer-Aided Static Procedure for Endodontic Access Cavities: An in Vitro Study.

Authors:  Álvaro Zubizarreta-Macho; Ana de Pedro Muñoz; Elena Riad Deglow; Rubén Agustín-Panadero; Jesús Mena Álvarez
Journal:  J Clin Med       Date:  2020-01-02       Impact factor: 4.241

Review 10.  Effect of Computer-Aided Navigation Techniques on the Accuracy of Endodontic Access Cavities: A Systematic Review and Meta-Analysis.

Authors:  Álvaro Zubizarreta-Macho; Sara Valle Castaño; José María Montiel-Company; Jesús Mena-Álvarez
Journal:  Biology (Basel)       Date:  2021-03-10
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.