| Literature DB >> 30336557 |
Casper Hyttel Clausen1, Maria Dimaki2, Christian Vinther Bertelsen3,4, Gustav Erik Skands5, Romen Rodriguez-Trujillo6,7, Joachim Dahl Thomsen8, Winnie E Svendsen9.
Abstract
Monitoring of bacteria concentrations is of great importance in drinking water management. Continuous real-time monitoring enables better microbiological control of the water and helps prevent contaminated water from reaching the households. We have developed a microfluidic sensor with the potential to accurately assess bacteria levels in drinking water in real-time. Multi frequency electrical impedance spectroscopy is used to monitor a liquid sample, while it is continuously passed through the sensor. We investigate three aspects of this sensor: First we show that the sensor is able to differentiate Escherichia coli (Gram-negative) bacteria from solid particles (polystyrene beads) based on an electrical response in the high frequency phase and individually enumerate the two samples. Next, we demonstrate the sensor's ability to measure the bacteria concentration by comparing the results to those obtained by the traditional CFU counting method. Last, we show the sensor's potential to distinguish between different bacteria types by detecting different signatures for S. aureus and E. coli mixed in the same sample. Our investigations show that the sensor has the potential to be extremely effective at detecting sudden bacterial contaminations found in drinking water, and eventually also identify them.Entities:
Keywords: bacteria counting; bacteria detection; bacteria differentiation; electrical impedance spectroscopy; water quality
Mesh:
Substances:
Year: 2018 PMID: 30336557 PMCID: PMC6210208 DOI: 10.3390/s18103496
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Figure 1Schematic drawings of the bacteria and particles used in this report (A) Gram-positive S. Aureus (B) Gram- negative E. coli. (C) Polystyrene beads.
Figure 2(A) Electric fields generated by two electrodes at low frequencies (1 and 2) and high frequencies (3 and 4) on beads (1 and 3) and bacteria (2 and 4). (B) Schematic of the measuring principle using front facing (1) or coplanar (2) electrodes. A differential signal between the two measuring electrodes is recorded for further analysis. (C) Schematic drawing of the measurement setup, in the case of coplanar electrodes. The sample is injected into the system by a syringe pump. A multi-frequency lock-in amplifier is used to generate and detect the signal. The signal from the measuring electrodes is passed through a current preamplifier before it is returned to the lock-in amplifier. (D) Schematic showing the detection principle. When the particle is not influencing the electric field generated by the electrodes, the differential signal is zero. As the particle travels in the channel it will only disturb the field of one set of electrodes at a time, which gives rise to a differential signal. When the particle leaves the electrodes area a transition looking like a sinusoidal signal will have been recorded, indicating the transition of one particle.
Figure 3(A) Opacity plotted against the low frequency signal of E. coli and polystyrene beads in drinking water. (B) Phase shift at 7 MHz plotted against the low frequency signal of E. coli and polystyrene beads in drinking water. (C,D) Same as (A,B), but with the bacteria and beads dispersed in 1/20× PBS.
Details of the concentration determination experiments, along with the measured transitions and the calculated concentrations by EIS and plating.
| Sample | Measurement Time (s) | Beads (#) | Beads (/mL) | Plate Count (/mL) | ||
|---|---|---|---|---|---|---|
| A | 1551.46 | 377 | 191,202 | 7 | 6221 | 3000 |
| B | 1554.65 | 426 | 201,612 | 92 | 48,181 | 154,000 |
| C | 1551.71 | 398 | 192,603 | 224 | 118,006 | 326,000 |
| D | 1551.9 | 366 | 185,596 | 483 | 274,769 | 643,000 |
| E | 1861.85 | 445 | 186,478 | 843 | 396,767 | 1.00 × 106 |
Figure 4Bacteria concentration measured with the system (impedance spectroscopy) and colony-forming unit counting. The beads were only counted using the system and their concentration was kept constant at 2 × 105 mL−1. The dashed line represents the 1:1 relation between the two methods. The dotted trend line for the E. coli data has an R2 value of 0.9937.
Figure 5(A) Phase shift at 8 MHz plotted against the low frequency signal of the bacteria E. coli and S. aureus and polystyrene beads in diluted PBS (B) Histogram of the high frequency phase angle, clearly showing the two populations of bacteria.