Barbara Lawrenz1,2, Junard Sibal3, Nicolas Garrido4, Emmanuel Abu3, Alliza Jean3, Laura Melado1, Human M Fatemi1.
Abstract
In recent years there is increasing evidence that elevated progesterone levels during ovarian stimulation for IVF / ICSI have a negative impact on the ART-outcome. However, different progesterone assays were used in the previous studies and different assays might produce varying results. This retrospective study evaluated the reproducibility and reliability of different progesterone assays with a special focus on progesterone levels below 1.5 ng/ml, as this range is crucial for early detection of progesterone rise during ovarian stimulation for IVF. A total of 413 blood samples were categorized in different progesterone ranges and whether they were retrieved on the day of final oocyte maturation and the results were compared regarding their reproducibility and reliability. To compare the reproducibility between the different progesterone assays, the Intraclass Correlation Coefficient (ICC) was calculated and interpretation of the ICC results was done according to Cicchetti, ranging from poor to excellent. The correlation of the assays was excellent when all samples were compared including samples retrieved on day of final oocyte maturation, however in the ranges of progesterone levels 1.0 ng/ml to < 1.5 ng/ml, 0.8 ng/ml to < 1.0 ng/ml and < 0.8 ng/ml, the ICC varied between poor and excellent. The assays "gen III" and "Architect" showed an excellent reproducibility of progesterone results throughout all ranges of progesterone levels. This analysis demonstrates, that different progesterone assays have a limited reproducibility and that the results depend on the assay used and the range of progesterone level. This fact leads to two important conclusions. Firstly the limited reproducibility might lead to substantially different treatment decisions in ovarian stimulation treatment for IVF and secondly critical interpretation of thresholds, provided by meta-analysis, is crucial despite the risk that the so far gained clinical experience might become irrelevant and has to be adjusted to the results, obtained by each assay.Entities:
Mesh:
Substances:
Year: 2018 PMID: 30383804 PMCID: PMC6211677 DOI: 10.1371/journal.pone.0206098
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Means, 95% confidence intervalls (95%CI) and standard deviations (SD) for the progesterone results, results below the detection range are calculated as „0.00“.
| Type of sample | Number of samples | Mean (ng/ml) | 95% Confidence Intervall (95%CI) | Standard-deviation |
|---|---|---|---|---|
| „gen II” | 413 | 1.13 | 0.78–1.47 | 3.55 |
| „gen III” | 413 | 0.76 | 0.47–1.05 | 2.95 |
| „Architect“ | 121 | 0.75 | 0.67–0.83 | 0.46 |
| „gen II” | 34 | 7.42 | 3.73–11.11 | 10.57 |
| „gen III” | 34 | 6.44 | 3.47–9.40 | 8.50 |
| „Architect“ | 14 | 1.57 | 1.31–1.84 | 0.46 |
| „gen II” | 45 | 1.22 | 1.17–1.26 | 0.13 |
| „gen III” | 45 | 0.68 | 0.60–0.77 | 0.27 |
| „Architect” | 26 | 1.02 | 0.89–1.14 | 0.31 |
| „gen II” | 30 | 0.89 | 0.87–0.91 | 0.57 |
| „gen III“ | 30 | 0.51 | 0.42–0.60 | 0.24 |
| „Architect” | 9 | 0.76 | 0.58–0.93 | 0.22 |
| „gen II” | 304 | 0.44 | 0.42–0.46 | 0.18 |
| „gen III” | 304 | 0.16 | 0.15–0.18 | 0.16 |
| „Architect” | 70 | 0.49 | 0.43–0.54 | 0.23 |
| „gen II” | 121 | 0.86 | 0.77–0.94 | 0.49 |
| „gen III” | 121 | 0.46 | 0.39–0.54 | 0.42 |
| „Architect” | 121 | 0.75 | 0.67–0.84 | 0.46 |
| „gen II” | 72 | 0.89 | 0.78–1.01 | 0.49 |
| „gen III” | 72 | 0.48 | 0.38–0.57 | 0.40 |
| „Architect” | 72 | 0.77 | 0.67–0.87 | 0.43 |
| „gen II” | 49 | 0.79 | 0.65–0.94 | 0.49 |
| „gen III” | 49 | 0.44 | 0.31–0.56 | 0.44 |
| „Architect” | 49 | 0.74 | 0.59–0.88 | 0.51 |
Reproducibility of the progesterone assays „gen 2“, „gen 3”and „Architect”according to Intraclass Correlation Coefficient (ICC) and interpretation according to Cicchetti et al. [19].
| Comparison between | No of cases | ICC single measures | Inter-pretation | ICC average measures | Inter-pretation | |
|---|---|---|---|---|---|---|
| gen 2 vs gen 3 | 413 | 0.973 | Excellent | 0.986 | Excellent | |
| gen 2 vs Architect | 121 | 0.814 | Excellent | 0.897 | Excellent | |
| gen 3 vs Architect | 121 | 0.957 | Excellent | 0.978 | Excellent | |
| gen 2 vs gen 3 | 34 | 0.966 | Excellent | 0.983 | Excellent | |
| gen 2 vs Architect | 14 | 0.287 | Poor | 0.446 | Fair | |
| gen 3 vs Architect | 14 | 0.938 | Excellent | 0.968 | Excellent | |
| gen 2 vs gen 3 | 45 | 0.288 | Poor | 0.488 | Fair | |
| gen 2 vs Architect | 26 | 0.315 | Poor | 0.479 | Fair | |
| gen 3 vs Architect | 26 | 0.887 | Excellent | 0.940 | Excellent | |
| gen 2 vs gen 3 | 30 | 0.138 | Poor | 0.242 | Poor | |
| gen 2 vs Architect | 9 | 0.127 | Poor | 0.225 | Poor | |
| gen 3 vs Architect | 9 | 0.779 | Excellent | 0.876 | Excellent | |
| gen 2 vs gen 3 | 304 | 0.544 | Fair | 0.704 | Good | |
| gen 2 vs Architect | 70 | 0.359 | Poor | 0.529 | Fair | |
| gen 3 vs Architect | 70 | 0.846 | Excellent | 0.917 | Excellent | |
| gen 2 vs gen 3 | 72 | 0.851 | Excellent | 0.919 | Excellent | |
| gen 2 vs Architect | 72 | 0.803 | Excellent | 0.890 | Excellent | |
| gen 3 vs Architect | 72 | 0.955 | Excellent | 0.977 | Excellent |
Fig 1Reproducibility of the progesterone assays „gen 2“, „gen 3”and „Architect“.
Progesterone levels are represented as scatter plots and with the confidence interval 95% (outer lines). X- and Y-axis presenting the progesterone levels in ng/ml in all diagrams.