Puntiwa Oonsiri1, Kitwadee Saksornchai1, Sivalee Suriyapee2.
Abstract
Radiation protection in the scrotum to reduce the risk of genetic effect in the future is very important. This study aimed to measure the scrotal dose outside the treatment fields by using the radio-photoluminescence glass dosimeter (RPLGD). The characteristics of RPLGD model GD-302M were studied. Scattered dose to scrotum was measured in one liposarcoma case with the prescribed dose of 60 Gy. RPLGDs were placed in three different locations: one RPLGD was positioned at the posterior area which closer to the scrotum, and the other two RPLGDs were placed between the penis and the scrotum. Three RPLGDs were employed in each location. The scattered doses were measured in every fraction during the whole course of treatment. The entire number of 100 RPLGDs showed the uniformity within ±2%. The signal from RPLGD demonstrated linear proportion to the radiation dose (r = 0.999). The relative energy response correction factor was 1.05. The average scrotal dose was 4.1 ± 0.9 cGy per fraction. The results presented a wide range since there was a high uncertainty during RPLGD placement. The total scrotal dose for the whole course of treatment was 101.9 cGy (1.7% of the prescribed dose). The RPLGD model GD-302M could be used to measure scattered dose after applying the relative energy correction factor.Entities:
Keywords: Radio-photoluminescent glass dosimeter (RPLGD); Scrotal dose; Testicular shielding
Year: 2018 PMID: 30309217 PMCID: PMC6226135 DOI: 10.3857/roj.2018.00143
Source DB: PubMed Journal: Radiat Oncol J ISSN: 2234-1900
Fig. 1.(A) The setup position of radio-photoluminescence glass dosimeter (RPLGD) under calibration condition for 6 MV photon beams and (B) the setup position of RPLGD, Ir-192, and FC-65G inside the cylindrical PMMA phantom for brachytherapy.
Fig. 2.(A) The setup of testicular shield for liposarcoma case. (B) The radio-photoluminescence glass dosimeter (RPLGD) placement inside the testicular shield. (C, D) The irradiation field for liposarcoma case: coronal plane and digital reconstructed radiograph.
Fig. 3.The linearity response of radio-photoluminescence glass dosimeters.
Fig. 4.The average scattered dose to the scrotum for each fraction.