Literature DB >> 26692248

3-year follow-up of a long-term registry-based multicentre study on vitamin E diffused polyethylene in total hip replacement.

Nanna H Sillesen1, Meridith E Greene1, Audrey K Nebergall1, James I Huddleston2, Roger Emerson3, Peter Gebuhr4, Anders Troelsen4, Henrik Malchau1.   

Abstract

PURPOSE: Preclinical studies of vitamin E diffused highly cross-linked polyethylene (E-XLPE) has shown enhanced mechanical properties with less wear. The purpose of our study was to document the early clinical outcome of E-XLPE to ensure, for the safety of patients, that there are no unforeseen early adverse events from using this new biomaterial.
METHODS: The enrolled patients (n = 977) have received either a porous titanium coated or porous plasma sprayed acetabular shell with either a E-XLPE liner or a medium cross-linked (AXL) liner. At each follow-up 5 patient-reported outcome measures (PROMs) were completed: Harris Hip Score, Case Mix Indicator, UCLA Activity Score, SF-36, and EQ-5D. Radiographs were measured for cup and stem position, as well as femoral head penetration into the liner (wear). Postoperative complications and revisions were collected.
RESULTS: At 3 years follow-up, there were 13 revisions due to: 4 periprosthetic fractures; 1 sepsis; 6 instabilities; and 2 implant mismatches at surgery. Wear of the hip implant, calculated by software analysis of pelvis x-rays, from the postoperative interval to 3 years showed a liner penetration rate of 0.027 mm/year for AXL and 0.005 mm/year for E-XLPE with no significant difference between the 2 (p = 0.24). Improvement was seen in all PROMs from the preoperative interval to 3 years after surgery (p<0.0001).
CONCLUSIONS: Early follow-up of the E-XLPE and AXL liners show low penetration. PROMs indicate improvement after total hip arthroplasty in functionality and quality of life across the centers. We have not observed any early in vivo adverse effects from diffusing the liners with vitamin E. TRIAL REGISTRATION: ClinicalTrials.gov Identifier: NCT00545285.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26692248     DOI: 10.5301/hipint.5000297

Source DB:  PubMed          Journal:  Hip Int        ISSN: 1120-7000            Impact factor:   2.135


  7 in total

1.  Reasons for Revision, Oxidation, and Damage Mechanisms of Retrieved Vitamin E-Stabilized Highly Crosslinked Polyethylene in Total Knee Arthroplasty.

Authors:  Hannah Spece; Jaclyn T Schachtner; Daniel W MacDonald; Gregg R Klein; Michael A Mont; Gwo-Chin Lee; Steven M Kurtz
Journal:  J Arthroplasty       Date:  2019-07-18       Impact factor: 4.757

2.  Cup positioning and its effect on polyethylene wear of vitamin E- and non-vitamin E-supplemented liners in total hip arthroplasty: radiographic outcome at 5-year follow-up.

Authors:  Josef Baghdadi; Shareef Alkhateeb; Alexander Roth; Marcus Jäger
Journal:  Arch Orthop Trauma Surg       Date:  2022-04-10       Impact factor: 3.067

3.  Implant survival of 2,723 vitamin E-infused highly crosslinked polyethylene liners in total hip arthroplasty: data from the Finnish Arthroplasty Register.

Authors:  Matias Hemmilä; Inari Laaksonen; Markus Matilainen; Antti Eskelinen; Jaason Haapakoski; Ari-Pekka Puhto; Jukka Kettunen; Konsta Pamilo; Keijo T Mäkelä
Journal:  Acta Orthop       Date:  2021-02-01       Impact factor: 3.717

Review 4.  Ultra-High Molecular Weight Polyethylene: Influence of the Chemical, Physical and Mechanical Properties on the Wear Behavior. A Review.

Authors:  Pierangiola Bracco; Anuj Bellare; Alessandro Bistolfi; Saverio Affatato
Journal:  Materials (Basel)       Date:  2017-07-13       Impact factor: 3.623

5.  Vitamin E-Enhanced Liners in Primary Total Hip Arthroplasty: A Systematic Review and Meta-Analysis.

Authors:  Qian-Yue Cheng; Bin-Fei Zhang; Peng-Fei Wen; Jun Wang; Lin-Jie Hao; Tao Wang; Hui-Guang Cheng; Ya-Kang Wang; Jian-Bin Guo; Yu-Min Zhang
Journal:  Biomed Res Int       Date:  2021-12-06       Impact factor: 3.411

6.  Catastrophic failure of a dual mobility bearing in a revision total hip arthroplasty.

Authors:  Brett G Brazier; Ethan R Wren; Michael J Blackmer; Dominic V Marino; Jason M Cochran
Journal:  Arthroplast Today       Date:  2018-04-22

7.  Superior rim fracture of a vitamin E-infused highly cross-linked polyethylene (HXLPE) liner leading to total hip arthroplasty revision.

Authors:  Brett G Brazier; J Wesley Mesko
Journal:  Arthroplast Today       Date:  2018-03-21
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.