Overexpansion of the SAPIEN 3 Transcatheter Heart Valve: An Ex Vivo Bench Study.

Janarthanan Sathananthan1, Stephanie Sellers2, Aaron Barlow2, Rob Fraser3, Viktória Stanová4, Anson Cheung1, Jian Ye1, Abdullah Alenezi1, Dale J Murdoch1, Mark Hensey1, Danny Dvir5, Philipp Blanke1, Régis Rieu4, David Wood1, Phillippe Pibarot6, Jonathan Leipsic1, John Webb7.   

Abstract

OBJECTIVES: This study assessed the effect of overexpansion beyond labeled size (diameter) of transcatheter heart valves through an ex vivo bench study.
BACKGROUND: Transcatheter heart valves function optimally when expanded to specific dimensions. However, clinicians may sometimes wish to overexpand balloon-expandable valves to address specific clinical challenges. The implications of overexpansion have assumed considerable importance, and objective information to guide practice is limited.
METHODS: We evaluated SAPIEN 3 transcatheter heart valves (Edwards Lifesciences, Irvine, California). Valves (diameters of 23, 26, and 29 mm) were expanded to nominal dimensions, and then incrementally overexpanded with balloons sized 1-, 2-, and 3-mm larger than the recommended diameter. Valves underwent visual, microcomputed tomography, and hydrodynamic evaluation at various degrees of overexpansion.
RESULTS: SAPIEN 3 valves with labeled diameters of 23, 26, and 29 mm could be incrementally overexpanded to midvalve diameters of 26.4, 28.4, and 31.2 mm, respectively. With overexpansion, there was visible restriction of the valve leaflets, which was particularly evident with the smaller valves. After maximal overexpansion of a 26-mm valve a leaflet tear was observed. High-speed video demonstrated impaired leaflet motion of both the 23- and 26-mm valves and hydrodynamic testing documented a regurgitant fraction for the 23- and 26-mm valves above accepted international standards. The maximally overexpanded 29-mm SAPIEN 3 still had relatively normal leaflet motion and excellent hydrodynamic function. Durability was not specifically evaluated.
CONCLUSIONS: Overexpansion of balloon-expandable valves is possible. However, excessive overexpansion may be associated with impaired hydrodynamic function, acute leaflet failure, and reduced durability. Smaller valves may be at greater risk with overexpansion than larger valves. Overexpansion is best avoided unless clinical circumstances are compelling.
Copyright © 2018 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  overexpansion; transcatheter heart valve

Mesh:

Year:  2018        PMID: 30190060     DOI: 10.1016/j.jcin.2018.06.027

Source DB:  PubMed          Journal:  JACC Cardiovasc Interv        ISSN: 1936-8798            Impact factor:   11.195


  7 in total

1.  A geometrically adaptable heart valve replacement.

Authors:  Sophie C Hofferberth; Mossab Y Saeed; Lara Tomholt; Matheus C Fernandes; Christopher J Payne; Karl Price; Gerald R Marx; Jesse J Esch; David W Brown; Jonathan Brown; Peter E Hammer; Richard W Bianco; James C Weaver; Elazer R Edelman; Pedro J Del Nido
Journal:  Sci Transl Med       Date:  2020-02-19       Impact factor: 17.956

2.  Treating patients with excessively large annuli with self-expanding transcatheter aortic valves: insights into supra-annular structures that anchor the prosthesis.

Authors:  Tian-Yuan Xiong; Yan-Biao Liao; Yi-Jian Li; Fei Chen; Yuanweixiang Ou; Xi Wang; Zi-Jie Wang; Xi Li; Zhen-Gang Zhao; Wei Meng; Yuan Feng; Mao Chen
Journal:  Herz       Date:  2020-09-03       Impact factor: 1.443

3.  Percutaneous aortic and mitral valve repair - from bench testing to simulators and clinical data: Percutaneous aortic and mitral valve repair.

Authors:  Paul T L Chiam; Huay Cheem Tan
Journal:  AsiaIntervention       Date:  2020-12-02

Review 4.  Transcatheter Bioprosthetic Aortic Valve Dysfunction: What We Know So Far.

Authors:  Fadi Sawaya; Troels H Jørgensen; Lars Søndergaard; Ole De Backer
Journal:  Front Cardiovasc Med       Date:  2019-10-04

5.  SAPIEN 3 transcatheter heart valve in patients with borderline annulus: "stretched better than bigger?"

Authors:  Giuseppe Tarantini; Francesco Cardaioli
Journal:  Catheter Cardiovasc Interv       Date:  2022-09       Impact factor: 2.585

Review 6.  Tissue Engineered Transcatheter Pulmonary Valved Stent Implantation: Current State and Future Prospect.

Authors:  Xiling Zhang; Thomas Puehler; Jette Seiler; Stanislav N Gorb; Janarthanan Sathananthan; Stephanie Sellers; Assad Haneya; Jan-Hinnerk Hansen; Anselm Uebing; Oliver J Müller; Derk Frank; Georg Lutter
Journal:  Int J Mol Sci       Date:  2022-01-10       Impact factor: 5.923

Review 7.  Choice of transcatheter heart valve: should we select the device according to each patient's characteristics or should it be "one valve fits all"?

Authors:  Matthias Renker; Won-Keun Kim
Journal:  Ann Transl Med       Date:  2020-08
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.