| Literature DB >> 30186906 |
Brett G Brazier1, J Wesley Mesko2.
Abstract
This case report looks at the failure of a vitamin E-infused highly cross-linked polyethylene acetabular liner that fractured at the superior rim of the locking mechanism. This calls to question whether there is a problem with the vitamin E-infused product or a flaw in the design of the liner or possibly a combination of the 2. Although there has been discussion of a possible minimum rim polyethylene thickness, there has not been a minimum thickness established. In addition, in the case of offset liners, indentations which are a part of a locking mechanism can cause the minimum rim dimension to be much less than the specified nominal thickness. This case serves as an interesting observation and contributes to the discussion of rim liner polyethylene thickness and minimum rim thickness.Entities:
Keywords: Liner; Locking mechanism; Polyethylene; Rim fracture; Rim thickness; Vitamin E infused
Year: 2018 PMID: 30186906 PMCID: PMC6123179 DOI: 10.1016/j.artd.2018.02.001
Source DB: PubMed Journal: Arthroplast Today ISSN: 2352-3441
Figure 1Anteroposterior (AP) radiograph of the pelvis made 6 weeks postoperatively showing acceptable position of the components.
Figure 2Cross-table lateral radiograph of the left hip 6 weeks postoperatively showing 45° of anteversion of the acetabular component.
Figure 3AP radiograph of the pelvis made 25 months postoperatively showing anterior and superior dislocation of the femoral head and was later discovered to be due to polyethylene fracture.
Figure 4Photograph of fractured polyethylene component at the superior rim into 2 fragments with no visible thinning of the liner.
Figure 5AP radiograph of the pelvis made 6 weeks postoperatively following revision of hip components.