| Literature DB >> 30185196 |
Dasheng Lin1,2, Lei Wang3, Zhaoliang Yu4, Deqing Luo3, Xigui Zhang5, Kejian Lian3.
Abstract
BACKGROUND: Treatment for osteonecrosis of the femoral head (ONFH) in young individuals remains controversial. We developed a lantern-shaped screw, which was designed to provide mechanical support for the femoral head to prevent its collapse, for the treatment of ONFH. The purpose of this study was to investigate the efficacy and safety of the lantern-shaped screw loaded with autologous bone for the treatment of pre-collapse stages of ONFH.Entities:
Keywords: Autogenous bone graft; Core decompression; Lantern-shaped screw; Osteonecrosis of the femoral head
Mesh:
Year: 2018 PMID: 30185196 PMCID: PMC6123930 DOI: 10.1186/s12891-018-2243-z
Source DB: PubMed Journal: BMC Musculoskelet Disord ISSN: 1471-2474 Impact factor: 2.362
Fig. 1The lantern-shaped screw. a Photo of the folding lantern-shaped screw. b Photo of the unfolding lantern-shaped screw. c Photo of the structure and composition of the lantern-shaped screw. d Photo of the supporting equipment for the lantern-shaped screw. e Photo of the lantern-shaped screw and the angle plate with a locking tail cap (black arrow)
Baseline patient characteristics
| Characteristic | Lantern-shaped screw group ( | Control group ( | |
|---|---|---|---|
| Mean age (range), yr. | 31.5 ± 1.9 (16–47) | 32.6 ± 1.7 (18–44) | 0.6622 |
| Female/ male, n | 3/ 13 | 4/ 12 | 1.0000 |
| Right/ left hip involved, n | 9/ 7 | 10/ 6 | 0.7189 |
| Body mass index (range), kg/cm2 | 24.9 ± 0.5 (21–27) | 24.3 ± 0.5 (21–26) | 0.3978 |
| Etiology | 0.4652 | ||
| Nontraumatic, n | 9 | 11 | |
| Traumatic, n | 7 | 5 | |
| Duration of illness (range), mo | 18.8 ± 2.0 (9–36) | 20.5 ± 1.8 (11–38) | 0.5379 |
| ARCO stage (II/ III) | 7/ 9 | 8/ 8 | 0.7232 |
| Bone marrow edema of hip (III) | 5 | 6 | 0.7097 |
| Mean operation time (range), min | 80.13 ± 3.3 (56–105) | 71.4 ± 3.1 (45–90) | 0.0621 |
| Mean blood loss (range), mL | 180.6 ± 14.9 (110–350) | 158.8 ± 13.7 (90–280) | 0.2892 |
| Harris hip score | |||
| Pre-operation (range) | 60.3 ± 1.8 (48–72) | 61.1 ± 2.0 (48–76) | 0.7457 |
| 12 months (range) | 86.3 ± 0.98 (81–94) | 80.8 ± 1.5 (67–90) | 0.0054 |
| 24 months (range) | 85.6 ± 1.3 (70–92) | 79.4 ± 1.7 (64–91) | 0.0078 |
| 36 months (range) | 85.2 ± 1.7 (68–95) | 78.2 ± 2.2 (64–93) | 0.0173 |
| Successful clinical results (%) | 15 (94) | 10 (63) | 0.0325 |
| Successful radiological results (%) | 14 (88) | 8 (50) | 0.0221 |
Fig. 2Intraoperative radiography for evaluation of surgical procedure. a Insertion of the lantern-shaped screw. b The lantern-shaped screw was being unfolded. c The lantern-shaped screw was the achievement of surface at surface support for the subchondral bone of femoral head
Fig. 3Representative radiographic images from both preoperative and postoperative taken at immediately after the lantern-shaped screw implantation and 36 months. a-c Preoperative X-ray, CT and sagittal T2-weighted magnetic resonance image showing ARCO stage III ONFH in a man aged 29 years. d and e X-ray anteroposterior view on the day of surgery. f-h X-ray and CT scans at 36 months after surgery showing union
Fig. 4The figure shows survival with requirement for further hip surgery as the endpoint. The survival rate was different between the lantern-shaped screw group (94%) and the control group (69%) at 36 months (P = 0.0628)