Passive limb movement test facilitates subthalamic deep brain stimulation under general anesthesia without influencing awareness.

Sheng-Tzung Tsai1, Shee-Ping Chen2, Sheng-Huang Lin3, Shinn-Zong Lin1, Shin-Yuan Chen1.   

Abstract

OBJECTIVES: We have shown that neuronal activity in the subthalamic nucleus (STN) in patients with Parkinson's disease can be accurately recorded during deep brain stimulation (DBS) with general anesthesia (GA). However, a vigorous passive range of motion (PROM) test might exert awakening effects on patients who are lightly anesthetized. We will explore the effects of PROM on the heart rate (HR) and mean arterial pressure (MAP) during microelectrode recording (MER) and confirm whether it facilitates identifying the sensory motor portion of the STN under GA.
MATERIALS AND METHODS: 3T magnetic resonance image targeting of the STN was done to guide MER during frame-based stereotactic procedures for DBS. Regular induction and endotracheal intubation for GA were performed and then maintained with a volatile anesthetic agent and muscle relaxant only. The depth of anesthesia was monitored by the bispectral index (BIS).
RESULTS: A total of ten patients were enrolled in this study. Their mean age was 48.5 ± 10.8 years old with a disease duration 8.6 ± 2.4 years at the time of surgery. During MER, PROM significantly decreased recording tract numbers and still reached the STN at a recorded length at 5.5 ± 0.8 mm. Compared with baseline, PROM increased HR by a mean 0.5 beats/min and MAP by a mean 1.4 mmHg (P = 0.1178 and 0.0525). The change in BIS was -0.7 (P = 0.4941), and the mean alveolar concentration of the anesthetic agent changed little throughout surgery.
CONCLUSIONS: PROM was effective in triggering and magnifying neuronal firing signal without influencing patient awareness during MER for STN-DBS under GA.

Entities:  

Keywords:  Deep brain stimulation; General anesthesia; Microelectrode recording; Passive range of movement; Subthalamic nucleus

Year:  2018        PMID: 30305788      PMCID: PMC6172905          DOI: 10.4103/tcmj.tcmj_17_18

Source DB:  PubMed          Journal:  Ci Ji Yi Xue Za Zhi


INTRODUCTION

Conventional deep brain stimulation (DBS) surgery for Parkinson's disease (PD) requires a patient to be awake to obtain proper neuronal signals during microelectrode recording (MER) and to perform test stimulation [123]. Nevertheless, patients with intractable medication off symptoms may not tolerate a prolonged awake surgical procedure. These off symptoms may include tremor, rigidity, and choking by drooling in the advanced stage. Thirty years after the first DBS procedure in a PD patient, changes are occurring in the paradigm in approaching DBS surgery, especially in the choice of anesthesia regimen, to improve the comfort of patients without jeopardizing outcomes [45]. Identifying the sensory motor portion of the subthalamic nucleus (STN) is crucial during MER for optimal final electrode implantation. Passive range of motion (PROM) of the patient's limbs enhances neuronal firing during MER of the sensory motor region of the STN. In our previous research, we proved that MER can be adequately performed while the patient receives desflurane under general anesthesia (GA), and the improvement in PD motor function was comparable to patients who underwent local anesthesia [67]. Nevertheless, more vigorous PROM may be needed to trigger neuronal signals effectively during GA than during an awake procedure. During GA for DBS, the end-tidal anesthetic concentration is usually maintained between 0.5 and 1.0 minimal alveolar concentration (MAC), and the anesthetic state is monitored by the patient heart rate (HR) and blood pressure [8]. Nevertheless, unresponsiveness in motor activity is not equal to unconsciousness of an anesthetized patient [910]. A combination of electroencephalography, electromyography, and standard vital sign measurement could be more realistic in monitoring the depth of anesthesia to ensure the quality and safety of GA [111213]. In this study, through continuous monitoring of the depth of anesthesia through the HR, mean arterial pressure (MAP), bispectral index (BIS), and minimal alveolar concentration (MAC) of the volatile anesthetic agent, we tried to clarify if vigorous PROM exerts awakening effects on patients who are lightly anesthetized. The goal is to find the most effective way to identify the sensory motor portion of the STN without influencing the depth of anesthesia during MER.

MATERIALS AND METHODS

This was a prospective cohort study of ten PD patients who underwent STN-DBS surgery. This work was approved by the Institutional Review Board of Tzu Chi General Hospital, Hualien, Taiwan (TCRD 104-35, IRB 103-122-A).

Patient selection

Ten patients who received GA during STN-DBS for PD were enrolled in this prospective study.

Imaging and targeting

The image fusion technique was applied in all cases. On the day of surgery, computed tomography (CT) of the brain was performed axially, in 1.25 mm intervals, with a stereotactic localizer. The standard 3T magnetic resonance (MR) images' (GE, 3.0 Tesla, USA) settings were T1 three-dimensional fast spoiled gradient echo (FSPGR) axial images of 1.0 mm thickness, with and without contrast enhancement and T2 FSPGR (T2W FSE) axial images at a 2 mm thickness. Each of these sequences was performed in contiguous slices. All images were transferred in a DICOM database through the PACS system to an S7 workstation (Medtronic, Minneapolis, MN, USA). Image fusion software was used to fuse all sets of MR images to CT scan images. The tentative surgical target coordinates for the tip of the permanently implantable electrode were set at the central lowest border of the STN through direct visualization on MR images and adjusted with indirect coordinates.

Stereotactic procedures

A Leksell G-frame (Elekta Instrument, Inc., Atlanta, GA, USA) was used for the stereotactic procedure. The patients rested comfortably on the operating table in a supine position with the head ring secured on a Mayfield adaptor. The target coordinates were applied to the stereotactic frame and the working stage. After MER procedures, the permanent electrodes were implanted.

Anesthetic procedures

Regular induction and endotracheal intubation for GA were performed and then maintained with a volatile anesthetic agent (desflurane or sevoflurane) and muscle relaxant only. The depth of anesthesia was monitored by the BIS. HR and MAP were recorded before and after each PROM.

Microelectrode recording procedures

Each microelectrode had a 10 mm bare tungsten tip and was 10 to 40 μm in diameter; the recording impedance was 0.5–1.5 MΩ (FHC, USA). The microelectrode was mounted on a microdriver. The signal recorded from the tip of the microelectrode was magnified and displayed by a Leadpoint intraoperative microrecording system (Medtronic). PROM was performed on both the upper and lower limbs to see if there was movement-related neuronal firing. Raw spike firings in the STN were transmitted through a Power 1401 High-Performance Data Acquisition Interface (CED, England) for off-line analysis by Spike 2. The microelectrode was advanced from a distance of 200 to 500 μm to detect the robust firing of neurons. After electrode implantation, intraoperative fluoroscopy was used to record the position of the microelectrode tip, which served as a reference for final electrode implantation. A macrostimulation test was not performed. The length of the recorded STN had to be at least 4.5 mm to be considered adequate for a good clinical outcome.

Protocol for data collection

Ten seconds of STN neuronal firings were recorded by MER before (baseline) and after PROM [Figure 1]. At the same time, the BIS, HR, and blood pressure were collected through Biopac amplifiers then through an AD converter (CED Micro 1401) to be stored for off-line analysis.
Figure 1

Microelectrode recordings of the subthalamic nucleus show a remarkable difference between raw neuronal firings under general anesthesia ([a] mean firing rate: 0), and firings under general anesthesia with passive movement ([b] mean firing rate: 47)

Microelectrode recordings of the subthalamic nucleus show a remarkable difference between raw neuronal firings under general anesthesia ([a] mean firing rate: 0), and firings under general anesthesia with passive movement ([b] mean firing rate: 47)

Postoperative course

Brain CT in 1.25 mm consecutive slices was performed before the patient was sent to the ward, to exclude intracranial complications and to evaluate the initial postoperative electrode coordinates through image fusion with the preoperative MRI. Another MRI was done to confirm the final electrode coordinates within 3 months postoperatively.

Statistical analysis

Paired t-tests were used to compare the pre- and post-PROM changes in the HR, MAP, BIS, and MAC of the volatile anesthetic agent. P < 0.05 was considered statistically significant.

RESULTS

A total of 10 patients (female/male: 3/7) were enrolled in this study. Their mean age at surgery was 48.5 ± 10.8 years and mean disease duration was 8.6 ± 2.4 years [Table 1]. The average MER trajectory tract was 1.1 ± 0.2 for each side and recorded depth of the STN was 5.5 ± 0.8 mm. A total of 103 MER neuronal firings were recorded with and without PROM [mean = 5.2 ± 2.9, Table 2]. When compared to baseline (without the PROM test), PROM increased the HR by a mean 0.5 beats/min and MAP by a mean 1.4 mmHg (P = 0.1178 and 0.0525). The associated change in the BIS was − 0.7 (P = 0.4941) and the MAC of the anesthetic agent changed little throughout surgery (P = 0.1825) [Table 3]. No patient was awakened during any PROM test.
Table 1

Characteristics of the ten Parkinson’s disease patients

Mean±SDRange
Sex (female/male)3/7
Age at onset (years)48.5±10.832-65
Disease duration (years)8.6±2.46-14
Age at surgery (years)57.1±9.540-72

SD: Standard deviation

Table 2

Data from microelectrode recording

Case numberBrainMER trajectorySTN recording length (mm)PROM (n)
1Right15.6266
Left15.1004
2Right14.9434
Left15.4965
3Right16.5812
Left16.3972
4Right15.0941
Left15.7201
5Right14.7425
Left14.8245
6Right16.2004
Left16.2657
7Right24.75912
Left15.3157
8Right13.8766
Left16.0049
9Right14.97110
Left14.9274
10Right16.7973
Left16.0876
Mean±SD1.1±0.25.5±0.85.2±2.9

STN: Subthalamic nucleus, PROM: Passive range of motion, MER: Microelectrode recording, SD: Standard deviation, n: Recorded neuronal firings

Table 3

Comparisons between baseline (no passive range of motion) and passive range of motion during microelectrode recording

BaselinePROMP


Mean±SDRangeMean±SDRange
HR (beats/min)69.3±8.961-8969.8±9.060-900.1178
MAP (mmHg)76.1±9.560-9077.5±9.262-920.0525
MAC0.7±0.00.6-0.70.7±0.00.6-0.70.1825
BIS49.0±9.532-6248.3±8.733-620.4941

PROM: Passive range of motion, HR: Heart rate, MAP: Mean arterial pressure, MAC: Minimum alveolar concentration, BIS: Bispectral index, SD: Standard deviation

Characteristics of the ten Parkinson’s disease patients SD: Standard deviation Data from microelectrode recording STN: Subthalamic nucleus, PROM: Passive range of motion, MER: Microelectrode recording, SD: Standard deviation, n: Recorded neuronal firings Comparisons between baseline (no passive range of motion) and passive range of motion during microelectrode recording PROM: Passive range of motion, HR: Heart rate, MAP: Mean arterial pressure, MAC: Minimum alveolar concentration, BIS: Bispectral index, SD: Standard deviation

DISCUSSION

This was the first report showing the effect of PROM on the level of anesthesia during MER under GA. We proved that, under light anesthetic conditions such as between 0.5 and 0.7 MAC, we could obtain effective provoked neuronal firings to identify the somatosensory STN. In addition, PROM did not alter the stable anesthetic status of the patient, as measured by the BIS, HR, and MAP. Our results also answered skepticism on whether the depth of anesthesia would be so light that the patient might awaken during successful MER under GA [6]. The depth of anesthesia is defined as the degree of central nervous system depression by an anesthetic agent. The BIS was introduced in 1992 to monitor the depth of anesthesia. This index is an arbitrary number from 0 (isoelectric) to 100 (awake). Our patients were maintained between a BIS reading of 33–62 during PROM, which might indicate that those patients were in a properly hypnotic state [11]. Since the BIS is a measurement comprising electromyography and electroencephalography, it might sometimes be misrepresented. For example, using medications for neuromuscular block would lead to declines in the BIS in awake persons [14]. Concomitant monitoring with HR and MAP could be a convenient and effective alternative to the BIS, and it has been suggested that standard clinical monitoring may have effects similar to electrical monitoring in preventing the risk of awareness during surgery [1516]. In this study, there were no significant changes in HR and MAP before or after PROM during MER. These patients reported no awareness during surgery. However, there is still controversy in the literature regarding the efficacy of the BIS in monitoring awareness under general anesthesia. The score may not truly represent the level of consciousness [17]. Rare cases of contradictory scores versus level of consciousness have been reported [18]. Despite the aid of HR and MAP, the reliability of BIS monitoring in assisting measurement of awareness or an asleep state requires further verification. It is well documented that, during awake MER procedures, active or passive movement-related changes in neuronal firing within the STN can be indicated as a typical neuronal firing originating from the sensorimotor portion of the STN [23]. We proved in our earlier report that, under volatile inhalation general anesthesia, typical STN firing patterns could be properly observed and triggered by PROM, with clinical outcomes comparable to STN-DBS under awake procedures [67]. Thirty years after the first DBS paper was published at 1987, issues about patient comfort have elicited concern about whether an “awake’ surgical technique is still necessary to obtain reliable and accurate MER signals followed by test stimulation. A recent meta-analysis by Ho et al. revealed that only 16 of 145 validated studies between the years 2004 and 2015 were related to GA. They concluded that DBS under GA may lead to a lower complication rate and there were no significant differences in clinical motor outcomes between awake DBS and DBS under GA [4]. DBS should be recognized as a minimally invasive procedure only if patients can accept the surgery without fear and discomfort. We hope the evolution from awake STN-DBS procedures to surgeries under general anesthesia will become a trend and help provide the benefits of DBS to neuropsychiatric patients who are not suitable for awake DBS. This study had some limitations. First, we designed this study to record both clinical and neurophysiological data, which resulted in a small group of patients. Second, we followed this small cohort without including a control group. Both caveats might explain the insignificant statistical results.

CONCLUSION

PROM was effective in triggering and magnifying neuronal firing signals without influencing patient awareness during MER for STN-DBS under GA. GA is suggested as an alternative choice for the comfort of patients and still ensures uncompromised clinical outcomes.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.
  18 in total

Review 1.  Surgery of the subthalamic nucleus: use of movement-related neuronal activity for surgical navigation.

Authors:  Philip A Starr; Philip V Theodosopoulos; Robert Turner
Journal:  Neurosurgery       Date:  2003-11       Impact factor: 4.654

2.  A combination of electroencephalogram and auditory evoked potentials separates different levels of anesthesia in volunteers.

Authors:  Bettina Horn; Stefanie Pilge; Eberhard F Kochs; Gudrun Stockmanns; Andreas Hock; Gerhard Schneider
Journal:  Anesth Analg       Date:  2009-05       Impact factor: 5.108

3.  Subthalamic deep brain stimulation after anesthetic inhalation in Parkinson disease: a preliminary study.

Authors:  Sheng-Huang Lin; Tsung-Ying Chen; Shinn-Zong Lin; Ming-Hwang Shyr; Yu-Cheng Chou; Wanhua Annie Hsieh; Sheng-Tzung Tsai; Shin-Yuan Chen
Journal:  J Neurosurg       Date:  2008-08       Impact factor: 5.115

4.  Bispectral index aware or minimum alveolar concentration aware?: Alerting thresholds for prevention of awareness.

Authors:  Gerhard Schneider
Journal:  Eur J Anaesthesiol       Date:  2015-05       Impact factor: 4.330

5.  Monitoring depth of anesthesia utilizing a combination of electroencephalographic and standard measures.

Authors:  Gerhard Schneider; Denis Jordan; Gerhard Schwarz; Petra Bischoff; Cornelis J Kalkman; Hermann Kuppe; Ingrid Rundshagen; Adem Omerovic; Matthias Kreuzer; Gudrun Stockmanns; Eberhard F Kochs
Journal:  Anesthesiology       Date:  2014-04       Impact factor: 7.892

6.  Relationship between bispectral index values and volatile anesthetic concentrations during the maintenance phase of anesthesia in the B-Unaware trial.

Authors:  Elizabeth L Whitlock; Alexander J Villafranca; Nan Lin; Ben J Palanca; Eric Jacobsohn; Kevin J Finkel; Lini Zhang; Beth A Burnside; Heiko A Kaiser; Alex S Evers; Michael S Avidan
Journal:  Anesthesiology       Date:  2011-12       Impact factor: 7.892

7.  Subthalamic deep brain stimulation in Parkinson's disease under different anesthetic modalities: a comparative cohort study.

Authors:  Shin-Yuan Chen; Sheng-Tzung Tsai; Sheng-Huang Lin; Tsung-Ying Chen; Hsiang-Yi Hung; Chi-Wei Lee; Wan-Hsiang Wang; Shee-Ping Chen; Shinn-Zong Lin
Journal:  Stereotact Funct Neurosurg       Date:  2011-11-19       Impact factor: 1.875

8.  Awake versus asleep deep brain stimulation for Parkinson's disease: a critical comparison and meta-analysis.

Authors:  Allen L Ho; Rohaid Ali; Ian D Connolly; Jaimie M Henderson; Rohit Dhall; Sherman C Stein; Casey H Halpern
Journal:  J Neurol Neurosurg Psychiatry       Date:  2017-03-01       Impact factor: 10.154

Review 9.  Anaesthetic interventions for prevention of awareness during surgery.

Authors:  Anthony G Messina; Michael Wang; Marshall J Ward; Chase C Wilker; Brett B Smith; Daniel P Vezina; Nathan Leon Pace
Journal:  Cochrane Database Syst Rev       Date:  2016-10-18

Review 10.  Spectral entropy monitoring for adults and children undergoing general anaesthesia.

Authors:  Anjolie Chhabra; Rajeshwari Subramaniam; Anurag Srivastava; Hemanshu Prabhakar; Mani Kalaivani; Saloni Paranjape
Journal:  Cochrane Database Syst Rev       Date:  2016-03-14
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.