Literature DB >> 30279900

Pacemaker malfunction after acute myocardial infarction in a patient with wrap-around left anterior descending artery supplying the right ventricular apex.

Tomohiko C Umei1, Toru Awaya1, Osamu Okazaki1, Hisao Hara1, Yukio Hiroi1.   

Abstract

A 50-year-old man with a dual-chamber pacemaker was admitted to our hospital complaining of chest pain. Anterior ST segment elevation myocardial infarction (STEMI) was diagnosed. Emergency coronary angiography revealed total occlusion of the proximal left anterior descending artery (LAD), and primary percutaneous coronary intervention was performed. Angiograms showed that the LAD was wrapped around the apex of both ventricles. On day 8, ventricular fibrillation and cardiopulmonary arrest occurred due to elevation of the pacing threshold because of pacemaker malfunction. The pacemaker was upgraded to an implantable cardioverter-defibrillator and the lead was inserted into the right ventricular septum. Myocardial scintigraphy with thallium-201 and technetium-99m pyrophosphate located the infarct zone around the apex of both ventricles. We conclude that pacing failure of the right ventricular lead occurred in this case of LAD occlusion due to a LAD supplying the right ventricular apex. Clinicians should be aware of the possibility of pacemaker failure in patients presenting with anterior STEMI due to a wrap-around LAD. <Learning objective: Pacemaker failure due to right ventricular lead malfunction can also occur in patients presenting with myocardial infarction of the left coronary arteries. In our case the culprit lesion was located in the left anterior descending artery, which wrapped around the apex to supply parts of the right ventricle.>.

Entities:  

Keywords:  Myocardial infarction; Myocardial scintigraphy; Pacemaker malfunction; Wrap-around left anterior descending artery

Year:  2018        PMID: 30279900      PMCID: PMC6149590          DOI: 10.1016/j.jccase.2018.03.001

Source DB:  PubMed          Journal:  J Cardiol Cases        ISSN: 1878-5409


  10 in total

1.  Reversible atrial lead dysfunction of DDD pacemaker after recent inferior myocardial infarction--a case report.

Authors:  Chiliang Chen; Ching-Chang Fang; Shih-Pu Wang
Journal:  Angiology       Date:  2004 Jul-Aug       Impact factor: 3.619

Review 2.  The pathology of myocardial infarction in the pre- and post-interventional era.

Authors:  M Pasotti; F Prati; E Arbustini
Journal:  Heart       Date:  2006-04-18       Impact factor: 5.994

3.  Evaluating microvascular obstruction after acute myocardial infarction using cardiac magnetic resonance imaging and 201-thallium and 99m-technetium pyrophosphate scintigraphy.

Authors:  Takayuki Onishi; Isshi Kobayashi; Yuko Onishi; Tomoyuki Kawashima; Hirotaka Muramoto; Hiroaki Nakamura; Yasutoshi Nagata; Shigeo Umezawa; Akihiro Niwa
Journal:  Circ J       Date:  2010-10-21       Impact factor: 2.993

4.  Usefulness of the Left Anterior Descending Artery Wrapping Around the Left Ventricular Apex to Predict Adverse Clinical Outcomes in Patients With Anterior Wall ST-Segment Elevation Myocardial Infarction (an INFUSE-AMI Substudy).

Authors:  Nobuaki Kobayashi; Akiko Maehara; Gary S Mintz; Steven D Wolff; Philippe Généreux; Ke Xu; Roxana Mehran; C Michael Gibson; Sorin J Brener; Gregg W Stone
Journal:  Am J Cardiol       Date:  2015-02-19       Impact factor: 2.778

5.  Pacemaker Troubleshooting: Common Clinical Scenarios.

Authors:  Payam Safavi-Naeini; Mohammad Saeed
Journal:  Tex Heart Inst J       Date:  2016-10-01

6.  Length of left anterior descending coronary artery determines prognosis in acute anterior wall myocardial infarction.

Authors:  Reuben Ilia; Jean Marc Weinstein; Arik Wolak; Harel Gilutz; Carlos Cafri
Journal:  Catheter Cardiovasc Interv       Date:  2014-04-30       Impact factor: 2.692

7.  Technetium-99m pyrophosphate/thallium-201 dual-isotope SPECT imaging predicts reperfusion injury in patients with acute myocardial infarction after reperfusion.

Authors:  Yasushi Akutsu; Kyouichi Kaneko; Yusuke Kodama; Hui-Ling Li; Hideki Nishimura; Yuji Hamazaki; Jumpei Suyama; Akira Shinozuka; Takehiko Gokan; Youichi Kobayashi
Journal:  Eur J Nucl Med Mol Imaging       Date:  2008-09-12       Impact factor: 9.236

Review 8.  Myocardial stimulation threshold in patients with cardiac pacemakers: effect of physiologic variables, pharmacologic agents, and lead electrodes.

Authors:  M L Dohrmann; N F Goldschlager
Journal:  Cardiol Clin       Date:  1985-11       Impact factor: 2.213

9.  The stunned atrial lead: Transient malfunction of a permanent atrial pacer lead following acute myocardial infarction.

Authors:  Shailendra Upadhyay; Stephen Marshalko; Craig McPherson
Journal:  Int J Crit Illn Inj Sci       Date:  2011-07

10.  Early improvement of pacing threshold following primary right coronary angioplasty.

Authors:  Fernando Pivatto Júnior; Diego Chemello; Geris Mazzutti; Maurício Pimentel; Paola Rabaioli; Leandro Zimerman
Journal:  HeartRhythm Case Rep       Date:  2016-09-28
  10 in total