Literature DB >> 30189220

Glistening Formation and Light Scattering in Six Hydrophobic-Acrylic Intraocular Lenses.

Grzegorz Łabuz1, Dominik Knebel1, Gerd U Auffarth2, Hui Fang1, Thomas Jtp van den Berg3, Timur M Yildirim1, Hyeck-Soo Son1, Ramin Khoramnia1.   

Abstract

PURPOSE: To study the glistening formation in various hydrophobic-acrylic intraocular lens (IOL) models, and to evaluate the effect of glistenings on light scattering in these IOLs.
DESIGN: Laboratory investigation.
METHODS: The susceptibility of the hydrophobic-acrylic material to develop glistenings was evaluated in 6 IOL models. Accelerated lens aging was induced by immersing the IOLs in a solution at 45 C for 24 hours and cooled to 37 C for 2.5 hours. Light microscopy and image acquisition were performed. Glistening statistics-that is, microvacuoles' (MV) number and size-were derived from image analysis. Light scattering was measured using a clinical device featuring an adaptation for in vitro IOL assessment.
RESULTS: The number of glistenings differed among the studied IOLs and ranged from 0 to 3532 MV/mm2. In 1 model, glistenings were found only at the periphery, with diffuse light scattering observed centrally despite the absence of microvacuoles. The mean size of glistenings ranged from 5.2 to 10.2 μm. The mean straylight parameter of the IOLs increased from 0.6 to 5.0 deg2/sr after accelerated aging. Straylight elevation demonstrated a proportional relationship with the glistening number.
CONCLUSIONS: We showed that hydrophobic-acrylic lenses differ in their resistance to glistenings, as one group proved to be glistening-free, but the other models revealed varying grades of glistenings. Moreover, we demonstrated that the presence of glistenings results in increased straylight, and that straylight proportionally depends on the glistenings number irrespective of the IOL model. However, more research is needed to confirm that the relationship we found holds for all hydrophobic-acrylic IOLs.
Copyright © 2018 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2018        PMID: 30189220     DOI: 10.1016/j.ajo.2018.08.032

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  10 in total

1.  IOLs glistenings and quality of vision.

Authors:  Andrzej Grzybowski; Piotr Kanclerz; George H H Beiko
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2019-10-19       Impact factor: 3.117

Review 2.  Etiology, pathogenesis, and management of acute intraocular lens opacification: a systematic review.

Authors:  Bita Momenaei; Mohammad Reza Akbari; Seyed Ali Tabatabaei; Mohammad Soleimani; Mahdi Soleymanzadeh; Kaveh Fadakar; Ahmed Alshaheeb; Mahtab Malekpour Khazari; Yasaman Vaseghi; Kasra Cheraqpour
Journal:  Int J Ophthalmol       Date:  2022-07-18       Impact factor: 1.645

3.  Variation in intraocular lens calcification under different environmental conditions in eyes with supplementary sulcus-supported lenses.

Authors:  Timur M Yildirim; Ramin Khoramnia; Sonja K Schickhardt; Donald J Munro; Patrick R Merz; Hyeck-Soo Son; Ingo Lieberwirth; Gerd U Auffarth
Journal:  Am J Ophthalmol Case Rep       Date:  2020-06-24

4.  Development of a standardized in vitro model to reproduce hydrophilic acrylic intraocular lens calcification.

Authors:  Leoni Britz; Sonja Katrin Schickhardt; Timur Mert Yildirim; Gerd Uwe Auffarth; Ingo Lieberwirth; Ramin Khoramnia
Journal:  Sci Rep       Date:  2022-05-10       Impact factor: 4.996

5.  Polarized glasses may help in symptomatic cases of intraocular lens glistenings.

Authors:  Andreas F Borkenstein; Eva-Maria Borkenstein
Journal:  Clin Optom (Auckl)       Date:  2019-05-06

6.  Quantitative evaluation of microvacuole formation in five intraocular lens models made of different hydrophobic materials.

Authors:  Timur M Yildirim; Sonja K Schickhardt; Qiang Wang; Elfriede Friedmann; Ramin Khoramnia; Gerd U Auffarth
Journal:  PLoS One       Date:  2021-04-30       Impact factor: 3.240

7.  Effectiveness and Safety of the Clareon Monofocal Intraocular Lens: Outcomes from a 12-Month Single-Arm Clinical Study in a Large Sample.

Authors:  Robert Lehmann; Andrew Maxwell; David M Lubeck; Raymond Fong; Thomas R Walters; Anna Fakadej
Journal:  Clin Ophthalmol       Date:  2021-04-20

8.  Refractive and Visual Outcomes of a Monofocal Non-Constant Aberration Aspheric Intraocular Lens.

Authors:  Adrián Hernández-Martínez; Miguel A Díaz-Del-Rio; María Ruiz-Santos; Ramón Ruiz-Mesa; Pedro Tañá-Rivero
Journal:  Clin Ophthalmol       Date:  2022-08-10

9.  Glistening formation in a new hydrophobic acrylic intraocular lens.

Authors:  Timur M Yildirim; Hui Fang; Sonja K Schickhardt; Qiang Wang; Patrick R Merz; Gerd U Auffarth
Journal:  BMC Ophthalmol       Date:  2020-05-06       Impact factor: 2.209

10.  Comparative analysis of in vitro accelerated glistening formation in foldable hydrophobic intraocular lenses.

Authors:  Tamer Tandogan; Gerd U Auffarth; Chul Young Choi; Hyeck-Soo Son; Ramin Khoramnia
Journal:  Int Ophthalmol       Date:  2021-05-19       Impact factor: 2.031

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.