Considering Blood Dilution improves the Precision of Continuous Whole Blood Glucose Measurements.

Carsten Benesch1, Mareike Kuhlenkötter1, Tim Heise1.   

Abstract

BACKGROUND: One major advantage of automated over manual clamps are continuous measurements of blood glucose concentrations (BG) allowing frequent adaptations in glucose infusion rates (GIR). However, BG measurements might be affected by changes in blood dilution. ClampArt®, a modern automated clamp device, corrects BG measurements for blood dilution, but the impact of this correction is unclear.
METHODS: The authors performed a retrospective analysis of BG during glucose clamps comparing values with a fixed dilution factor with those corrected for the actual blood dilution.
RESULTS: Clamp quality substantially improved with the consideration of blood dilution: Mean accuracy fell from 8.1% ± 2.9% to 4.1% ± 0.8%, precision improved from 9.6 ± 3.6 mg/dl to 3.7 ± 1.3 mg/dl and control deviation from -2.6 ± 4.2 mg/dl to 0.2 ± 0.2 mg/dl.
CONCLUSIONS: Correcting continuous BG measurements for blood dilution significantly increases BG measurement and clamp quality.

Entities:  

Keywords:  ClampArt; clamp quality; control deviation; glucose clamp; pharmacodynamics; type 1 diabetes

Year:  2018        PMID: 30255710      PMCID: PMC6610612          DOI: 10.1177/1932296818803621

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  8 in total

1.  Continuous Blood Glucose Monitoring (Cgm) System Based on Intravenous Microdialysis and Glucose Sensing.

Authors:  H Greiner; R Schaller-Ammann; A Huber; M O'Connell; J Krejci; G Porro; S Korsatko; J Gurban; S Deller; M Brunner; A Berghofer; J Priedl; E Zijlstra; T Heise; T R Pieber; L Schaupp
Journal:  Biomed Tech (Berl)       Date:  2013-09-07       Impact factor: 1.411

2.  Impact of newly developed, next-generation artificial endocrine pancreas.

Authors:  Noriko Kambe; Shinji Kawahito; Naoji Mita; Kazumi Takaishi; Toshiko Katayama; Yoko Sakai; Tomohiro Soga; Hiroaki Kawano; Munehide Matsuhisa; Mitsuo Shimada; Tetsuya Kitagawa; Hiroshi Kitahata
Journal:  J Med Invest       Date:  2015

3.  How to Assess the Quality of Glucose Clamps? Evaluation of Clamps Performed With ClampArt, a Novel Automated Clamp Device.

Authors:  Carsten Benesch; Tim Heise; Oliver Klein; Lutz Heinemann; Sabine Arnolds
Journal:  J Diabetes Sci Technol       Date:  2015-04-07

Review 4.  Euglycaemic glucose clamp: what it can and cannot do, and how to do it.

Authors:  Tim Heise; Eric Zijlstra; Leszek Nosek; Sascha Heckermann; Leona Plum-Mörschel; Thomas Forst
Journal:  Diabetes Obes Metab       Date:  2016-07-13       Impact factor: 6.577

5.  Morning administration of 0.4U/kg/day insulin glargine 300U/mL provides less fluctuating 24-hour pharmacodynamics and more even pharmacokinetic profiles compared with insulin degludec 100U/mL in type 1 diabetes.

Authors:  T S Bailey; J Pettus; R Roussel; W Schmider; M Maroccia; N Nassr; O Klein; G B Bolli; R Dahmen
Journal:  Diabetes Metab       Date:  2017-11-16       Impact factor: 6.041

6.  Development of the Biostator Glucose clamping algorithm.

Authors:  A H Clemens; D L Hough; P A D'Orazio
Journal:  Clin Chem       Date:  1982-09       Impact factor: 8.327

Review 7.  Pulse oximetry.

Authors:  Amal Jubran
Journal:  Crit Care       Date:  2015-07-16       Impact factor: 9.097

Review 8.  The theory and fundamentals of bioimpedance analysis in clinical status monitoring and diagnosis of diseases.

Authors:  Sami F Khalil; Mas S Mohktar; Fatimah Ibrahim
Journal:  Sensors (Basel)       Date:  2014-06-19       Impact factor: 3.576

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.