Accuracy of Quantification of Iodine and Hounsfield Unit Values on Virtual Monochromatic Imaging Using Dual-Energy Computed Tomography: Comparison of Dual-Layer Computed Tomography With Fast Kilovolt-Switching Computed Tomography.

Hayate Washio, Shingo Ohira, Tsukasa Karino, Yuya Nitta, Masaaki Hayashi1, Masayoshi Miyazaki, Teruki Teshima.   

Abstract

OBJECTIVE: The aim of the study was to compare the accuracy of quantification of iodine and Hounsfield unit (HU) values on virtual monochromatic imaging (VMI) using dual-layer computed tomography (DLCT) and fast kilovolt-switching computed tomography (FKSCT).
MATERIALS AND METHODS: This study was performed in 2 phantoms (large and small) using 16 rods representing different materials (iodine, calcium, blood, and adipose tissue) with different dimensions and concentrations. The absolute percentage errors (absolute ratio of measurement error to true iodine concentration) for iodine concentration and HU value on VMI at 50, 70, and 100 keV were compared between DLCT and FKSCT. The Mann-Whitney U test was used to assess statistical significance.
RESULTS: Overall, the absolute percentage errors for iodine concentration and HU value on VMI were smaller for DLCT than for FKSCT.
CONCLUSIONS: Overall, the accuracy of iodine and HU values was higher for DLCT than for FKSCT.

Entities:  

Mesh:

Substances:

Year:  2018        PMID: 30252833     DOI: 10.1097/RCT.0000000000000798

Source DB:  PubMed          Journal:  J Comput Assist Tomogr        ISSN: 0363-8715            Impact factor:   1.826


  2 in total

1.  Accuracy of virtual monochromatic images generated by the decomposition of photoelectric absorption and Compton scatter in dual-energy computed tomography.

Authors:  Kazuhiro Sato; Chifumi Sato; Ayami Takahashi; Hirokazu Takano; Shingo Kayano; Ayana Ishiguro; Yumi Takane; Tomohiro Kaneta
Journal:  Phys Eng Sci Med       Date:  2022-01-28

2.  Verification of the Dose Reduction Effect via Diluted Injection in Dual-Energy Computed Tomography Using a Human Blood Flow Phantom.

Authors:  Hironobu Tomita; Koichi Shibata
Journal:  Radiol Res Pract       Date:  2019-04-01
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.