Sachin Tayal1, Arun Chandanmal Gandhi1, Abbas Ali1.
Abstract
A failure in any of the hardware components can lead to degradation in image quality and accuracy of quantification. Few artifacts may bring serious impacts on the image quality and finally lead to the wrong diagnosis. We encountered a sudden appearance of the cold area, being more prominently visible on Positron Emission Tomography image in comparison to the fused image. A daily quality assurance scan was performed to evaluate the status of the machine after giving a restart to the system. It was discovered that the module number 9-11 showed dark black color in the graph due to scarcity of counts indicating a failure/malfunctioning of the detector or electronic board-Cassette Electronic Module board (CEM board) supporting the module 9-11. Further evaluation of the system helped us diagnose that one of the electronic board (CEM Board) had become nonfunctional. A new functional electronic board (CEM board) was ordered and placed and the error rectified.Entities:
Keywords: Artifact; daily quality assurance; detector module; electronic board
Year: 2018 PMID: 30386062 PMCID: PMC6194775 DOI: 10.4103/ijnm.IJNM_70_18
Source DB: PubMed Journal: Indian J Nucl Med ISSN: 0974-0244
Figure 1Artifact as seen in axial images
Figure 2(a) Computed tomography image. (b) Positron emission tomography image showing loss of counts in image on the left upper region. (c) Positron emission tomography–computed tomography fusion image. (d) Maximum intensity projection image of the study
Figure 3Daily quality assurance images as seen after a restart of the system, showing a defect in module 9–11
Figure 4Daily quality assurance image showing the same defect being shifted to module 33 and 34 after assigning the malfunctioning electronic board (cassette electronic module board) to module 33 and 34
Figure 5Daily quality assurance image after replacement of faulty electronic board (cassette electronic module board) with a new and functional one, showing complete acceptance of all parameters