Paul S Rosen1,2, Herman Sahlin3, Rudolf Seemann4, Ari S Rosen5.
Abstract
BACKGROUND: This retrospective consecutive case series study was performed to determinate the survival rate and implant stability of short (7 mm length) dental implants with an electrowetted hydrophilic surface that were in function from 1 to 7 years.Entities:
Keywords: Cumulative survival rate; Electrowetted surface; Implant stability; Short implants
Year: 2018 PMID: 30136010 PMCID: PMC6105540 DOI: 10.1186/s40729-018-0136-4
Source DB: PubMed Journal: Int J Implant Dent ISSN: 2198-4034
Fig. 1a Initial radiograph exposed at abutment installation on a 5-mm-wide by 7-mm-length dental implant used to help support a removable partial denture for a 71-year-old Caucasian male. b Radiograph of the area taken at 82 months demonstrates good bone stability. c Clinical image of the area 82 months later. Soft tissue remains healthy. The two teeth anterior are in the process of receiving a two-unit fixed splint
Fig. 2a Radiographic image of a 3.5-mm-wide by 7-mm-length dental implant at the time of its placement at the mandibular left second premolar in a 63-year-old Caucasian female. b Radiographic image of a three-unit fixed partial denture upon its initial placement. The dental implant is the anterior abutment with the prosthesis screw retained to it. The distal abutment is the mandibular left second molar. A coping has been placed in the event of intrusion. c Clinical image of the three-unit fixed partial denture at 58 months post-loading. Soft tissue health has been maintained around both the tooth and the dental implant. d Radiograph exposed at 58 months demonstrating the steady state of bone on both the dental implant and the natural tooth abutments. There is no suggestion of any intrusion occurring with the mandibular left second molar
Patient demographics
| Number | Percent | ||
|---|---|---|---|
| Age (years) | 20–29 | 1 | 1.3 |
| 30–39 | 4 | 5.3 | |
| 40–49 | 6 | 8.0 | |
| 50–59 | 19 | 25.3 | |
| 60–69 | 27 | 36.0 | |
| 70–79 | 12 | 16.0 | |
| 80–89 | 6 | 8.0 | |
| Sex | Female | 48 | 64.0 |
| Male | 27 | 26.0 | |
| Smoker | No | 69 | 92.0 |
| Yes | 6 | 8.0 | |
| Diabetes | No | 71 | 94.7 |
| Yes | 4 | 5.3 | |
| Osteoporosis | No | 66 | 88.0 |
| Osteopenia | 5 | 6.7 | |
| Osteoporosis | 4 | 5.3 | |
| Penicillin-VK allergy | No | 65 | 86.7 |
| Yes | 10 | 13.3 | |
Implant and site-related specifications
| Number | Percent | ||
|---|---|---|---|
| Jaw | Maxilla | 60 | 69.8 |
| Mandible | 26 | 30.2 | |
| Position | Anterior | 2 | 2.3 |
| Posterior | 84 | 87.7 | |
| Implant diameter | 3.5 mm | 4 | 4.7 |
| 4.0 mm | 14 | 16.3 | |
| 4.5 mm | 13 | 15.1 | |
| 5.0 mm | 21 | 24.4 | |
| 5.5 mm | 32 | 37.2 | |
| 6.0 mm | 2 | 2.3 | |
| Site type | Healed site | 48 | 55.8 |
| Immediate extraction | 31 | 36.0 | |
| Implant replacement | 1 | 1.2 | |
| Sinus | 6 | 7.0 | |
| Bone quality (Lekholm/Zarb) | Type 1 | 4 | 4.7 |
| Type 2 | 54 | 62.8 | |
| Type 3 | 26 | 30.2 | |
| Type 4 | 2 | 2.3 | |
| Insertion torque (Ncm) | 10–19 | 6 | 7.0 |
| 20–29 | 14 | 16.3 | |
| 30–39 | 59 | 68.6 | |
| 40–49 | 4 | 4.7 | |
| 50 | 3 | 3.5 | |
| Implant stability at placement (ISQ) | 30–39 | 1 | 1.2 |
| 40–49 | 1 | 1.2 | |
| 50–59 | 1 | 1.2 | |
| 60–69 | 17 | 19.8 | |
| 70–79 | 44 | 51.2 | |
| 80–89 | 22 | 25.6 | |
| Surgery | One-stage | 44 | 51.2 |
| Two-stage | 42 | 48.8 | |
| Type of restoration | Crown | 48 | 55.8 |
| Bridge | 35 | 40.7 | |
| Locator | 3 | 3.5 | |
Specification of failed implants (three implants in three patients)
| Sex | Age | Smoker | Risk factors | Position | Implant diameter | Insert. torque | ISQ at insert. | Time of failure |
|---|---|---|---|---|---|---|---|---|
| Male | 75 | No | Diabetes | Maxillary premolar | 5.5 mm | 20 | 75 | 0 months |
| Male | 73 | No | No | Maxillary premolar | 4.5 mm | 32 | 80 | 60 months |
| Female | 77 | No | Osteoporosis | Mandibular molar | 5.0 mm | 32 | 70 | 0 months |
Implant survival, life table analysis
| Interval | Implants | Failed | Not followed | CSR (%) |
|---|---|---|---|---|
| Insertion to 1 year | 86 | 2 | 3 | 97.7 |
| 1 to 2 years | 81 | 0 | 19 | 97.7 |
| 2 to 3 years | 62 | 0 | 6 | 97.7 |
| 3 to 4 years | 56 | 0 | 6 | 97.7 |
| 4 to 5 years | 50 | 0 | 16 | 97.7 |
| 5 to 6 years | 34 | 1 | 16 | 94.8 |
| 6 to 7 years | 17 | 0 | 15 | 94.8 |
| 7 years | 2 | – | – | – |
Resonance frequency analysis
| Time | ISQ |
|
|---|---|---|
| Implant insertion | 73.6 ± 8.1 | 86 |
| 1–4 weeks | 79.4 ± 4.1 | 28 |
| 5–8 weeks | 77.3 ± 5.0 | 34 |
| 9–12 weeks | 74.9 ± 5.6 | 20 |
| 13–16 weeks | 73.3 ± 4.4 | 11 |