Literature DB >> 30229423

Dual-energy CT in the obese: a preliminary retrospective review to evaluate quality and feasibility of the single-source dual-detector implementation.

Noah E Atwi1, David L Smith1, Carson D Flores2, Ekta Dharaiya3, Raman Danrad1, Avinash Kambadakone4, Aran M Toshav5.   

Abstract

PURPOSE: To determine the feasibility of performing dual-energy CT with a single-source spectral detector system in obese patients.
MATERIALS AND METHODS: Retrospective, IRB-approved review of 28 patients weighing ≥ 270 lbs (122 kg) who underwent CT of the abdomen on a single-source spectral detector system was performed. Two blinded, independent radiologists rated relative preference between conventional CT images taken at 120 kVp (CCT120) and monoenergetic 70 keV equivalent (MonoE70) as well as iodine map image quality in the spleen, pancreas, kidneys, and liver. Signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were compared between conventional CT and MonoE70 images and correlated with body habitus markers of weight, height, and abdominal diameter.
RESULTS: MonoE70 images were preferred by radiologists 100% of the time (1-sample t test, p < 0.0001) over conventional CCT120 images. Noise was significantly lower; SNR and CNR were significantly higher in MonoE70 images than in CCT120 images (paired t tests, p < 0.0001). Mean iodine map rating (scale 1-5) was 4.54 ± 0.58, denoting near homogenous and complete iodine mapping through the spleen, pancreas, kidneys, and liver for the majority of patients. Body habitus markers were not significantly correlated with image preference score; noise; MonoE70 SNR; MonoE70 CNR; change in noise, SNR, or CNR from CCT120 to MonoE70, or iodine map quality; ordinal and linear regression, p = 0.2547, p = 0.6837, p =  0.1888, p = 0.5489, p = 0.9830, p = 0.8849, p = 0.8741, p = 0.1522, respectively.
CONCLUSION: The single-source spectral detector implementation of dual-energy CT provides viable, high-quality imaging for obese patients.

Entities:  

Keywords:  DECT; Dual-layer detector DECT; Iodine map; Obesity; Virtual monoenergetic imaging

Mesh:

Year:  2019        PMID: 30229423     DOI: 10.1007/s00261-018-1774-y

Source DB:  PubMed          Journal:  Abdom Radiol (NY)


  2 in total

Review 1.  Recognizing and Minimizing Artifacts at Dual-Energy CT.

Authors:  Anushri Parakh; Chansik An; Simon Lennartz; Prabhakar Rajiah; Benjamin M Yeh; Frank J Simeone; Dushyant V Sahani; Avinash R Kambadakone
Journal:  Radiographics       Date:  2021-02-19       Impact factor: 5.333

2.  The Effect of Image Resolution on Deep Learning in Radiography.

Authors:  Carl F Sabottke; Bradley M Spieler
Journal:  Radiol Artif Intell       Date:  2020-01-22
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.