Literature DB >> 30278703

Automated reactive accelerated aging for rapid in vitro evaluation of neural implant performance.

Matthew G Street1, Cristin G Welle2, Pavel A Takmakov1.   

Abstract

Novel therapeutic applications for neural implants require miniaturized devices. Miniaturization imposes stricter requirements for reliability of materials. Pilot clinical studies suggest that rapid failure of the miniaturized neural implants in the body presents a major challenge for this type of technology. Traditional evaluations of neural implant performance over clinically relevant durations present time- and resource-intensive experiments in animals. Reactive accelerated aging (RAA) is an in vitro test platform that was developed to expedite durability testing of neural implants, as a screening technique designed to simulate the aggressive physiological environment experienced by the implants. This approach employs hydrogen peroxide, which mimics reactive oxygen species, and a high temperature to accelerate chemical reactions that lead to device degradation similar to that found with devices implanted in vivo. The original RAA system required daily manual maintenance and was prone to variability in performance. To address these limitations, this work introduces automated reactive accelerated aging (aRAA) with closed-loop monitoring components that make the system simple, robust, and scalable. The core novel technology in the aRAA is electrochemical detection for feedback control of hydrogen peroxide concentration, implemented with simple off-the-shelf components. The aRAA can run multiple parallel experiments for high-throughput device testing and optimization. For this reason, the aRAA provides a simple tool for rapid in vitro evaluation of the durability of neural implants, ultimately expediting the development of a new generation of miniaturized devices with a long functional lifespan.

Entities:  

Mesh:

Substances:

Year:  2018        PMID: 30278703     DOI: 10.1063/1.5024686

Source DB:  PubMed          Journal:  Rev Sci Instrum        ISSN: 0034-6748            Impact factor:   1.523


  3 in total

1.  Activation of UCP1-Independent Ca2+ Cycling Thermogenesis by Wireless Optogenetics.

Authors:  Kenji Ikeda; Kazuki Tajima; Yuji Tanabe; Ada S Y Poon; Shingo Kajimura
Journal:  Methods Mol Biol       Date:  2022

Review 2.  Evaluation methods for long-term reliability of polymer-based implantable biomedical devices.

Authors:  Dong Hyeon Lee; Chae Hyun Kim; Jiman Youn; Joonsoo Jeong
Journal:  Biomed Eng Lett       Date:  2021-04-15

3.  Wireless optogenetics protects against obesity via stimulation of non-canonical fat thermogenesis.

Authors:  Kazuki Tajima; Kenji Ikeda; Yuji Tanabe; Ella A Thomson; Takeshi Yoneshiro; Yasuo Oguri; Marc D Ferro; Ada S Y Poon; Shingo Kajimura
Journal:  Nat Commun       Date:  2020-04-07       Impact factor: 14.919

  3 in total