Ambient Noise Production by High-Frequency Neonatal Ventilators.

Justin Goldstein1, Alyse Laliberte2, Martin Keszler2.   

Abstract

OBJECTIVE: To assess sound levels of 4 high-frequency neonatal ventilators to determine whether there is a safety benefit in using modern high-frequency ventilators compared with older models. STUDY
DESIGN: We performed a bench study comparing noise production of the Sensormedics 3100A Oscillator, Bunnell Life Pulse Jet Ventilators Model 203 and Model 204, and Dräger VN500 in high-frequency mode. A wide range of ventilation settings was examined. All measurements were performed in triplicate using a high-fidelity sound meter, with data analyzed using ANOVA and regression analyses.
RESULTS: The Dräger ventilator was quietest overall, with average sound levels of 49.8 ± 0.49 dB across all settings. The average noise from the Sensormedics was 53.6 ± 2.01 dB, for Bunnell Model 203 was 54.1 ± 1.09 dB, and for Bunnell Model 204 was 53.7 ± 1.45 dB. Adjustments made to frequency/rate and mean airway pressure/positive end-expiratory pressure had minimal effect on noise, and increasing amplitude/peak inspiratory pressure resulted in significantly more noise by all ventilators. At all settings, the Sensormedics and Bunnell ventilators were louder than the Dräger, and the difference became greater as amplitude/peak inspiratory pressure was increased.
CONCLUSIONS: The Dräger VN500 in high-frequency mode produces significantly less noise that both the Sensormedics and Bunnell ventilators. These data suggest that using the Dräger VN500 as a high-frequency ventilator may reduce the potential for adverse outcomes created by ventilator noise.
Copyright © 2018 Elsevier Inc. All rights reserved.

Keywords:  adverse noise effects; decibel level; jet ventilator; oscillator

Mesh:

Year:  2018        PMID: 30268396     DOI: 10.1016/j.jpeds.2018.08.029

Source DB:  PubMed          Journal:  J Pediatr        ISSN: 0022-3476            Impact factor:   4.406


  2 in total

1.  Investigating Noise Exposure to Newborn Infants From Respiratory Support: Methodological Considerations.

Authors:  Devika Singh; Gerhard Fusch
Journal:  Cureus       Date:  2021-11-08

2.  [Noise Level and Frequency Experienced by Premature Infants Receiving Incubator Care in the Neonatal Intensive Care Unit].

Authors:  So Hyun Joo; Tae Im Kim
Journal:  Child Health Nurs Res       Date:  2020-04-30
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.