Maged M Roshdy1,2, Sherine S Wahba1,2, Ramy R Fikry2,3.
Abstract
PURPOSE: The aim of this study was to find indices for keratoconus (KC) detection with higher sensitivity and specificity than the currently available ones with the rotating Scheimpflug imaging (Oculyzer I, Pentacam) software. PATIENTS AND METHODS: Data of 103 normal and 73 KC eyes imaged by Oculyzer I, including slit-lamp examination, refraction, Scheimpflug images, and follow-up data were collected. Logistic regression was done to find new indices with higher area under receiver operating characteristic curves using the currently available ones.Entities:
Keywords: elevation; keratoconus; keratoconus diagnosis; keratoconus indices; pachymetry; pentacam
Year: 2018 PMID: 30197498 PMCID: PMC6112792 DOI: 10.2147/OPTH.S171827
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Figure 1Getting the elevation of the thinnest point on mouse click at this exact point.
Abbreviation: BFS, best-fit sphere.
The indices with highest AUROC
| Indices | AUROC | 95% CI | Criterion | Sensitivity | Specificity | AUROC compared with that of CAIRO 9 (P) | AUROC compared with that of CAIRO 8 (P) |
|---|---|---|---|---|---|---|---|
| CAIRO using 9 mm BFS | 0.997 | 0.973–1.0 | ≥0.5 | 95.89 | 97.09 | 0.268 | |
| CAIRO using 8 mm BFS | 0.995 | 0.970–1.0 | ≥0.5 | 93.15 | 98.06 | 0.268 | |
| ART-Max | 0.987 | 0.957–0.998 | ≤412 | 97.26 | 93.2 | 0.046 | 0.087 |
| PPI-Max | 0.987 | 0.958–0.998 | >1.4 | 91.78 | 98.06 | 0.033 | 0.074 |
| Elevation of the thinnest point from the 9 mm anterior BFS | 0.979 | 0.945–0.995 | >7 | 95.89 | 92.23 | 0.108 | 0.138 |
| Elevation of the thinnest point from the 8 mm posterior BFS | 0.979 | 0.945–0.995 | >20 | 86.3 | 100 | 0.017 | 0.024 |
| PPI-Avg | 0.978 | 0.944–0.994 | >1.1 | 87.67 | 98.06 | 0.038 | 0.049 |
| Elevation of the thinnest point from the 9 mm posterior BFS | 0.977 | 0.942–0.994 | >22 | 89.04 | 96.12 | 0.027 | 0.04 |
| ART-Avg | 0.976 | 0.942–0.993 | ≤496 | 94.52 | 94.17 | 0.028 | 0.035 |
| Elevation of the thinnest point from the 7 mm posterior BFS | 0.971 | 0.934–0.990 | >10 | 87.67 | 100 | 0.049 | 0.061 |
| Elevation of the thinnest point from the 8 mm anterior BFS | 0.968 | 0.930–0.989 | >5 | 91.78 | 96.12 | 0.096 | 0.099 |
| Elevation of the thinnest point from the 7 mm posterior BFTE | 0.953 | 0.910–0.979 | >10 | 87.67 | 99.03 | <0.001 | <0.001 |
Abbreviations: 95% CI, 95% confidence interval of the AUROC; ART-Avg, Ambrosio’s relational thickness average; ART-Max, Ambrosio’s relational thickness maximum; AUROC, area under the receiver operating characteristic curve; BFS, best-fit sphere; BFTE, best-fit toric ellipsoid; CAIRO, Corneal Assessment Index from the Relational thickness and other Oculus values using 9 mm BFS (CAIRO 9) and 8 mm BFS (CAIRO 8); P from CAIRO indices, probability of chance that the AUROC of the index is less than that of CAIRO 9 or CAIRO 8 indices; PPI-Avg, pachymetric progression index average; PPI-Max, pachymetric progression index maximum.
Fallacies of each index included in the CAIRO indices
| PPI-Max | ART-Max | Elevation of the thinnest point from the 8 mm anterior BFS | Elevation of the thinnest point from the 9 mm anterior BFS | CAIRO 8 | CAIRO 9 | |
|---|---|---|---|---|---|---|
| False diagnosis rate | 4.55% | 5.11% | 6.25% | 6.82% | 3.98% | 3.41% |
Abbreviations: ART-Max, Ambrosio relational thickness in maximum meridian; BFS, best-fit sphere; CAIRO, Corneal Assessment Index from Relational thickness and other OCULUS values using 8 mm BFS (CAIRO 8) and 9 mm BFS (CAIRO 9).
Figure 2AUROC of CAIRO indices compared with the 4 indices of highest AUROC throughout all cases.
Abbreviations: ART-Max, Ambrosio’s relational thickness maximum; CAIRO, Corneal Assessment Index from the Relational thickness and other Oculus values using 9 mm BFS (CAIRO 9) and 8 mm BFS (CAIRO 8); PPI-max, pachymetric progression index maximum; AE, anterior elevation from 9 mm BFS (AE 9) 8 mm BFS (AE 8).
Figure 3AUROC of CAIRO indices compared with the 4 indices of highest AUROC in differentiating early KC from normal corneas.
Abbreviations: ART-Max, Ambrosio’s relational thickness maximum; AUROC, Area Under Receiver Operating Characteristic curves; CAIRO, Corneal Assessment Index from the Relational thickness and other Oculus values using 9 mm BFS (CAIRO 9) and 8 mm BFS (CAIRO 8); PPI-max, pachymetric progression index maximum; KC, keratoconus; AE, anterior elevation from 9 mm BFS (AE 9) and 8 mm BFS (AE 8).
AUROC of CAIRO indices and some of the conventional indices in both normal and each subdivision of KC groups
| Indices | TKC <2 (n= 23) vs normal (n=103)
| TKC 2–4 (n= 50) vs normal (n=103)
| ||||
|---|---|---|---|---|---|---|
| AUROC | AUROC | |||||
| CAIRO 8 | 0.989 | 0.533 | 0.997 | 0.341 | ||
| CAIRO 9 | 0.992 | 0.533 | – | 0.999 | 0.341 | |
| ART-Avg | 0.934 | 0.028 | 0.033 | 0.996 | 0.651 | 0.135 |
| ART-Max | 0.968 | 0.072 | 0.083 | 0.996 | 0.609 | 0.127 |
| Elevation of the thinnest point from the 8 mm anterior BFS | 0.937 | 0.191 | 0.182 | 0.982 | 0.318 | 0.318 |
| Elevation of the thinnest point from the 8 mm posterior BFS | 0.938 | 0.011 | 0.012 | 0.998 | 0.612 | 0.359 |
Abbreviations: ART-Avg, Ambrosio relational thickness average; ART-Max, Ambrosio relational thickness maximum; AUROC, area under the receiver operating characteristic curve; CAIRO, Corneal Assessment Index from the Relational thickness and other Oculus values; TKC, Topographic Keratoconus Classification.