Construction and validation of a low cost paediatric pelvis phantom.

Ali Mohammed Ali1, Peter Hogg2, Safora Johansen3, Andrew England4.   

Abstract

PURPOSE: Imaging phantoms can be cost prohibitive, therefore a need exists to produce low cost alternatives which are fit for purpose. This paper describes the development and validation of a low cost paediatric pelvis phantom based on the anatomy of a 5-year-old child.
METHODS: Tissue equivalent materials representing paediatric bone (Plaster of Paris; PoP) and soft tissue (Poly methyl methacrylate; PMMA) were used. PMMA was machined to match the bony anatomy identified from a CT scan of a 5-year-old child and cavities were created for infusing the PoP. Phantom validation comprised physical and visual measures. Physical included CT density comparison between a CT scan of a 5-year old child and the phantom and Signal to Noise Ratio (SNR) comparative analysis of anteroposterior phantom X-ray images against a commercial anthropomorphic phantom. Visual analysis using a psychometric image quality scale (face validity).
RESULTS: CT density, the percentage difference between cortical bone, soft tissue and their equivalent tissue substitutes were -4.7 to -4.1% and -23.4%, respectively. For SNR, (mAs response) there was a strong positive correlation between the two phantoms (r > 0.95 for all kVps). For kVp response, there was a strong positive correlation between 1 and 8 mAs (r = 0.85), this then decreased as mAs increased (r = -0.21 at 20 mAs). Psychometric scale results produced a Cronbach's Alpha of almost 0.8.
CONCLUSIONS: Physical and visual measures suggest our low-cost phantom has suitable anatomical characteristics for X-ray imaging. Our phantom could have utility in dose and image quality optimisation studies.
Copyright © 2018 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Development; Dose optimisation; Low-cost; Pelvis phantom; Validation

Mesh:

Year:  2018        PMID: 30396676     DOI: 10.1016/j.ejrad.2018.09.015

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  3 in total

1.  Fabrication of 3D printed head phantom using plaster mixed with polylactic acid powder for patient-specific QA in intensity-modulated radiotherapy.

Authors:  Sung Yeop Kim; Jae Won Park; Jaehyeon Park; Ji Woon Yea; Se An Oh
Journal:  Sci Rep       Date:  2022-10-19       Impact factor: 4.996

2.  Proposal of a Lab Bench for the Unobtrusive Monitoring of the Bladder Fullness with Bioimpedance Measurements.

Authors:  Valentin Gaubert; Hayriye Gidik; Vladan Koncar
Journal:  Sensors (Basel)       Date:  2020-07-17       Impact factor: 3.576

3.  Optimization of Image Quality and Organ Absorbed Dose for Pediatric Chest X-Ray Examination: In-House Developed Chest Phantom Study.

Authors:  Thanyawee Pengpan; Natch Rattanarungruangchai; Juthathip Dechjaithat; Phawinee Panthim; Puntarika Siricharuwong; Ausanai Prapan
Journal:  Radiol Res Pract       Date:  2022-04-16
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.