Literature DB >> 30276109

Preclinical Comparison of Thermal Tissue Effects from Traditional Electrosurgery and a Low-Temperature Electrosurgical Device during Anterior Cervical Discectomy and Fusion.

Kris Radcliff1, Palaniswamy Vijay2, Ruba F Sarris2, Molly Speltz3, Joshua G Vose2.   

Abstract

BACKGROUND: Exposure of the anterior cervical spine requires dissection in proximity to critical neurovascular structures. Monopolar electrosurgical (ES) devices generate heat in contacted tissues, resulting in thermal damage and temperature change. This study examined depth of thermal injury and temperature change associated with use of a low-temperature electrosurgical device (LTD) compared to traditional electrosurgery during a cadaveric anterior cervical discectomy and fusion (ACDF) dissection.
METHODS: ACDF was performed, using ES or LTD, on cervical spines (C3-4 and C4-5) from 2 fresh human cadavers with intact neck soft tissues and no history of surgery. Cadavers were maintained at 22-23°C, and fiber-optic temperature sensors (Neoptix, Québec City, Québec, Canada) were placed near relevant structures to measure changes during dissection. Depth of thermal injury was assessed by hematoxylin and eosin and Masson's trichrome histology of fixed tissue specimens.
RESULTS: Use of the LTD resulted in a statistically significant reduction in temperature change at platysma (3.0 ± 1.04 vs. 11.41 ± 3.10°C, P = .003), carotid sheath (7.32 ± 1.13 vs. 15.57 ± 2.56°C, P = .007), and longus colli (6.11 ± 1.32 vs. 12.9 ± 3.62°C, P = .016) compared to ES. Temperature change at the trachea was similar between groups (6.06 ± 1.99 vs. 4.96 ± 1.89°C, P = .528). Histology showed that LTD produced less mean and maximal depth of thermal injury compared to ES (mean: 0.5 vs. 1.2 mm; max: 0.9 vs. 1.8 mm; P < .05).
CONCLUSIONS: The results of this pilot study demonstrate that anterior cervical spine exposure using an LTD reduces tissue temperature change and depth of thermal injury compared to ES. CLINICAL RELEVANCE: Although exploratory, these results suggest that use of an LTD during ACDF may reduce the extent of thermal tissue injury during dissection. Future studies in live animal models are warranted to determine if thermal injury is a potential cause of common exposure-related complications, such as dysphagia and dysphonia.

Entities:  

Keywords:  anterior cervical discectomy and fusion; electrosurgery; spine surgery; temperature change; thermal injury

Year:  2018        PMID: 30276109      PMCID: PMC6159608          DOI: 10.14444/5059

Source DB:  PubMed          Journal:  Int J Spine Surg        ISSN: 2211-4599


  18 in total

1.  Electrosurgery: history, principles, and current and future uses.

Authors:  Nader N Massarweh; Ned Cosgriff; Douglas P Slakey
Journal:  J Am Coll Surg       Date:  2006-03       Impact factor: 6.113

2.  Comparison of the effects of surgical dissection devices on the rabbit liver.

Authors:  Joel D MacDonald; Christian A Bowers; Steven S Chin; Greg Burns
Journal:  Surg Today       Date:  2013-09-05       Impact factor: 2.549

3.  Comparative healing of surgical incisions created by the PEAK PlasmaBlade, conventional electrosurgery, and a scalpel.

Authors:  Shang A Loh; Grace A Carlson; Edward I Chang; Eric Huang; Daniel Palanker; Geoffrey C Gurtner
Journal:  Plast Reconstr Surg       Date:  2009-12       Impact factor: 4.730

4.  Quantitative comparison of surgical margin histology following excision with traditional electrosurgery and a low-thermal-injury dissection device.

Authors:  Manuel E Ruidiaz; Maria J Cortes-Mateos; Sergio Sandoval; David T Martin; Jessica Wang-Rodriguez; Farnaz Hasteh; Anne Wallace; Joshua G Vose; Andrew C Kummel; Sarah L Blair
Journal:  J Surg Oncol       Date:  2011-07-08       Impact factor: 3.454

5.  Comparative healing of human cutaneous surgical incisions created by the PEAK PlasmaBlade, conventional electrosurgery, and a standard scalpel.

Authors:  Manuel E Ruidiaz; Davorka Messmer; Dominique Y Atmodjo; Joshua G Vose; Eric J Huang; Andrew C Kummel; Howard L Rosenberg; Geoffrey C Gurtner
Journal:  Plast Reconstr Surg       Date:  2011-07       Impact factor: 4.730

6.  Surgical complications of anterior cervical diskectomy and fusion for cervical degenerative disk disease: a single surgeon's experience of 1,576 patients.

Authors:  Anil Nanda; Mayur Sharma; Ashish Sonig; Sudheer Ambekar; Pappireddy Bollam
Journal:  World Neurosurg       Date:  2013-09-18       Impact factor: 2.104

7.  Thermal damage assessment of novel bipolar forceps in a sheep model of spinal surgery.

Authors:  Ebonia W Elliott-Lewis; Jacquelin Jolette; Jandira Ramos; Edward C Benzel
Journal:  Neurosurgery       Date:  2010-07       Impact factor: 4.654

8.  Comparison of the thermal and histopathological effects of bipolar and monopolar electrosurgical resection of the prostate in a canine model.

Authors:  Raymond Ko; Andrew H H Tan; Ben H Chew; P Elaine Rowe; Hassan Razvi
Journal:  BJU Int       Date:  2009-10-08       Impact factor: 5.588

Review 9.  Vertebral artery injury in cervical spine surgery: anatomical considerations, management, and preventive measures.

Authors:  Chan W Peng; Benedict T Chou; John A Bendo; Jeffrey M Spivak
Journal:  Spine J       Date:  2008-05-27       Impact factor: 4.166

10.  Traditional electrosurgery and a low thermal injury dissection device yield different outcomes following bilateral skin-sparing mastectomy: a case report.

Authors:  Richard E Fine; Joshua G Vose
Journal:  J Med Case Rep       Date:  2011-05-28
View more
  1 in total

1.  Predictive nomogram of cage nonunion after anterior cervical discectomy and fusion: A retrospective study in a spine surgery center.

Authors:  Kai Zhou; Longfei Ji; Shuwei Pang; You Tang; Changliang Liu
Journal:  Medicine (Baltimore)       Date:  2022-09-30       Impact factor: 1.817

  1 in total