Alastair Beaven1, Adrian C Gardner1, David S Marks1, Jwalant S Mehta1, Matthew Newton-Ede1, Jonathan B Spilsbury1.
Abstract
STUDYEntities:
Keywords: Complication; Scoliosis; Spine
Year: 2018 PMID: 30213160 PMCID: PMC6147884 DOI: 10.31616/asj.2018.12.5.794
Source DB: PubMed Journal: Asian Spine J ISSN: 1976-1902
Fig. 1.Method of measuring rod length. Region of Interest Window to ensure adequate exposure of the radiographic area of interest. a)Bottom of the measuring aperture (greatest opacity). b)This is found just below the rod collar. c)Measure to top of the measuring aperture (greatest opacity). Express length in millimeters to nearest 0.1 mm, i.e., 18.9 mm in this figure.
Demographics of patients with MCGR rods in this cohort (n=28)
| Patients with MCGR | Value |
|---|---|
| Male | 9 |
| Female | 19 |
| Age | 7 yr 1 mo (61-129 mo) |
| Cobb angle (°) | 61.5 (45-73) |
| Rods inserted | 52 |
| Clinic lengthening episodes per patient | 9 (6-12) |
| Radiographs per patient | 4 (3-5.5) |
Values are presented as number or median (interquartile range).
MCGR, magnetically controlled growth rod.
Outcomes of all rods (n=28)
| Failure | Outcome | Time to failure (yr, mo) | Age at failure (yr, mo) |
|---|---|---|---|
| N | Continue to lengthen | NA | NA |
| N | Continue to lengthen | NA | NA |
| N | Continue to lengthen | NA | NA |
| N | Continue to lengthen | NA | NA |
| N | Continue to lengthen | NA | NA |
| N | Continue to lengthen | NA | NA |
| N | Continue to lengthen | NA | NA |
| N | Continue to lengthen | NA | NA |
| N | Continue to lengthen | NA | NA |
| N | Continue to lengthen | NA | NA |
| N: fail to lengthen in clinic | Lengthened successfully under GA | NA | NA |
| N: fail to lengthen in clinic | Lengthened successfully under GA | NA | NA |
| Y: fail to lengthen under GA | Final fusion | 1, 3 | 6, 10 |
| Y: fail to lengthen under GA | Plan for final fusion | 2, 0 | 13, 1 |
| Y: fail to lengthen under GA | Final fusion | 2, 3 | 11, 7 |
| Y: fail to lengthen under GA | Plan to exchange for further MCGR | 3, 1 | 7, 6 |
| Y: fail to lengthen under GA | Converted to traditional rod | 1, 3 | 12, 0 |
| Y: fail to lengthen under GA | Rods explanted and braced | 1, 3 | 7, 3 |
| Y: fail to lengthen under GA | Final fusion | 1/6 | 12, 11 |
| Y: broken rod | Revised to further MCGR | 1/6 | 7, 1 |
| Y: ran out of rod length | Rods exchanged | 2, 3 | 6, 8 |
| Y: broken actuator pin | Plan to explant | 3, 8 | 7, 8 |
| Y: broken actuator pin | Final fusion | 5, 4 | 14, 5 |
| Y: broken actuator pin | Final fusion | 1, 8 | 8, 0 |
| Y: broken actuator pin | Plan for final fusion | 1, 3 | 12, 2 |
| Y: screws pulling out | Final fusion | 2, 7 | 10, 0 |
| ?: fail to lengthen in clinic | Plan for lengthening under GA | 2, 3 | 6, 8 |
| ?: fail to lengthen in clinic | Plan for lengthening under GA | 4, 0 | 9, 1 |
NA, not applicable; GA, general anesthesia; MCGR, magnetically controlled growth rod.
Patient under investigation for metal allergy.
Single rod revised to two-rod construct.
Syndromic child underwent final fusion in accordance with family’s wishes.