Intra-aortic balloon pump experience: a single center study comparing with and without sheath insertion.

Yücel Özen1, Mehmet Aksut1, Davut Cekmecelioglu1, Mehmet Dedemoglu1, Ozge Altas1, Sabit Sarikaya1, Murat Bulent Rabus1, Kaan Kirali1.   

Abstract

Introduction: The mechanical circulation support used in treatment of low cardiac output at most is the intra-aortic balloon pump (IABP). Its usage fields are the complications occurring due to ischemic heart disease, disrupted left ventricle function, and the low cardiac output syndrome occurring during coronary artery by-pass surgery.
Methods: During 28 years from 1985 to 2013, IABP support has been implemented to 3135 patients in our cardiac surgery operating theater and intensive care unit. The mean age of the patients was 61.4 ± 13.2 years (16-82). 2506 patients (80%) were the ones whom the cardiac surgery has been implemented. IABP support has been provided for 629 (20%) patients for medical treatment. We utilized IABP most frequently in coronary artery patients (70%). The first choice for placing the balloon catheter is the femoral artery in 3093 cases (98.7%).
Results: The most frequently observed balloon complication was the lower extremity ischemia in 383 cases (12.2%).The leg ischemia was statistically significantly more frequent in patients with sheath (P=0.004). The extremity ischemia has developed in 4 of 12 patients with balloon placed from upper extremity. The local bleeding and balloon rupture were more frequent in patients whom the balloon has been placed without sheath. The mortality due to IABP has occurred in only 5 patients.
Conclusion: Despite increase in IABP usage frequency rapidly, the complications due to catheter are still seen. We believe that the leg ischemia that is the most frequently seen complication can be prevented via IABP use without sheath.

Entities:  

Keywords:  Intra-aortic Balloon Pump; Mechanical Support; Sheath Usage; Vascular Access

Year:  2018        PMID: 30386534      PMCID: PMC6203873          DOI: 10.15171/jcvtr.2018.23

Source DB:  PubMed          Journal:  J Cardiovasc Thorac Res        ISSN: 2008-5117


Introduction

Intra-aortic balloon pump (IABP) is the mechanical support device which is the one most frequently used in low cardiac output where the medical treatment is unsuccessful.[1] By decreasing the afterload and increasing the pressure in roots of aorta, IABP leads the coronary arterial perfusion to increase. Hence, the myocardial oxygen requirement decreases. Since its use by Kantrowitz et al[2] in 1968, IABP has been widely used in treatment of low cardiac output in many health care centers. The most frequent usage fields are unstable angina, complications developing secondary to ischemic cardiac disease, insufficient left ventricle function, and the low cardiac output syndrome developing during coronary heart surgery. Our aim in this study is to compare the complications in patients to whom the IABP with and without sheath has been implemented in our clinic.

Materials and Methods

Between the years of 1985 and 2013, IABP has been implemented on a total of 3135 patients in our clinic. The study data has been obtained by retrospectively scanning the intensive care and operating theater records. The patients to whom the surgery has been implemented were grouped into 3 groups by IABP placement times as preoperative, intraoperative, and postoperative. The patients whom the balloon has been placed in intensive care unit before taking into operation were clustered in preoperative group, the ones whom the balloon has been placed during operation were clustered in intraoperative group, and the patients whom the balloon has been placed in intensive care unit follow-up after the operation were clustered in postoperative. The patients where the trans-femoral path has been used were clustered into 2 groups as use with and without sheath. These groups were compared in terms of developed complications. The patients where the IABP used with ECMO (extracorporeal membrane oxygenation) were excluded from the study.

Statistical methods

Statistical analyses were carried out by utilizing the SPSS (Statistical Package for the Social Sciences) version 15.0 software. The definitive analyses were provided by using frequency tables of categorical variables. While comparing the intergroup variables, the “Pearson chi-square test” was utilized. P < 0.05 is accepted to be statistically significant.

Results

The mean age of the patients was 61.4 ± 13.2 years (16-82), and 2880 (92%) were males. Most of these patients had (70%) coronary artery disease (CAD). The demographical data of the patients are presented in Table 1. Cardiac surgery was applied to 2506 (80%) of the patients. Isolated CABG was implemented on 1842 (58.7%) patients. The CABGs performed off-pump were included in this group. IABP support was provided for 629 (20%) patients for medical treatment purpose. Among these patients, 235 (7.5%) patients were the ones with cardiomyopathy waiting for heart transplantation. The IABP usage indications are presented in Table 2. IABP has been implemented on 372 patients (11.8%) preoperatively, on 1246 (39.7%) patients postoperatively. The most frequent balloon implementation point was femoral artery with 3093 cases (98.7%). The balloon has been placed in percutaneous path in 2670 (85.2%) patients, and surgically in 423 (13.5%) patients. In percutaneous implementation, the balloon has been placed by sending to femoral artery through sheath until the year 2000. In this period, the balloon was implemented without sheath in patients having thin femoral artery. Since this date, the sheathless balloon placement method is adopted as routine. The diameter of the used balloon catheter was 8.5F or 9.0F. In 42 (1.3%) patients where it has not been placed in trans-femoral method, the alternative methods have been preferred. IABP implementation points are given in Table 3.
Table 1

Demographic characteristics of patients

No. %
The mean age61.4 ± 13.2 (16-82) Male288092
Female2558
Diabetes mellitus59018.8
Hypertension60219.2
Chronic obstructive pulmonary disease2588.2
Hyperlipidemia1896
Smoking101232.3
Peripheral artery disease2217
Coronary artery disease220470.3
Rheumatic valvular disease43814
Ischemic valve disease1133.6
Table 2

Indications for use of IABP

No. %
Surgery group 250680
I- Preoperative application (372)CAD32810.5
Valve disease341.1
Combinated100.3
II- Intraoperative application (888)(end of pump ) Isolated CABG83126.5
Double proceduresa 471.5
Complex surgeryb 100.3
III- Postoperative Application (1246)Isolated CABG101132.2
Double proceduresa 2116.7
Complex surgeryb 240.8
Medical group 62920
Acute MI and complications, 304 (%)1-Postinfarct VSD812.6
2-Ischemic mitral insufficiency521.6
3- Left ventricular dysfunction1504.8
4- Arrhythmia210.7
Unstable angina pectoris331
LMCA571.8
Cardiomyopathy2357.5

IABP: Inrtaaortic balloon pump, CAD: coronary artery disease, CABG: coronary artery bypass graft, MI: myocardial infarctus, VSD: ventricular septal defect, LMCA: left main coronary artery

a The patients who applied 2 procedures with CABG either valvular or carotid surgery

b The patients who applied 3 procedures as Bentall and CABG or CABG and double valve replacement.

Table 3

IABP application areas

No. %
Transfemoral, 3093 (98.7%) Percutaneous2670 (85.2%) With sheath93429.8
Without sheath173655.4
Surgical 423 (13.5%) With sheath1605.1
Without sheath2638.4
Alternative routes42 (1.3%) Transthoracic30 (0.1%) Direct100.3
Via graft200.6
Transaxillary12 (0.4%) Direct80.2
Via graft40.1

IABP: Intraaortic aort balloon pump.

a The total number of patients used IABP between 1985 and 2013: 3135

IABP: Inrtaaortic balloon pump, CAD: coronary artery disease, CABG: coronary artery bypass graft, MI: myocardial infarctus, VSD: ventricular septal defect, LMCA: left main coronary artery a The patients who applied 2 procedures with CABG either valvular or carotid surgery b The patients who applied 3 procedures as Bentall and CABG or CABG and double valve replacement. IABP: Intraaortic aort balloon pump. a The total number of patients used IABP between 1985 and 2013: 3135 The most frequently seen complication in patients whom the balloon has been placed was lower extremity ischemia (383, 12.2%). Surgical intervention was implemented on 204 (6.5%) patients. The embolectomy has been implemented mostly (153, 4.9%). In 4 patients with advanced ischemia, the amputation was implemented. In 4 of 42 patients whose IABPs were placed in alternative paths, the upper extremity ischemia developed. In 2 of them, the surgical intervention was required. Also in 44 (1.4%) patients, there was bleeding in balloon placement points. Even though hospital mortality was 814 (25.9%), only 5 of them occurred as a result of complications due to balloon (Table 4).
Table 4

IABP complications

Complications No. %
1-Extremity ischemia, 387 (12.3%) Low extremity, 379 (12%) Medical1795.7
Surgical204 (6.5%) Embolectomy1534.9
Embolectomy + saphenous patch plasty280.9
Embolectomy + graft interposition80.2
Femoro-femoral bypass20.06
Fasciotomy50.1
Amputation40.1
Upper extremity, 4 (0.12%) Medical20.06
Surgical2 (0.06%) Embolectomy10.03
Embolectomy + graft interposition10.03
2. Local bleeding 441.4
3. Hemolysis 210.7
4. Thrombocytopenia 160.5
5. Infections 210.7
6. Chronic serous discharge 230.7
7. Paraplegia 50.1
8. Aortic complications, 7 (% 0.2) Dissection >3>0.1
Rupture 40.1
9. Balloon Rupture 822..6
10. Cerebral Embolism 30.1

IABP: Intraaortic aort balloon pump.

IABP: Intraaortic aort balloon pump. Paraplegia, which is a rare complication, developed in 5 patients. Thrombocytopenia developed in 16 patients, while hemolysis developed in 21 patients. In 7 patients, the aorta complication developed. Dissection occurred in 3 of these patients, and aorta rupture occurred in 4. And 4 of them were lost in early period. One of 3 patients where the cerebral emboli developed was lost in postoperative 35th day. The mortality due to balloon was found to be 0.16% with 5 patients (Table 4).

Discussion

The need for increased use of IABP during cardiac surgery in the recent years has been reported by many groups.[3,4] The first clinical application of IABP was in 1968 by Kantrowitz et al to support patients with cardiogenic shock due to myocardial infarction.[2] The most frequently used one among the mechanical circulation support devices nowadays is the IABP. The most important reasons of that are its easy implementation, and its clear price advantage against other devices.[5] IABP has two principal physiological effects; it improves coronary blood flow and myocardial oxygen supply by increasing diastolic perfusion pressure and reduces afterload by rapid balloon deflation in systole leading to reduced ventricular work and consequently decreased myocardial oxygen consumption.[6] The real usage field of IABP, which is widely used in cardiology clinics and intensive care units, is the centers where the open cardiac surgery is applied. The aim of use in these centers can be divided into 3 as preventing the low cardiac output and severe myocardial ischemia in preoperative period, providing assistive support for patients who cannot leave the heart-lung device in intraoperative period, and preventing the low output in intensive care unit or medical-treatment-resistant arrhythmia in postoperative period.[7] The most frequent balloon usage in patients whom we applied surgery in our clinic is in postoperative period with 1011 cases (32.2%). Despite the positive hemodynamic effects of IABP, the complication rates related to IABPs are not low. In a study which reviewed the recent literature, the complication rates related to IABPs were reported to be between 32.6% and 50%.[8] IABPs may lead to complications like thrombocytopenia, bleeding, injury to the aorta and iliac arteries, dissection, thromboembolism and leg ischemia. Thrombocytopenia and bleeding are the most commons complications.[9] Additionally, complications related to malposition of the balloon catheter can be observed.[10] Even though the incidence of arterial thrombosis and emboli development is decreased by implementing systemic heparinizing during IABP usage, the implementation’s effect increasing the bleeding in postoperative period may lead to stop or delay the usage. Among these complications, the most frequently observed one was the upper extremity ischemia with 387 cases (12.3%) (Table 4). The extremity ischemia was observed in lower extremity where the balloon catheter was placed and in patients whom the device was implemented with sheath. Comparing with patients without sheath, the result was found to be statistically significant (Table 5). The longer the IABP support continues, the more the ischemia development rate in that extremity increases. The reason of that ischemia which the balloon causes is the damage on vessel wall during placing the balloon catheter, the prevention of blood circulation by catheter, and vasospasm occurring due to low cardiac output. Female gender, advanced age, diabetes, and atherosclerotic heart disease are the factors speeding this complication up, and the development of lower extremity ischemia in such patients was found to be more frequent. In patients where the balloon catheter cannot be placed via femoral artery, the transthoracic, axillary or subclavian artery path is an alternative.[11] In our clinic implementation, we placed IABP to 42 patients where the trans-femoral path could not be used (Table 2). Among these patients, upper extremity ischemia developed in 4 patients, and surgical intervention was required in 2 of them. In order to minimize the vascular complications in IABP implementation, we generally prefer sheathless balloon catheter implementation. The advantage of this method is to decrease the mechanical obstruction. But there may be the bleeding as severe leakage in insertion point of balloon catheter (Table 5). Local bleeding was found to be more frequent in sheathless group (P = 0.03). This bleeding can be taken under control easily via purse sutures around the catheter. Bleeding occurred in 44 of our patients at balloon insertion point. In 21 patients, the infection developed in insertion point of balloon. 18 of these patients were the ones where the balloon was installed surgically because it could not be installed in percutaneous path. Thrombocytopenia developed in 16 patients. In these patients, the thrombocytopenia cleared up after heparin implementation was cut, and balloon was removed. Balloon rupture developed in 82 patients. The balloon rupture was found to be statistically significantly more frequent in sheathless group (P = 0.02). As a result of increased quality of used material and the technological developments, the balloon rupture is observed less frequently in recent years. The rapid flow of blood into the balloon catheter or inability of performing filling must make us think balloon rupture. In that case, the balloon must be removed immediately. Cerebral emboli developed in 3 patients. The reason of the emboli may be removal of atherosclerotic plaque from aortic or carotid tips of balloon catheter during installing the balloon. Or the air emboli due to balloon rupture may have led that. For this reason, the length must be measure on the patient before installing the balloon, and the location of the balloon must be checked through transesophageal echocardiography or teleradiograph after the balloon is installed. One of the patients where cerebral emboli developed due to balloon rupture died on postoperative 43th day due to low cardiac output. Aorta rupture developed in 4 patients. By taking 2 of these patients under surgery, the rupture in aorta was restored with graft. But other 2 patients died. Aorta dissection developed in 3 patients. Two of them died. And 1 of the patients was not intervened. In control tomography, it was seen that the rupture has limited. It is very hard to predict this catastrophic situation. In cases where it is hard to push the guide wire forward during the process, we recommend to assess the abdominal aorta pathologies through abdominal ultrasonography.
Table 5

Hospital mortality

No. Exitus %
Surgical group 250670928.3
Preoperative group3726116.4
Intraoperative group88822024.8
Postoperative group124642834.3
Transfemoral access309380125.9
Alternative access421330.1
Medical group 62910516.7
Balloon-related mortality 313550.16
Total hospital mortality 313581425.9
Also the timing of IABP support is a controversial issue nowadays. The mortality is less than 20% in preoperative installation. But the mortalities in intraoperative and postoperative installations are almost 30% and 40%, respectively.[12] Lavana et al evaluated the association between timing of IABP insertion and outcomes of the patients reporting mortality rates as 10% in preoperative group, 16% in the intraoperative group and 29% in the postoperative group. They concluded that preoperative IABP use was associated with reduction in hospital mortality.[13] In study of Parissis et al, the preoperative mortality was found to be 18.2%, while the postoperative mortality was found to be 58.3%.[14] Given results led to the suggestion of preoperative prophylactic IABP insertion in high risk patients before observing any hemodynamic compromise.[15] In our study, the preoperative mortality was found to be 16.4%, while intraoperative mortality was found to be 24.8% and postoperative mortality was found to be 34.3%. Even though the total hospital mortality was 25.9%, the mortality due to balloon was found to be 0.16% (Table 6). The total hospital mortality in study carried out in our clinic at 1999 has been found to be 26.6%.[16]
Table 6

Complication comparison in transfemoral IABP applications

With sheath (n=1999) Without sheath (n=1094) P
No. % No. %
Extremity ischemia28214.11019.20.
Amputation40.2000.04
Local bleeding110.5242.20.03
Local infection90.490.81
Balloon rupture261.3423.80.02

IABP: Intraaortic aort balloon pump.

IABP: Intraaortic aort balloon pump.

Conclusion

The IABP of nowadays is a mechanical support device which is still used safely and widely in treatment of low cardiac output developing after cardiac surgery. The easiness of implementation of this mechanical support is the main reason of that it is first preference in all kinds of intensive care units and open heart surgery. As a result of our IABP implementations in single center, we think that the preoperative and early-intraoperative balloon implementations are safer than intraoperative and postoperative implementations. Even though the complications due to balloon catheter decreased in recent years, they still occur. We recommend sheathless usage in order to prevent the extremity ischemia which is the most frequent complication developing due to IABP.

Limitations

This is a retrospective study, and not all the records of old patients could be reached. Since the medium- and long-term follow-ups of patients could not be reached, the long-term mortality rates could not be calculated.

Ethical approval

The data have been gathered and analyzed retrospectively, after approval of local ethical committee.

Competing interests

The authors declare that they have no competing interests.
  14 in total

1.  Effective surgical management of high-risk coronary patients using preoperative intra-aortic balloon counterpulsation therapy.

Authors:  J T Christenson; M Schmuziger; F Simonet
Journal:  Cardiovasc Surg       Date:  2001-08

2.  Visceral arterial compromise during intra-aortic balloon counterpulsation therapy.

Authors:  Ardawan Julian Rastan; Eugen Tillmann; Sreekumar Subramanian; Lukas Lehmkuhl; Anne Kathrin Funkat; Sergej Leontyev; Torsten Doenst; Thomas Walther; Matthias Gutberlet; Friedrich Wilhelm Mohr
Journal:  Circulation       Date:  2010-09-14       Impact factor: 29.690

3.  Intraaortic balloon counterpulsation: patterns of usage and outcome in cardiac surgery patients.

Authors:  L L Creswell; M Rosenbloom; J L Cox; T B Ferguson; N T Kouchoukos; T L Spray; M K Pasque; T B Ferguson; T H Wareing; C B Huddleston
Journal:  Ann Thorac Surg       Date:  1992-07       Impact factor: 4.330

4.  Transthoracic intraaortic balloon pump: a simplified technique.

Authors:  J H Burack; P Uceda; J N Cunningham
Journal:  Ann Thorac Surg       Date:  1996-07       Impact factor: 4.330

5.  Efficacy and cost-effectiveness of preoperative IABP in patients with ejection fraction of 0.25 or less.

Authors:  C A Dietl; M D Berkheimer; E L Woods; C L Gilbert; W F Pharr; C H Benoit
Journal:  Ann Thorac Surg       Date:  1996-08       Impact factor: 4.330

6.  Trends in intraaortic balloon counterpulsation complications and outcomes in cardiac surgery.

Authors:  Jan T Christenson; Marc Cohen; James J Ferguson; Robert J Freedman; Michael F Miller; E Magnus Ohman; Ramachandra C Reddy; Gregg W Stone; Philip M Urban
Journal:  Ann Thorac Surg       Date:  2002-10       Impact factor: 4.330

7.  Intra-aortic balloon counterpulsation with special reference to determinants of survival.

Authors:  J Hedenmark; H Ahn; A Henze; S O Nyström; R Svedjeholm; H Tydén
Journal:  Scand J Thorac Cardiovasc Surg       Date:  1989

8.  The need for intra aortic balloon pump support following open heart surgery: risk analysis and outcome.

Authors:  Haralabos Parissis; Michael Leotsinidis; Mohammad Tauqeer Akbar; Efstratios Apostolakis; Dimitrios Dougenis
Journal:  J Cardiothorac Surg       Date:  2010-04-05       Impact factor: 1.637

9.  Influence of timing of intraaortic balloon placement in cardiac surgical patients.

Authors:  Jayshree D Lavana; John F Fraser; Susan E Smith; Lesley Drake; Peter Tesar; Daniel V Mullany
Journal:  J Thorac Cardiovasc Surg       Date:  2009-11-11       Impact factor: 5.209

Review 10.  Intra aortic balloon pump: literature review of risk factors related to complications of the intraaortic balloon pump.

Authors:  Haralabos Parissis; Alan Soo; Bassel Al-Alao
Journal:  J Cardiothorac Surg       Date:  2011-11-02       Impact factor: 1.637

View more
  2 in total

1.  Intra-Aortic Balloon Pump Catheter Insertion Using a Novel Left External Iliac Artery Approach in A Case of Chronic Heart Failure Due to Dilated Cardiomyopathy.

Authors:  Mirosław Dziekiewicz; Marek Banaszewski; Mateusz Kuć; Janina Stępińska
Journal:  Am J Case Rep       Date:  2019-12-07

Review 2.  Complications of Temporary Percutaneous Mechanical Circulatory Support for Cardiogenic Shock: An Appraisal of Contemporary Literature.

Authors:  Anna V Subramaniam; Gregory W Barsness; Saarwaani Vallabhajosyula; Saraschandra Vallabhajosyula
Journal:  Cardiol Ther       Date:  2019-10-23
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.