Impact of newly developed, next-generation artificial endocrine pancreas.

Noriko Kambe1, Shinji Kawahito, Naoji Mita, Kazumi Takaishi, Toshiko Katayama, Yoko Sakai, Tomohiro Soga, Hiroaki Kawano, Munehide Matsuhisa, Mitsuo Shimada, Tetsuya Kitagawa, Hiroshi Kitahata.   

Abstract

BACKGROUND: Recent studies have shown that strict perioperative blood glucose management may reduce mortality and morbidity in critically ill adult patients. The purpose of this study was to assess the accuracy and efficacy of the intraoperative application of a newly developed, next-generation artificial endocrine pancreas (STG-55, Nikkiso Co., Ltd., Tokyo, Japan).
METHODS: Twenty patients scheduled to undergo surgery were enrolled in this study. The STG-55 is designed to be more user-friendly than its conventional counterpart (STG-22) while maintaining the latter's fundamental functions, such as a closed-loop system using algorithms for insulin and glucose infusion. After anesthetic induction, a 20G intravenous catheter was inserted into a peripheral forearm vein and connected to a continuous blood glucose monitor. The resultant 105 scores for paired blood glucose values were compared by Bland-Altman analysis.
RESULTS: Stable blood glucose values were maintained automatically, and there were no complications related to use of the STG-55. A close correlation (r=0.96) was observed between continuous glucose measurements using the STG-55 and conventional intermittent glucose measurements. The difficulty of manipulation using this system was decreased by improved preparation procedures.
CONCLUSION: The glycemic control system using the STG-55 could provide an alternative way to achieve effective and safe perioperative glycemic control.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 25817282     DOI: 10.2152/jmi.62.41

Source DB:  PubMed          Journal:  J Med Invest        ISSN: 1343-1420


  3 in total

1.  Accuracy and Time Delay of Glucose Measurements of Continuous Glucose Monitoring and Bedside Artificial Pancreas During Hyperglycemic and Euglycemic Hyperinsulinemic Glucose Clamp Study.

Authors:  Akio Kuroda; Satoshi Taniguchi; Yuko Akehi; Hiroyasu Mori; Motoyuki Tamaki; Reiko Suzuki; Yinhua Otsuka; Munehide Matsuhisa
Journal:  J Diabetes Sci Technol       Date:  2017-10-09

2.  Considering Blood Dilution improves the Precision of Continuous Whole Blood Glucose Measurements.

Authors:  Carsten Benesch; Mareike Kuhlenkötter; Tim Heise
Journal:  J Diabetes Sci Technol       Date:  2018-09-26

3.  Precision and accuracy of hyperglycemic clamps in a multicenter study.

Authors:  Kieren J Mather; Ashley H Tjaden; Adam Hoehn; Kristen J Nadeau; Thomas A Buchanan; Steven E Kahn; Silva A Arslanian; Sonia Caprio; Karen M Atkinson; Melanie Cree-Green; Kristina M Utzschneider; Sharon L Edelstein
Journal:  Am J Physiol Endocrinol Metab       Date:  2021-03-01       Impact factor: 4.310

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.