Literature DB >> 25708211

Comparison of multimetric D index with keratometric, pachymetric, and posterior elevation parameters in diagnosing subclinical keratoconus in fellow eyes of asymmetric keratoconus patients.

Orkun Muftuoglu1, Orhan Ayar2, Volkan Hurmeric2, Faik Orucoglu2, Ilkay Kılıc2.   

Abstract

PURPOSE: To compare the multimetric D index and other keratoconus-screening parameters in patients with clinical keratoconus in 1 eye and subclinical keratoconus in the fellow eye.
SETTING: Medipol University Hospital and Birinci Eye Hospital, Istanbul, Turkey.
DESIGN: Retrospective case-control study.
METHODS: Patients with clinical keratoconus in 1 eye and subclinical keratoconus in the fellow eye and eyes of normal subjects were evaluated with a rotating Scheimpflug imaging system (Pentacam). Parameters included anterior curve analysis, keratometry (K) values, minimum corneal thickness, pachymetric progression index, Ambrósio relational thickness, posterior elevation, back difference elevation, and D-index values. The receiver operating characteristic (ROC) curves were analyzed to evaluate the area under curve (AUC), sensitivity, and specificity of each parameter.
RESULTS: Forty-five patients and 67 normal subjects were evaluated. The pachymetric progression indices, posterior elevation, and the D-index measurements were statistically significantly higher whereas corneal thickness and Ambrósio relational thickness measurements were significantly lower in eyes with keratoconus or subclinical keratoconus than in eyes of normal subjects (P < .05). Using the ROC analysis, the AUC values of the mean steep K, minimum corneal thickness, pachymetric progression index minimum, Ambrósio relational thickness maximum, posterior elevation, back difference elevation, and D index to distinguish between subclinical keratoconus from control subjects were 0.52, 0.64, 0.71, 0.72, 0.71, 0.76, and 0.83, respectively.
CONCLUSION: The new multimetric D index seems to be better than other single-metric parameters in diagnosing keratoconus and subclinical keratoconus with good specificity. However, the sensitivity levels of all parameters were relatively limited in the diagnosis of subclinical keratoconus. FINANCIAL DISCLOSURE: No author has a financial or proprietary interest in any material or method mentioned.
Copyright © 2015 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2015        PMID: 25708211     DOI: 10.1016/j.jcrs.2014.05.052

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  22 in total

1.  Distinguishing Highly Asymmetric Keratoconus Eyes Using Combined Scheimpflug and Spectral-Domain OCT Analysis.

Authors:  Eric S Hwang; Claudia E Perez-Straziota; Sang Woo Kim; Marcony R Santhiago; J Bradley Randleman
Journal:  Ophthalmology       Date:  2018-07-25       Impact factor: 12.079

Review 2.  The correlation between keratoconus and eye rubbing: a review.

Authors:  Hatim Najmi; Yara Mobarki; Khalid Mania; Bashaer Altowairqi; Mohammed Basehi; Mohammed Salih Mahfouz; Mona Elmahdy
Journal:  Int J Ophthalmol       Date:  2019-11-18       Impact factor: 1.779

3.  Best indices of dual Scheimpflug/Placido tomographer for keratoconus detection.

Authors:  Zahra Heidari; Ebrahim Jafarzadehpour; Mehrdad Mohammadpour; Hassan Hashemi
Journal:  Int Ophthalmol       Date:  2022-09-23       Impact factor: 2.029

4.  Performances of Corneal Topography and Tomography in the Diagnosis of Subclinical and Clinical Keratoconus.

Authors:  Cristina Ariadna Nicula; Adriana Elena Bulboacă; Dorin Nicula; Ariadna Patricia Nicula; Karin Ursula Horvath; Sorana D Bolboacă
Journal:  Front Med (Lausanne)       Date:  2022-05-26

5.  Clinical Study on the Differential Diagnosis of High Myopia Astigmatism and Subclinical Keratoconus in Adolescents by Pentacam Anterior Segment Analyzer.

Authors:  Yimeng Ruan; Ying Zhang; Xiaohong Ying
Journal:  Contrast Media Mol Imaging       Date:  2022-05-16       Impact factor: 3.009

6.  Characteristic of entire corneal topography and tomography for the detection of sub-clinical keratoconus with Zernike polynomials using Pentacam.

Authors:  Zhe Xu; Weibo Li; Jun Jiang; Xiran Zhuang; Wei Chen; Mei Peng; Jianhua Wang; Fan Lu; Meixiao Shen; Yuanyuan Wang
Journal:  Sci Rep       Date:  2017-11-28       Impact factor: 4.379

7.  Changes of corneal tomography in patients with congenital blepharoptosis.

Authors:  Tiepei Zhu; Xin Ye; Peifang Xu; Jingyi Wang; Huina Zhang; Hailong Ni; Zhaoan Su; Juan Ye
Journal:  Sci Rep       Date:  2017-07-26       Impact factor: 4.379

8.  A predictive model for early diagnosis of keratoconus.

Authors:  Gracia Castro-Luna; Antonio Pérez-Rueda
Journal:  BMC Ophthalmol       Date:  2020-07-02       Impact factor: 2.209

9.  New corneal assessment index from the relational thickness and other OCULUS values (CAIRO Index).

Authors:  Maged M Roshdy; Sherine S Wahba; Ramy R Fikry
Journal:  Clin Ophthalmol       Date:  2018-08-23

10.  Objective and subjective diagnostic parameters in the fellow eye of unilateral keratoconus.

Authors:  Eman A Awad; Waleed A Abou Samra; Magda A Torky; Amr M El-Kannishy
Journal:  BMC Ophthalmol       Date:  2017-10-06       Impact factor: 2.209